Effect of Mobile Phase pH on the Electrospray Ionization Efficiency and Qualitative Analysis of Pharmaceuticals in ESI + LC-MS/MS

The effect of mobile phase pH on positive ionization process and retention time of nine pharmaceuticals on ultra-performance liquid chromatography-electrospray-tandem mass spectrometry (LC-MS/MS) was discussed. The effective use of high and low mobile phase pH in LC-MS/MS qualitative analyses method...

Full description

Bibliographic Details
Main Authors: Bakar, N.K.A (Author), Kafeenah, H.I.S (Author), Osman, R. (Author)
Format: Article
Language:English
Published: Oxford University Press 2019
Subjects:
ion
pH
Online Access:View Fulltext in Publisher
View in Scopus
LEADER 02878nam a2200457Ia 4500
001 10.1093-chromsci-bmz061
008 220121s2019 CNT 000 0 und d
020 |a 00219665 (ISSN) 
245 1 0 |a Effect of Mobile Phase pH on the Electrospray Ionization Efficiency and Qualitative Analysis of Pharmaceuticals in ESI + LC-MS/MS 
260 0 |b Oxford University Press  |c 2019 
650 0 4 |a chemical model 
650 0 4 |a chemistry 
650 0 4 |a Chromatography, Liquid 
650 0 4 |a electrospray mass spectrometry 
650 0 4 |a Hydrogen-Ion Concentration 
650 0 4 |a ion 
650 0 4 |a Ions 
650 0 4 |a limit of detection 
650 0 4 |a Limit of Detection 
650 0 4 |a Linear Models 
650 0 4 |a liquid chromatography 
650 0 4 |a metformin 
650 0 4 |a Metformin 
650 0 4 |a Models, Chemical 
650 0 4 |a pH 
650 0 4 |a procedures 
650 0 4 |a reproducibility 
650 0 4 |a Reproducibility of Results 
650 0 4 |a simvastatin 
650 0 4 |a Simvastatin 
650 0 4 |a Spectrometry, Mass, Electrospray Ionization 
650 0 4 |a statistical model 
650 0 4 |a tandem mass spectrometry 
650 0 4 |a Tandem Mass Spectrometry 
856 |z View Fulltext in Publisher  |u https://doi.org/10.1093/chromsci/bmz061 
856 |z View in Scopus  |u https://www.scopus.com/inward/record.uri?eid=2-s2.0-85076328167&doi=10.1093%2fchromsci%2fbmz061&partnerID=40&md5=1e867fc652c817e608f115b905a9f486 
520 3 |a The effect of mobile phase pH on positive ionization process and retention time of nine pharmaceuticals on ultra-performance liquid chromatography-electrospray-tandem mass spectrometry (LC-MS/MS) was discussed. The effective use of high and low mobile phase pH in LC-MS/MS qualitative analyses method was also evaluated by comparing the instrument detection limit, quantification limit, precision, linearity and signal to noise (S/N) under low and high mobile phase pH. In this work, six mobile phase pH that ranged between pH 2 and pH 10 were used to evaluate the effect of the mobile phase pH changes on the ionization process in electrospray ionization. Results revealed that high mobile phase pH ionized more pharmaceuticals molecules and gave a higher signal than low mobile phase pH in positive ionization mode. The results proved that ammonium ion was better as a proton donor in high pH mobile phase than the hydronium ion in acidic mobile phase. The results revealed that the qualitative LC-MS/MS analyses method by using high mobile phase pH has better performance for most analytes in terms of sensitivity, precision, linearity and S/N than the low mobile phase pH. © 2019 The Author(s). Published by Oxford University Press. All rights reserved. 
700 1 0 |a Bakar, N.K.A.  |e author  
700 1 0 |a Kafeenah, H.I.S.  |e author  
700 1 0 |a Osman, R.  |e author  
773 |t Journal of Chromatographic Science  |x 00219665 (ISSN)  |g 57 9, 847-854