Two T waves following one QRS complex – Unusual treadmill test tracings due to printer malfunction

We observed two T waves following one QRS complex during treadmill stress testing of three patients evaluated on the same machine. Careful analysis revealed that this happened due to transient stopping of the printer of the machine.