Prosthetic Fracture of a Cemented Exeter Femoral Stem (Case Report)

We present a single case of fracture of an Exeter femoral prosthesis at the neck, occurring after a fall from standing height, in a lean 70yr lady at 6 years post implantation. The fracture propagated from the insertion dimple on the superior aspect of the prosthesis shoulder. Materials analysis sug...

Full description

Bibliographic Details
Main Authors: Michael Facek, Yasser Khatib, Eric Swarts
Format: Article
Language:English
Published: Joint Implant Surgery & Research Foundation 2016-12-01
Series:Reconstructive Review
Subjects:
Online Access:http://reconstructivereview.org/ojs/index.php/rr/article/view/159
Description
Summary:We present a single case of fracture of an Exeter femoral prosthesis at the neck, occurring after a fall from standing height, in a lean 70yr lady at 6 years post implantation. The fracture propagated from the insertion dimple on the superior aspect of the prosthesis shoulder. Materials analysis suggested variance in composition of the alloy, particularly with grain size heterogeneity. Whilst Exeter femoral prosthesis fracture is extremely rare, when it does occur the literature suggests it is often in the context of excessive mechanical stresses (obesity, high offset, falls). However, this case may represent a failure of materials rather than mechanical stresses alone.