Prosthetic Fracture of a Cemented Exeter Femoral Stem (Case Report)
We present a single case of fracture of an Exeter femoral prosthesis at the neck, occurring after a fall from standing height, in a lean 70yr lady at 6 years post implantation. The fracture propagated from the insertion dimple on the superior aspect of the prosthesis shoulder. Materials analysis sug...