Predictions of the Compressible Fluid Model and its Comparison to Experimental Measurements of 𝑄 Factors and Flexural Resonance Frequencies for Microcantilevers

The qualitative agreement between experimental measurements of the 𝑄 factors and flexural resonance frequencies in air of microcantilevers and calculations based on the compressible fluid model of Van Eysden and Sader (2009) is presented. The 𝑄 factors and resonance frequencies observed on two sets...

Full description

Bibliographic Details
Main Authors: Jason Jensen, Martin Hegner
Format: Article
Language:English
Published: Hindawi Limited 2012-01-01
Series:Journal of Sensors
Online Access:http://dx.doi.org/10.1155/2012/258381
Description
Summary:The qualitative agreement between experimental measurements of the 𝑄 factors and flexural resonance frequencies in air of microcantilevers and calculations based on the compressible fluid model of Van Eysden and Sader (2009) is presented. The 𝑄 factors and resonance frequencies observed on two sets of cantilever arrays were slightly lower than those predicted by the model. This is attributed to the individual design and geometry of the microfabricated hinged end of the cantilever beams in the array.
ISSN:1687-725X
1687-7268