Dosimetric evaluation of 120-leaf multileaf collimator in a Varian linear accelerator with 6-MV and 18-MV photon beams
In this study the dosimetric characteristics of 120-leaf multileaf collimators (MLCs) were evaluated for 6-MV and 18-MV photon beams. The dose rate, percentage depth dose, surface dose, dose in the build-up region, beam profile, flatness, symmetry, and penumbra width were measured using three field-...
Main Authors: | , , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Wolters Kluwer Medknow Publications
2008-01-01
|
Series: | Journal of Medical Physics |
Subjects: | |
Online Access: | http://www.jmp.org.in/article.asp?issn=0971-6203;year=2008;volume=33;issue=3;spage=114;epage=118;aulast=Mohan |