Clinical impact of applying strategic programming in patients with implantable cardioverter-defibrillators beyond reducing inappropriate shocks

Abstract Background The incidence of inappropriate shocks remains high at 30% in patients with implantable cardioverter-defibrillators (ICDs). This retrospective study sought to examine the efficacy of strategic programming (ICD programming with a long detection interval and high-rate cutoff) in red...

Full description

Bibliographic Details
Main Authors: You Mi Hwang, Jun Kim, Gi Byoung Nam, Kee Joon Choi, You Ho Kim, Seo Young Park
Format: Article
Language:English
Published: BMC 2020-03-01
Series:International Journal of Arrhythmia
Subjects:
Online Access:https://doi.org/10.1186/s42444-020-00012-w