A note on erythrocyte counts in cerebrospinal fluid with the automated hematology analyzer, Technicon H*3
Material and Methods: We tested the hematology analyzer, Technicon H*3, which was not designed for CSF analysis. Twenty CSF samples were analyzed. The counts obtained were then compared with the manual microscopic method. Results: Linearity was tested by normal saline solution (NSS). The results of...