Early fracture of the modular neck of a MODULUS femoral stem
We present the case of a 46-year-old woman who underwent revision surgery approximately 4 years after total hip arthroplasty because of a fracture of the modular neck of a MODULUS femoral stem. The fractured surfaces of the retrieved implant were inspected using optical and scanning electron microsc...
Main Authors: | , , , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Elsevier
2017-06-01
|
Series: | Arthroplasty Today |
Subjects: | |