Double twiddle trouble, a new variant of twiddler syndrome
The twiddler syndrome results in retraction and coiling of the lead in the pacemaker pocket with subsequent pacemaker malfunction. If the lead׳s parameters are not affected, then the problem can be corrected by lead repositioning. We describe the first reported case of a dramatic pacemaker lead coil...