Design and Construction of an Optical Computed Tomography Scanner for Polymer Gel Dosimetry Application

Polymer gel dosimeter is the only accurate three dimensional (3D) dosimeter that can measure the absorbed dose distribution in a perfect 3D setting. Gel dosimetry by using optical computed tomography (OCT) has been promoted by several researches. In the current study, we designed and constructed a p...

Full description

Bibliographic Details
Main Authors: Seyed Salman Zakariaee, Asghar Mesbahi, Ahmad Keshtkar, Vahid Azimirad
Format: Article
Language:English
Published: Wolters Kluwer Medknow Publications 2014-01-01
Series:Journal of Medical Signals and Sensors
Subjects: