Calibration of tri-axial MEMS accelerometers in the low-frequency range – Part 2: Uncertainty assessment
A comparison among three methods for the calibration of tri-axial accelerometers, in particular MEMS, is presented in this paper, paying attention to the uncertainty assessment of each method. The first method is performed according to the ISO 16063 standards. Two innovative methods are analysed...