Evaluation of six scatter correction methods based on spectral analysis in 99m Tc SPECT imaging using SIMIND Monte Carlo simulation

Compton-scattered photons included within the photopeak pulse-height window result in the degradation of SPECT images both qualitatively and quantitatively. The purpose of this study is to evaluate and compare six scatter correction methods based on setting the energy windows in 99m Tc spectrum. SIM...

Full description

Bibliographic Details
Main Authors: Mahsa Noori Asl, Alireza Sadremomtaz, Ahmad Bitarafan-Rajabi
Format: Article
Language:English
Published: Wolters Kluwer Medknow Publications 2013-01-01
Series:Journal of Medical Physics
Subjects:
Online Access:http://www.jmp.org.in/article.asp?issn=0971-6203;year=2013;volume=38;issue=4;spage=189;epage=197;aulast=Asl