Summary: | Ion chromatography coupled with a conductivity detector was used to investigate the analysis of Cr(III) and Cr(VI) in aqueous samples. An IC methodology for Cr(III) was developed using a cation column and an eluent containing tartaric acid, ethylenediamine, and acetonitrile at pH 2.9. The detection limit of this method can reach 0.1 ppm level with good precision. Several operational parameters were evaluated during the regular use of the method. Comparison of the IC method with AA method showed good agreement between the two methods. The anion exchange column was used for Cr(VI) determination. The best results were obtained with an eluent containing sodium gluconate, borate buffer, glycerin, and acetonitrile. The retention time for the Cr207 2 - sample was 11 min. and the calibration curve was linear between 1.0 and 100 ppm.
|