Beam quality specification in kilo-voltage radiotherapy

Purpose: The purpose of this work was to compare and analyse two clinically measurable beam quality specifiers, the half value layer (HVL) and the ratio of the doses at depths 2 cm and 5 cm (D2/D5) for a range of kilovoltage modalities, and to determine whether a practical, alternative and/or bet...

Full description

Bibliographic Details
Main Author: Jozela, Sibusiso
Format: Others
Language:en
Published: 2008
Subjects:
HVL
Online Access:http://hdl.handle.net/10539/4902
id ndltd-netd.ac.za-oai-union.ndltd.org-wits-oai-wiredspace.wits.ac.za-10539-4902
record_format oai_dc
spelling ndltd-netd.ac.za-oai-union.ndltd.org-wits-oai-wiredspace.wits.ac.za-10539-49022021-04-29T05:09:17Z Beam quality specification in kilo-voltage radiotherapy Jozela, Sibusiso HVL D2/D5 dosimetry Purpose: The purpose of this work was to compare and analyse two clinically measurable beam quality specifiers, the half value layer (HVL) and the ratio of the doses at depths 2 cm and 5 cm (D2/D5) for a range of kilovoltage modalities, and to determine whether a practical, alternative and/or better correlation exists. Methods and materials: Four x-ray units were used: two Philips RT 250 units, a Pantak HF 420 operated up to 250 kV, and a D3300 Gulmay Medical unit operated up to 300 kV. As not all these units were equipped with an internal monitor chamber, a system was used where either the first measurement was repeated at the end of each series or an external monitor chamber was employed in order to ensure output constancy. A range of HVL’s were measured on each of the energies investigated on this work, which were used clinically. A calibrated 0.6 cc ionization chamber was used in a 30 cm x 30 cm x 30 cm water phantom to measure the absorbed dose to water at depths 2 cm and 5 cm in order to investigate D2/D5 as the alternative quality index. Results: The effectiveness of using a monitor chamber in the determination of HVL has been shown to be significant in this work where HVLs differed by up to 3%. Errors incurred from using HVL have been identified. This work verified that the ratio of doses at depths 2 cm and 5 cm in water could be applied as a kilovoltage beam quality specifier in the clinical environment at low and medium energies with a well defined FSD and field size. Conclusions: The use of D2/D5 as a tool to verify the beam quality index would simplify quality control in the clinical environment. Further work would have to be done to investigate other energies. Lower energies may require the use of shallower depths in order to improve accuracy and ensure a more clinically relevant setup. 2008-05-29T13:00:02Z 2008-05-29T13:00:02Z 2008-05-29T13:00:02Z Thesis http://hdl.handle.net/10539/4902 en 2567374 bytes 16404 bytes application/pdf application/pdf application/pdf application/pdf
collection NDLTD
language en
format Others
sources NDLTD
topic HVL
D2/D5
dosimetry
spellingShingle HVL
D2/D5
dosimetry
Jozela, Sibusiso
Beam quality specification in kilo-voltage radiotherapy
description Purpose: The purpose of this work was to compare and analyse two clinically measurable beam quality specifiers, the half value layer (HVL) and the ratio of the doses at depths 2 cm and 5 cm (D2/D5) for a range of kilovoltage modalities, and to determine whether a practical, alternative and/or better correlation exists. Methods and materials: Four x-ray units were used: two Philips RT 250 units, a Pantak HF 420 operated up to 250 kV, and a D3300 Gulmay Medical unit operated up to 300 kV. As not all these units were equipped with an internal monitor chamber, a system was used where either the first measurement was repeated at the end of each series or an external monitor chamber was employed in order to ensure output constancy. A range of HVL’s were measured on each of the energies investigated on this work, which were used clinically. A calibrated 0.6 cc ionization chamber was used in a 30 cm x 30 cm x 30 cm water phantom to measure the absorbed dose to water at depths 2 cm and 5 cm in order to investigate D2/D5 as the alternative quality index. Results: The effectiveness of using a monitor chamber in the determination of HVL has been shown to be significant in this work where HVLs differed by up to 3%. Errors incurred from using HVL have been identified. This work verified that the ratio of doses at depths 2 cm and 5 cm in water could be applied as a kilovoltage beam quality specifier in the clinical environment at low and medium energies with a well defined FSD and field size. Conclusions: The use of D2/D5 as a tool to verify the beam quality index would simplify quality control in the clinical environment. Further work would have to be done to investigate other energies. Lower energies may require the use of shallower depths in order to improve accuracy and ensure a more clinically relevant setup.
author Jozela, Sibusiso
author_facet Jozela, Sibusiso
author_sort Jozela, Sibusiso
title Beam quality specification in kilo-voltage radiotherapy
title_short Beam quality specification in kilo-voltage radiotherapy
title_full Beam quality specification in kilo-voltage radiotherapy
title_fullStr Beam quality specification in kilo-voltage radiotherapy
title_full_unstemmed Beam quality specification in kilo-voltage radiotherapy
title_sort beam quality specification in kilo-voltage radiotherapy
publishDate 2008
url http://hdl.handle.net/10539/4902
work_keys_str_mv AT jozelasibusiso beamqualityspecificationinkilovoltageradiotherapy
_version_ 1719400120097701888