Phantom Study Incorporating A Diode Array Into The Treatment Planning System For Patient-Specific Quality Assurance

The purpose of this research is to accurately match the calculation environment, i.e. the treatment planning system (TPS) with the measurement environment (using a 2-D diode array) for lung Stereotactic Body Radiation Therapy (SBRT) patient-specific quality assurance (QA). Furthermore, a new phantom...

Full description

Bibliographic Details
Other Authors: Curley, Casey Michael (author)
Format: Others
Language:English
Published: Florida Atlantic University
Subjects:
Online Access:http://purl.flvc.org/fau/fd/FA00004744
http://purl.flvc.org/fau/fd/FA00004744
id ndltd-fau.edu-oai-fau.digital.flvc.org-fau_33908
record_format oai_dc
spelling ndltd-fau.edu-oai-fau.digital.flvc.org-fau_339082019-07-04T03:52:18Z Phantom Study Incorporating A Diode Array Into The Treatment Planning System For Patient-Specific Quality Assurance FA00004744 Curley, Casey Michael (author) Leventouri, Theodora (Thesis advisor) Ouhib, Zoubir (Thesis advisor) Florida Atlantic University (Degree grantor) Charles E. Schmidt College of Science Department of Physics 171 p. application/pdf Electronic Thesis or Dissertation Text English The purpose of this research is to accurately match the calculation environment, i.e. the treatment planning system (TPS) with the measurement environment (using a 2-D diode array) for lung Stereotactic Body Radiation Therapy (SBRT) patient-specific quality assurance (QA). Furthermore, a new phantom was studied in which the 2-D array and heterogeneities were incorporated into the patient-specific QA process for lung SBRT. Dual source dual energy computerized tomography (DSCT) and single energy computerized tomography (SECT) were used to model phantoms incorporating a 2-D diode array into the TPS. A water-equivalent and a heterogeneous phantom (simulating the thoracic region of a patient) were studied. Monte Carlo and pencil beam dose distributions were compared to the measured distributions. Composite and individual fields were analyzed for normally incident and planned gantry angle deliveries. The distributions were compared using γ-analysis for criteria 3% 3mm, 2% 2mm, and 1% 1mm. The Monte Carlo calculations for the DSCT modeled phantoms (incorporating the array) showed an increase in the passing percentage magnitude for 46.4 % of the fields at 3% 3mm, 85.7% at 2% 2mm, and 92.9% at 1% 1mm. The Monte Carlo calculations gave no agreement for the same γ-analysis criteria using the SECT. Pencil beam calculations resulted in lower passing percentages when the diode array was incorporated in the TPS. The DSCT modeled phantoms (incorporating the array) exhibited decrease in the passing percentage magnitude for 85.7% of the fields at 3% 3mm, 82.1% at 2% 2mm, and 71.4% at 1% 1mm. In SECT modeled phantoms (incorporating the array), a decrease in passing percentage magnitude were found for 92.9% of the fields at 3% 3mm, 89.3% at 2% 2mm, and 82.1% at 1% 1mm. In conclusion, this study demonstrates that including the diode array in the TPS results in increased passing percentages when using a DSCT system with a Monte Carlo algorithm for patient-specific lung SBRT QA. Furthermore, as recommended by task groups (e.g. TG 65, TG 101, TG 244) of the American Association of Physicists in Medicine (AAPM), pencil beam algorithms should be avoided in the presence of heterogeneous materials, including a diode array. Florida Atlantic University Cancer--Radiotherapy. Lungs--Cancer--Treatment. Monte Carlo method. Proton beams. Image-guided radiation therapy. Ion bombardment. Medical physics. Includes bibliography. Dissertation (Ph.D.)--Florida Atlantic University, 2016. FAU Electronic Theses and Dissertations Collection Copyright © is held by the author with permission granted to Florida Atlantic University to digitize, archive and distribute this item for non-profit research and educational purposes. Any reuse of this item in excess of fair use or other copyright exemptions requires permission of the copyright holder. http://purl.flvc.org/fau/fd/FA00004744 http://purl.flvc.org/fau/fd/FA00004744 http://rightsstatements.org/vocab/InC/1.0/ https://fau.digital.flvc.org/islandora/object/fau%3A33908/datastream/TN/view/Phantom%20Study%20Incorporating%20A%20Diode%20Array%20Into%20The%20Treatment%20Planning%20System%20For%20Patient-Specific%20Quality%20Assurance.jpg
collection NDLTD
language English
format Others
sources NDLTD
topic Cancer--Radiotherapy.
Lungs--Cancer--Treatment.
Monte Carlo method.
Proton beams.
Image-guided radiation therapy.
Ion bombardment.
Medical physics.
spellingShingle Cancer--Radiotherapy.
Lungs--Cancer--Treatment.
Monte Carlo method.
Proton beams.
Image-guided radiation therapy.
Ion bombardment.
Medical physics.
Phantom Study Incorporating A Diode Array Into The Treatment Planning System For Patient-Specific Quality Assurance
description The purpose of this research is to accurately match the calculation environment, i.e. the treatment planning system (TPS) with the measurement environment (using a 2-D diode array) for lung Stereotactic Body Radiation Therapy (SBRT) patient-specific quality assurance (QA). Furthermore, a new phantom was studied in which the 2-D array and heterogeneities were incorporated into the patient-specific QA process for lung SBRT. Dual source dual energy computerized tomography (DSCT) and single energy computerized tomography (SECT) were used to model phantoms incorporating a 2-D diode array into the TPS. A water-equivalent and a heterogeneous phantom (simulating the thoracic region of a patient) were studied. Monte Carlo and pencil beam dose distributions were compared to the measured distributions. Composite and individual fields were analyzed for normally incident and planned gantry angle deliveries. The distributions were compared using γ-analysis for criteria 3% 3mm, 2% 2mm, and 1% 1mm. The Monte Carlo calculations for the DSCT modeled phantoms (incorporating the array) showed an increase in the passing percentage magnitude for 46.4 % of the fields at 3% 3mm, 85.7% at 2% 2mm, and 92.9% at 1% 1mm. The Monte Carlo calculations gave no agreement for the same γ-analysis criteria using the SECT. Pencil beam calculations resulted in lower passing percentages when the diode array was incorporated in the TPS. The DSCT modeled phantoms (incorporating the array) exhibited decrease in the passing percentage magnitude for 85.7% of the fields at 3% 3mm, 82.1% at 2% 2mm, and 71.4% at 1% 1mm. In SECT modeled phantoms (incorporating the array), a decrease in passing percentage magnitude were found for 92.9% of the fields at 3% 3mm, 89.3% at 2% 2mm, and 82.1% at 1% 1mm. In conclusion, this study demonstrates that including the diode array in the TPS results in increased passing percentages when using a DSCT system with a Monte Carlo algorithm for patient-specific lung SBRT QA. Furthermore, as recommended by task groups (e.g. TG 65, TG 101, TG 244) of the American Association of Physicists in Medicine (AAPM), pencil beam algorithms should be avoided in the presence of heterogeneous materials, including a diode array. === Includes bibliography. === Dissertation (Ph.D.)--Florida Atlantic University, 2016. === FAU Electronic Theses and Dissertations Collection
author2 Curley, Casey Michael (author)
author_facet Curley, Casey Michael (author)
title Phantom Study Incorporating A Diode Array Into The Treatment Planning System For Patient-Specific Quality Assurance
title_short Phantom Study Incorporating A Diode Array Into The Treatment Planning System For Patient-Specific Quality Assurance
title_full Phantom Study Incorporating A Diode Array Into The Treatment Planning System For Patient-Specific Quality Assurance
title_fullStr Phantom Study Incorporating A Diode Array Into The Treatment Planning System For Patient-Specific Quality Assurance
title_full_unstemmed Phantom Study Incorporating A Diode Array Into The Treatment Planning System For Patient-Specific Quality Assurance
title_sort phantom study incorporating a diode array into the treatment planning system for patient-specific quality assurance
publisher Florida Atlantic University
url http://purl.flvc.org/fau/fd/FA00004744
http://purl.flvc.org/fau/fd/FA00004744
_version_ 1719219098132414464