New method of collecting output factors for commissioming linear accelerators with special emphasis on small fields and intensity modualted readiation therapy

Common methods for commissioning linear accelerators often neglect beam data for small fields. Examining the methods of beam data collection and modeling for commissioning linear accelerators revealed little to no discussion of the protocols for fields smaller than 4 cm x 4 cm. This leads to decreas...

Full description

Bibliographic Details
Other Authors: Smith, Cindy (author)
Format: Others
Language:English
Published: Florida Atlantic University
Subjects:
Online Access:http://purl.flvc.org/fau/fd/FA00004161
id ndltd-fau.edu-oai-fau.digital.flvc.org-fau_13507
record_format oai_dc
spelling ndltd-fau.edu-oai-fau.digital.flvc.org-fau_135072019-07-04T03:56:30Z New method of collecting output factors for commissioming linear accelerators with special emphasis on small fields and intensity modualted readiation therapy FA00004161 Smith, Cindy (author) Pella, Silvia (Thesis advisor) Florida Atlantic University (Degree grantor) Charles E. Schmidt College of Science Department of Physics 109 p. application/pdf Electronic Thesis or Dissertation Text English Common methods for commissioning linear accelerators often neglect beam data for small fields. Examining the methods of beam data collection and modeling for commissioning linear accelerators revealed little to no discussion of the protocols for fields smaller than 4 cm x 4 cm. This leads to decreased confidence levels in the dose calculations and associated monitor units (MUs) for Intensity Modulated Radiation Therapy (IMRT). The parameters of commissioning the Novalis linear accelerator (linac) on the Eclipse Treatment Planning System (TPS) led to the study of challenges collecting data for very small fields. The focus of this thesis is the examination of the protocols for output factor collection and their impact on dose calculations by the TPS for IMRT treatment plans. Improving output factor collection methods, led to significant improvement in absolute dose calculations which correlated with the complexity of the plans. Florida Atlantic University Coherence (Nuclear physics) Linear accelerators in medicine Medical physics Nuclear medicine Particle beams Radiation -- Dosage Radiotherapy -- Positioning Includes bibliography. Thesis (M.S.)--Florida Atlantic University, 2014. FAU Electronic Theses and Dissertations Collection Copyright © is held by the author, with permission granted to Florida Atlantic University to digitize, archive and distribute this item for non-profit research and educational purposes. Any reuse of this item in excess of fair use or other copyright exemptions requires permission of the copyright holder. http://purl.flvc.org/fau/fd/FA00004161 http://rightsstatements.org/vocab/InC/1.0/ https://fau.digital.flvc.org/islandora/object/fau%3A13507/datastream/TN/view/New%20method%20of%20collecting%20output%20factors%20for%20commissioming%20linear%20accelerators%20with%20special%20emphasis%20on%20small%20fields%20and%20intensity%20modualted%20readiation%20therapy.jpg
collection NDLTD
language English
format Others
sources NDLTD
topic Coherence (Nuclear physics)
Linear accelerators in medicine
Medical physics
Nuclear medicine
Particle beams
Radiation -- Dosage
Radiotherapy -- Positioning
spellingShingle Coherence (Nuclear physics)
Linear accelerators in medicine
Medical physics
Nuclear medicine
Particle beams
Radiation -- Dosage
Radiotherapy -- Positioning
New method of collecting output factors for commissioming linear accelerators with special emphasis on small fields and intensity modualted readiation therapy
description Common methods for commissioning linear accelerators often neglect beam data for small fields. Examining the methods of beam data collection and modeling for commissioning linear accelerators revealed little to no discussion of the protocols for fields smaller than 4 cm x 4 cm. This leads to decreased confidence levels in the dose calculations and associated monitor units (MUs) for Intensity Modulated Radiation Therapy (IMRT). The parameters of commissioning the Novalis linear accelerator (linac) on the Eclipse Treatment Planning System (TPS) led to the study of challenges collecting data for very small fields. The focus of this thesis is the examination of the protocols for output factor collection and their impact on dose calculations by the TPS for IMRT treatment plans. Improving output factor collection methods, led to significant improvement in absolute dose calculations which correlated with the complexity of the plans. === Includes bibliography. === Thesis (M.S.)--Florida Atlantic University, 2014. === FAU Electronic Theses and Dissertations Collection
author2 Smith, Cindy (author)
author_facet Smith, Cindy (author)
title New method of collecting output factors for commissioming linear accelerators with special emphasis on small fields and intensity modualted readiation therapy
title_short New method of collecting output factors for commissioming linear accelerators with special emphasis on small fields and intensity modualted readiation therapy
title_full New method of collecting output factors for commissioming linear accelerators with special emphasis on small fields and intensity modualted readiation therapy
title_fullStr New method of collecting output factors for commissioming linear accelerators with special emphasis on small fields and intensity modualted readiation therapy
title_full_unstemmed New method of collecting output factors for commissioming linear accelerators with special emphasis on small fields and intensity modualted readiation therapy
title_sort new method of collecting output factors for commissioming linear accelerators with special emphasis on small fields and intensity modualted readiation therapy
publisher Florida Atlantic University
url http://purl.flvc.org/fau/fd/FA00004161
_version_ 1719219233823391744