The effect of variations in component positioning and swing phase load on the occurrence and severity of edge loading and wear in hip joint replacements

Edge loading is multifactorial and has been identified as a risk leading to the revision of hip joint replacements. Pre-clinical testing requirements currently do not incorporate edge loading. Previous testing of edge loading has been carried out as controlled dynamic microseparation of 0.5 mm via a...

Full description

Bibliographic Details
Main Author: O'Dwyer Lancaster-Jones, Oscar Gilbert
Other Authors: Fisher, John ; Al-Hajjar, Mazen
Published: University of Leeds 2017
Subjects:
621
Online Access:https://ethos.bl.uk/OrderDetails.do?uin=uk.bl.ethos.739771
id ndltd-bl.uk-oai-ethos.bl.uk-739771
record_format oai_dc
spelling ndltd-bl.uk-oai-ethos.bl.uk-7397712019-03-05T15:48:14ZThe effect of variations in component positioning and swing phase load on the occurrence and severity of edge loading and wear in hip joint replacementsO'Dwyer Lancaster-Jones, Oscar GilbertFisher, John ; Al-Hajjar, Mazen2017Edge loading is multifactorial and has been identified as a risk leading to the revision of hip joint replacements. Pre-clinical testing requirements currently do not incorporate edge loading. Previous testing of edge loading has been carried out as controlled dynamic microseparation of 0.5 mm via a spring, where the femoral head contacts the rim section of the acetabular cup. The input conditions of these tests did not allow the individual evaluation of other variables due to the constraints of the methodology. The aim of these studies was to evaluate the effect on the occurrence and severity of edge loading, and wear, under edge loading due to different parameters in a hip joint simulator. A translational mismatch of 1, 2, 3 and 4 mm were applied between the centres of rotation of the head and the cup in order to evaluate the effect on the dynamic separation and severity of edge loading. This allowed differentiating the effect of the 45°, 55° and 65° cup inclination angle studied in the test matrix. Three studies were developed to determine the effect of the cup inclination angle (45°, 55° and 65°), the effect of the swing phase load (50-500 N) and the effect of the spring constant (50, 100 and 200 N/mm), where short biomechanical studies determined the effect and aided to provide an informed decision on the specific test conditions selected for wear testing. The results indicated how the decrease in cup inclination angle and increase in swing phase load decreased the dynamic separation and severity of edge loading and wear. These studies have elaborated on the potential conditions leading to edge loading and their outcome. Overall the severity of edge loading had a positive linear correlation (R2=0.90) with the wear rate for all the conditions tested. This project has emphasised the criteria required to individually evaluate different parameters when considering edge loading and presented a format to develop an ISO standard to consider the risks associated with edge loading.621University of Leedshttps://ethos.bl.uk/OrderDetails.do?uin=uk.bl.ethos.739771http://etheses.whiterose.ac.uk/19659/Electronic Thesis or Dissertation
collection NDLTD
sources NDLTD
topic 621
spellingShingle 621
O'Dwyer Lancaster-Jones, Oscar Gilbert
The effect of variations in component positioning and swing phase load on the occurrence and severity of edge loading and wear in hip joint replacements
description Edge loading is multifactorial and has been identified as a risk leading to the revision of hip joint replacements. Pre-clinical testing requirements currently do not incorporate edge loading. Previous testing of edge loading has been carried out as controlled dynamic microseparation of 0.5 mm via a spring, where the femoral head contacts the rim section of the acetabular cup. The input conditions of these tests did not allow the individual evaluation of other variables due to the constraints of the methodology. The aim of these studies was to evaluate the effect on the occurrence and severity of edge loading, and wear, under edge loading due to different parameters in a hip joint simulator. A translational mismatch of 1, 2, 3 and 4 mm were applied between the centres of rotation of the head and the cup in order to evaluate the effect on the dynamic separation and severity of edge loading. This allowed differentiating the effect of the 45°, 55° and 65° cup inclination angle studied in the test matrix. Three studies were developed to determine the effect of the cup inclination angle (45°, 55° and 65°), the effect of the swing phase load (50-500 N) and the effect of the spring constant (50, 100 and 200 N/mm), where short biomechanical studies determined the effect and aided to provide an informed decision on the specific test conditions selected for wear testing. The results indicated how the decrease in cup inclination angle and increase in swing phase load decreased the dynamic separation and severity of edge loading and wear. These studies have elaborated on the potential conditions leading to edge loading and their outcome. Overall the severity of edge loading had a positive linear correlation (R2=0.90) with the wear rate for all the conditions tested. This project has emphasised the criteria required to individually evaluate different parameters when considering edge loading and presented a format to develop an ISO standard to consider the risks associated with edge loading.
author2 Fisher, John ; Al-Hajjar, Mazen
author_facet Fisher, John ; Al-Hajjar, Mazen
O'Dwyer Lancaster-Jones, Oscar Gilbert
author O'Dwyer Lancaster-Jones, Oscar Gilbert
author_sort O'Dwyer Lancaster-Jones, Oscar Gilbert
title The effect of variations in component positioning and swing phase load on the occurrence and severity of edge loading and wear in hip joint replacements
title_short The effect of variations in component positioning and swing phase load on the occurrence and severity of edge loading and wear in hip joint replacements
title_full The effect of variations in component positioning and swing phase load on the occurrence and severity of edge loading and wear in hip joint replacements
title_fullStr The effect of variations in component positioning and swing phase load on the occurrence and severity of edge loading and wear in hip joint replacements
title_full_unstemmed The effect of variations in component positioning and swing phase load on the occurrence and severity of edge loading and wear in hip joint replacements
title_sort effect of variations in component positioning and swing phase load on the occurrence and severity of edge loading and wear in hip joint replacements
publisher University of Leeds
publishDate 2017
url https://ethos.bl.uk/OrderDetails.do?uin=uk.bl.ethos.739771
work_keys_str_mv AT odwyerlancasterjonesoscargilbert theeffectofvariationsincomponentpositioningandswingphaseloadontheoccurrenceandseverityofedgeloadingandwearinhipjointreplacements
AT odwyerlancasterjonesoscargilbert effectofvariationsincomponentpositioningandswingphaseloadontheoccurrenceandseverityofedgeloadingandwearinhipjointreplacements
_version_ 1718997219219079168