SIMULATIONS OF SCANNED FOCUSSED ULTRASOUND HYPERTHERMIA: THE EFFECTS OF SCANNING SPEED, SCANNING PATTERN AND MULTIPLE TILTED TRANSDUCERS

A transient three-dimensional simulation program was developed to study the effects of scanning speed, scanning pattern, blood perfusion, transducer choice and multiple tilted transducers with overlapping foci during scanned focussed ultrasound hyperthermia. The results showed that (1) the temperatu...

Full description

Bibliographic Details
Main Author: Moros, Eduardo Gerardo, 1960-
Language:en_US
Published: The University of Arizona. 1987
Subjects:
Online Access:http://hdl.handle.net/10150/276440
id ndltd-arizona.edu-oai-arizona.openrepository.com-10150-276440
record_format oai_dc
spelling ndltd-arizona.edu-oai-arizona.openrepository.com-10150-2764402015-10-23T05:01:30Z SIMULATIONS OF SCANNED FOCUSSED ULTRASOUND HYPERTHERMIA: THE EFFECTS OF SCANNING SPEED, SCANNING PATTERN AND MULTIPLE TILTED TRANSDUCERS Moros, Eduardo Gerardo, 1960- Fever therapy. Cancer -- Treatment. Ultrasonic waves -- Therapeutic use. A transient three-dimensional simulation program was developed to study the effects of scanning speed, scanning pattern, blood perfusion, transducer choice and multiple tilted transducers with overlapping foci during scanned focussed ultrasound hyperthermia. The results showed that (1) the temperature fluctuations increase linearly with decreasing scanning speed, (2) the temperature fluctuations are a weak, increasingly exponential function of the blood perfusion rate, and (3) that the largest temperature fluctuation is always located at the acoustical focal depth on the scan path independently of focal plane depth. Simulations using multiple scan paths showed that relatively uniform average temperature distributions can be achieved at the focal zone as long as the spacing between the concentric scans was not greater than the diameter of the focus of the power field. Finally, the results showed that using multiple tilted transducers with overlapping foci, increased focussing can be obtained at the focal depth. 1987 text Thesis-Reproduction (electronic) http://hdl.handle.net/10150/276440 17759068 1330559 .b1637759x en_US Copyright © is held by the author. Digital access to this material is made possible by the University Libraries, University of Arizona. Further transmission, reproduction or presentation (such as public display or performance) of protected items is prohibited except with permission of the author. The University of Arizona.
collection NDLTD
language en_US
sources NDLTD
topic Fever therapy.
Cancer -- Treatment.
Ultrasonic waves -- Therapeutic use.
spellingShingle Fever therapy.
Cancer -- Treatment.
Ultrasonic waves -- Therapeutic use.
Moros, Eduardo Gerardo, 1960-
SIMULATIONS OF SCANNED FOCUSSED ULTRASOUND HYPERTHERMIA: THE EFFECTS OF SCANNING SPEED, SCANNING PATTERN AND MULTIPLE TILTED TRANSDUCERS
description A transient three-dimensional simulation program was developed to study the effects of scanning speed, scanning pattern, blood perfusion, transducer choice and multiple tilted transducers with overlapping foci during scanned focussed ultrasound hyperthermia. The results showed that (1) the temperature fluctuations increase linearly with decreasing scanning speed, (2) the temperature fluctuations are a weak, increasingly exponential function of the blood perfusion rate, and (3) that the largest temperature fluctuation is always located at the acoustical focal depth on the scan path independently of focal plane depth. Simulations using multiple scan paths showed that relatively uniform average temperature distributions can be achieved at the focal zone as long as the spacing between the concentric scans was not greater than the diameter of the focus of the power field. Finally, the results showed that using multiple tilted transducers with overlapping foci, increased focussing can be obtained at the focal depth.
author Moros, Eduardo Gerardo, 1960-
author_facet Moros, Eduardo Gerardo, 1960-
author_sort Moros, Eduardo Gerardo, 1960-
title SIMULATIONS OF SCANNED FOCUSSED ULTRASOUND HYPERTHERMIA: THE EFFECTS OF SCANNING SPEED, SCANNING PATTERN AND MULTIPLE TILTED TRANSDUCERS
title_short SIMULATIONS OF SCANNED FOCUSSED ULTRASOUND HYPERTHERMIA: THE EFFECTS OF SCANNING SPEED, SCANNING PATTERN AND MULTIPLE TILTED TRANSDUCERS
title_full SIMULATIONS OF SCANNED FOCUSSED ULTRASOUND HYPERTHERMIA: THE EFFECTS OF SCANNING SPEED, SCANNING PATTERN AND MULTIPLE TILTED TRANSDUCERS
title_fullStr SIMULATIONS OF SCANNED FOCUSSED ULTRASOUND HYPERTHERMIA: THE EFFECTS OF SCANNING SPEED, SCANNING PATTERN AND MULTIPLE TILTED TRANSDUCERS
title_full_unstemmed SIMULATIONS OF SCANNED FOCUSSED ULTRASOUND HYPERTHERMIA: THE EFFECTS OF SCANNING SPEED, SCANNING PATTERN AND MULTIPLE TILTED TRANSDUCERS
title_sort simulations of scanned focussed ultrasound hyperthermia: the effects of scanning speed, scanning pattern and multiple tilted transducers
publisher The University of Arizona.
publishDate 1987
url http://hdl.handle.net/10150/276440
work_keys_str_mv AT moroseduardogerardo1960 simulationsofscannedfocussedultrasoundhyperthermiatheeffectsofscanningspeedscanningpatternandmultipletiltedtransducers
_version_ 1718102286146732032