SIMULATED ANNEALING AND ESTIMATION THEORY IN CODED-APERTURE IMAGING (RECONSTRUCTION, MONTE CARLO, WIENER FILTER).

Coded-aperture imaging without detector motion can be used to reconstruct three-dimensional radionuclide distributions in the context of nuclear medicine. This approach offers several advantages over the rotating gamma-ray camera systems presently employed in the clinic. These advantages include imp...

Full description

Bibliographic Details
Main Author: SMITH, WARREN ESCHHOLZ.
Other Authors: Barrett, Harry
Language:en
Published: The University of Arizona. 1985
Subjects:
Online Access:http://hdl.handle.net/10150/188135
id ndltd-arizona.edu-oai-arizona.openrepository.com-10150-188135
record_format oai_dc
spelling ndltd-arizona.edu-oai-arizona.openrepository.com-10150-1881352015-11-21T03:00:57Z SIMULATED ANNEALING AND ESTIMATION THEORY IN CODED-APERTURE IMAGING (RECONSTRUCTION, MONTE CARLO, WIENER FILTER). SMITH, WARREN ESCHHOLZ. Barrett, Harry Nuclear medicine -- Equipment and supplies. Imaging systems in medicine. Coded-aperture imaging without detector motion can be used to reconstruct three-dimensional radionuclide distributions in the context of nuclear medicine. This approach offers several advantages over the rotating gamma-ray camera systems presently employed in the clinic. These advantages include improved sensitivity, potentially better spatial resolution, and the capability of doing dynamic studies. There are two problems associated with the coded-aperture approach, however. First, the data is "multiplexed", which refers to the fact that many line integrals of the source distribution are combined together and not measured individually, so that information is lost. Second, the number of resolvable detector elements is typically an order of magnitude less than the number of object elements to be reconstructed, so that the reconstruction problem is underdetermined. Consequently, the reconstruction is not unique. By using various types of a priori information in forming the reconstruction, however, it is possible to augment the incomplete data set. Two algorithms are presented to reconstruct objects from their coded-image projections and various types of a priori information. The first, a Monte Carlo algorithm, is a flexible and computationally efficient approach using the a priori knowledge of positivity and nearest-neighbor correlation. This algorithm is used to qualitatively explore the effect of the data-taking geometry on reconstruction performance. The second algorithm is a linear estimator incorporating as a priori knowledge completely general first- and second-order statistical information about the object class to be reconstructed. The linear-estimator formalism also provides a minimum-variance expression for system optimization. This linear algorithm is used to explore the effects of correct and incorrect a priori information on reconstruction performance, and to quantitatively investigate reconstruction quality with respect to data-taking geometry. 1985 text Dissertation-Reproduction (electronic) http://hdl.handle.net/10150/188135 697290408 8603355 en Copyright © is held by the author. Digital access to this material is made possible by the University Libraries, University of Arizona. Further transmission, reproduction or presentation (such as public display or performance) of protected items is prohibited except with permission of the author. The University of Arizona.
collection NDLTD
language en
sources NDLTD
topic Nuclear medicine -- Equipment and supplies.
Imaging systems in medicine.
spellingShingle Nuclear medicine -- Equipment and supplies.
Imaging systems in medicine.
SMITH, WARREN ESCHHOLZ.
SIMULATED ANNEALING AND ESTIMATION THEORY IN CODED-APERTURE IMAGING (RECONSTRUCTION, MONTE CARLO, WIENER FILTER).
description Coded-aperture imaging without detector motion can be used to reconstruct three-dimensional radionuclide distributions in the context of nuclear medicine. This approach offers several advantages over the rotating gamma-ray camera systems presently employed in the clinic. These advantages include improved sensitivity, potentially better spatial resolution, and the capability of doing dynamic studies. There are two problems associated with the coded-aperture approach, however. First, the data is "multiplexed", which refers to the fact that many line integrals of the source distribution are combined together and not measured individually, so that information is lost. Second, the number of resolvable detector elements is typically an order of magnitude less than the number of object elements to be reconstructed, so that the reconstruction problem is underdetermined. Consequently, the reconstruction is not unique. By using various types of a priori information in forming the reconstruction, however, it is possible to augment the incomplete data set. Two algorithms are presented to reconstruct objects from their coded-image projections and various types of a priori information. The first, a Monte Carlo algorithm, is a flexible and computationally efficient approach using the a priori knowledge of positivity and nearest-neighbor correlation. This algorithm is used to qualitatively explore the effect of the data-taking geometry on reconstruction performance. The second algorithm is a linear estimator incorporating as a priori knowledge completely general first- and second-order statistical information about the object class to be reconstructed. The linear-estimator formalism also provides a minimum-variance expression for system optimization. This linear algorithm is used to explore the effects of correct and incorrect a priori information on reconstruction performance, and to quantitatively investigate reconstruction quality with respect to data-taking geometry.
author2 Barrett, Harry
author_facet Barrett, Harry
SMITH, WARREN ESCHHOLZ.
author SMITH, WARREN ESCHHOLZ.
author_sort SMITH, WARREN ESCHHOLZ.
title SIMULATED ANNEALING AND ESTIMATION THEORY IN CODED-APERTURE IMAGING (RECONSTRUCTION, MONTE CARLO, WIENER FILTER).
title_short SIMULATED ANNEALING AND ESTIMATION THEORY IN CODED-APERTURE IMAGING (RECONSTRUCTION, MONTE CARLO, WIENER FILTER).
title_full SIMULATED ANNEALING AND ESTIMATION THEORY IN CODED-APERTURE IMAGING (RECONSTRUCTION, MONTE CARLO, WIENER FILTER).
title_fullStr SIMULATED ANNEALING AND ESTIMATION THEORY IN CODED-APERTURE IMAGING (RECONSTRUCTION, MONTE CARLO, WIENER FILTER).
title_full_unstemmed SIMULATED ANNEALING AND ESTIMATION THEORY IN CODED-APERTURE IMAGING (RECONSTRUCTION, MONTE CARLO, WIENER FILTER).
title_sort simulated annealing and estimation theory in coded-aperture imaging (reconstruction, monte carlo, wiener filter).
publisher The University of Arizona.
publishDate 1985
url http://hdl.handle.net/10150/188135
work_keys_str_mv AT smithwarreneschholz simulatedannealingandestimationtheoryincodedapertureimagingreconstructionmontecarlowienerfilter
_version_ 1718133537575534592