Reproducibility of a continuous-wave Doppler ultrasound system for assessment of ascending aortic blood flow responses during graded exercise testing with healthy individuals

Continuous-wave (CW) Doppler recordings of ascending aortic maximal blood flow acceleration (PkA), maximal velocity (PkV) and systolic velocity integral (SVI) were taken at each stage of a graded exercise treadmill test on two separate days with 30 physically active adult males. Signals were measure...

Full description

Bibliographic Details
Main Author: Wetherill, Lindsay D.
Other Authors: Health and Physical Education
Format: Others
Language:en_US
Published: Virginia Polytechnic Institute and State University 2017
Subjects:
Online Access:http://hdl.handle.net/10919/80102
id ndltd-VTETD-oai-vtechworks.lib.vt.edu-10919-80102
record_format oai_dc
spelling ndltd-VTETD-oai-vtechworks.lib.vt.edu-10919-801022020-09-29T05:44:05Z Reproducibility of a continuous-wave Doppler ultrasound system for assessment of ascending aortic blood flow responses during graded exercise testing with healthy individuals Wetherill, Lindsay D. Health and Physical Education LD5655.V855 1987.W473 Doppler effect Echocardiography Aortic valve insufficiency Exercise tests Continuous-wave (CW) Doppler recordings of ascending aortic maximal blood flow acceleration (PkA), maximal velocity (PkV) and systolic velocity integral (SVI) were taken at each stage of a graded exercise treadmill test on two separate days with 30 physically active adult males. Signals were measured (Quinton Exerdop) for all cardiac cycles in the 3rd minute of each stage using a hand-held probe positioned at the suprasternal notch. A dedicated microcomputer, programmed to select "valid" beats on the basis of value consistency in the sample set, determined the acceptability of signals. No significant differences were found between the three trial means within each stage on either day for PkA, PkV or SVI. Significant (p <.01) intraclass reliability estimates ranged from r = 0.89 to 0.97 (Pk.A), r = 0.90 to 0.98 (PkV) and r = 0.85 to 0.95 (SVI) . Coefficients of variation were calculated at each stage to estimate the relative consistency of each measure. A gradual reduction of the coefficient of variation was observed for each blood flow measure between stages one and four. The test-retest (between days) reliability coefficients for PkA, PkV and SVI for stages one to four ranged between r = 0.51 to 0.78 (P <.004), but correlations for the pre-exercise baseline and stages five and six were lower. These results indicate that (1) PkA, PkV and SVI demonstrate greater measurement stability within each stage of a graded exercise test than is the case between separate days of measurement at the same stage; and (2) there is modest day-to-day response stability for clinical testing with the Doppler parameter of PkV. Reliability/stability was best in exercise stages which encompass the speed and grade range of 45.0 m•min⁻¹/10%. - 111.7 m•min⁻¹/14%, i.e., those in which all subjects can walk. Master of Science 2017-11-09T20:41:57Z 2017-11-09T20:41:57Z 1987 Thesis Text http://hdl.handle.net/10919/80102 en_US OCLC# 17344736 In Copyright http://rightsstatements.org/vocab/InC/1.0/ xi, 150 leaves application/pdf application/pdf Virginia Polytechnic Institute and State University
collection NDLTD
language en_US
format Others
sources NDLTD
topic LD5655.V855 1987.W473
Doppler effect
Echocardiography
Aortic valve insufficiency
Exercise tests
spellingShingle LD5655.V855 1987.W473
Doppler effect
Echocardiography
Aortic valve insufficiency
Exercise tests
Wetherill, Lindsay D.
Reproducibility of a continuous-wave Doppler ultrasound system for assessment of ascending aortic blood flow responses during graded exercise testing with healthy individuals
description Continuous-wave (CW) Doppler recordings of ascending aortic maximal blood flow acceleration (PkA), maximal velocity (PkV) and systolic velocity integral (SVI) were taken at each stage of a graded exercise treadmill test on two separate days with 30 physically active adult males. Signals were measured (Quinton Exerdop) for all cardiac cycles in the 3rd minute of each stage using a hand-held probe positioned at the suprasternal notch. A dedicated microcomputer, programmed to select "valid" beats on the basis of value consistency in the sample set, determined the acceptability of signals. No significant differences were found between the three trial means within each stage on either day for PkA, PkV or SVI. Significant (p <.01) intraclass reliability estimates ranged from r = 0.89 to 0.97 (Pk.A), r = 0.90 to 0.98 (PkV) and r = 0.85 to 0.95 (SVI) . Coefficients of variation were calculated at each stage to estimate the relative consistency of each measure. A gradual reduction of the coefficient of variation was observed for each blood flow measure between stages one and four. The test-retest (between days) reliability coefficients for PkA, PkV and SVI for stages one to four ranged between r = 0.51 to 0.78 (P <.004), but correlations for the pre-exercise baseline and stages five and six were lower. These results indicate that (1) PkA, PkV and SVI demonstrate greater measurement stability within each stage of a graded exercise test than is the case between separate days of measurement at the same stage; and (2) there is modest day-to-day response stability for clinical testing with the Doppler parameter of PkV. Reliability/stability was best in exercise stages which encompass the speed and grade range of 45.0 m•min⁻¹/10%. - 111.7 m•min⁻¹/14%, i.e., those in which all subjects can walk. === Master of Science
author2 Health and Physical Education
author_facet Health and Physical Education
Wetherill, Lindsay D.
author Wetherill, Lindsay D.
author_sort Wetherill, Lindsay D.
title Reproducibility of a continuous-wave Doppler ultrasound system for assessment of ascending aortic blood flow responses during graded exercise testing with healthy individuals
title_short Reproducibility of a continuous-wave Doppler ultrasound system for assessment of ascending aortic blood flow responses during graded exercise testing with healthy individuals
title_full Reproducibility of a continuous-wave Doppler ultrasound system for assessment of ascending aortic blood flow responses during graded exercise testing with healthy individuals
title_fullStr Reproducibility of a continuous-wave Doppler ultrasound system for assessment of ascending aortic blood flow responses during graded exercise testing with healthy individuals
title_full_unstemmed Reproducibility of a continuous-wave Doppler ultrasound system for assessment of ascending aortic blood flow responses during graded exercise testing with healthy individuals
title_sort reproducibility of a continuous-wave doppler ultrasound system for assessment of ascending aortic blood flow responses during graded exercise testing with healthy individuals
publisher Virginia Polytechnic Institute and State University
publishDate 2017
url http://hdl.handle.net/10919/80102
work_keys_str_mv AT wetherilllindsayd reproducibilityofacontinuouswavedopplerultrasoundsystemforassessmentofascendingaorticbloodflowresponsesduringgradedexercisetestingwithhealthyindividuals
_version_ 1719345880502370304