Evaluation of the ADVIA®60 on highvalue platelets

Platelets are the smallest cells in the blood. They are formed in the bone-marrow and are important for the blood coagulation. Platelet tranfusions are given to patients propyhlactically before an operation but also in therapeutical purpose in connection with bleeding. It’s importent that the qualit...

Full description

Bibliographic Details
Main Author: Ekbom, Lisa H
Format: Others
Language:English
Published: Uppsala universitet, Institutionen för medicinsk biokemi och mikrobiologi 2005
Subjects:
Online Access:http://urn.kb.se/resolve?urn=urn:nbn:se:uu:diva-6149
id ndltd-UPSALLA1-oai-DiVA.org-uu-6149
record_format oai_dc
spelling ndltd-UPSALLA1-oai-DiVA.org-uu-61492013-04-19T20:49:22ZEvaluation of the ADVIA®60 on highvalue plateletsengEkbom, Lisa HUppsala universitet, Institutionen för medicinsk biokemi och mikrobiologi2005Precisionautomated hematology analyzerblood cell counterquality controlimpedanceplatelet count.Cell biologyCellbiologiPlatelets are the smallest cells in the blood. They are formed in the bone-marrow and are important for the blood coagulation. Platelet tranfusions are given to patients propyhlactically before an operation but also in therapeutical purpose in connection with bleeding. It’s importent that the quality controls of the platelet concentrates are reliable. ADVIA®60 (Bayer HealthCare) is a fully automated cell counter which uses impedance principle to count platelets in blood samples. The purpose of the study was to evaluate this new instrument for use in the blood bank of Akademiska Sjukhuset in Uppsala. The instrument was bought to be used for quality control of platelet concentrates. 30 samples from platelet concentrates, from both apheresis and from buffy coats, were analyzed 10 times each on ADVIA®60 and the coefficient of variation (CV) was calculated for each sample. CV variated from 0,8 % to 2,9 % which is good considering that according to Bayer HealthCare the CV should be < 5 % for thrombocytes on ADVIA®60. The instrument was newly calibrated when the study was performed. Platelet count can also be performed by immunological or optical principles. Student thesisinfo:eu-repo/semantics/bachelorThesistexthttp://urn.kb.se/resolve?urn=urn:nbn:se:uu:diva-6149application/pdfinfo:eu-repo/semantics/openAccess
collection NDLTD
language English
format Others
sources NDLTD
topic Precision
automated hematology analyzer
blood cell counter
quality control
impedance
platelet count.
Cell biology
Cellbiologi
spellingShingle Precision
automated hematology analyzer
blood cell counter
quality control
impedance
platelet count.
Cell biology
Cellbiologi
Ekbom, Lisa H
Evaluation of the ADVIA®60 on highvalue platelets
description Platelets are the smallest cells in the blood. They are formed in the bone-marrow and are important for the blood coagulation. Platelet tranfusions are given to patients propyhlactically before an operation but also in therapeutical purpose in connection with bleeding. It’s importent that the quality controls of the platelet concentrates are reliable. ADVIA®60 (Bayer HealthCare) is a fully automated cell counter which uses impedance principle to count platelets in blood samples. The purpose of the study was to evaluate this new instrument for use in the blood bank of Akademiska Sjukhuset in Uppsala. The instrument was bought to be used for quality control of platelet concentrates. 30 samples from platelet concentrates, from both apheresis and from buffy coats, were analyzed 10 times each on ADVIA®60 and the coefficient of variation (CV) was calculated for each sample. CV variated from 0,8 % to 2,9 % which is good considering that according to Bayer HealthCare the CV should be < 5 % for thrombocytes on ADVIA®60. The instrument was newly calibrated when the study was performed. Platelet count can also be performed by immunological or optical principles.
author Ekbom, Lisa H
author_facet Ekbom, Lisa H
author_sort Ekbom, Lisa H
title Evaluation of the ADVIA®60 on highvalue platelets
title_short Evaluation of the ADVIA®60 on highvalue platelets
title_full Evaluation of the ADVIA®60 on highvalue platelets
title_fullStr Evaluation of the ADVIA®60 on highvalue platelets
title_full_unstemmed Evaluation of the ADVIA®60 on highvalue platelets