The Effects of a Carbon Fiber Table Top on Radiation Dose and Image Quality During Fluoroscopy

Fluoroscopic procedures are commonly used in today’s healthcare and involve ionizing radiation exposure to personnel and patients. During these procedures the patient is placed on a surgical table. The board on which the patient is lying on is referred to as the table top, and how different table to...

Full description

Bibliographic Details
Main Authors: Vinasco Korsfeldt, Ludvig, Tais, Mazin
Format: Others
Language:English
Published: KTH, Medicinsk teknik 2014
Online Access:http://urn.kb.se/resolve?urn=urn:nbn:se:kth:diva-172999
id ndltd-UPSALLA1-oai-DiVA.org-kth-172999
record_format oai_dc
spelling ndltd-UPSALLA1-oai-DiVA.org-kth-1729992015-09-23T04:30:39ZThe Effects of a Carbon Fiber Table Top on Radiation Dose and Image Quality During FluoroscopyengETT KOLFIBERBORDS EFFEKTER PÅ STRÅLDOS OCH BILDKVALITET VID FLUOROSKOPIVinasco Korsfeldt, LudvigTais, MazinKTH, Medicinsk teknikKTH, Medicinsk teknik2014Fluoroscopic procedures are commonly used in today’s healthcare and involve ionizing radiation exposure to personnel and patients. During these procedures the patient is placed on a surgical table. The board on which the patient is lying on is referred to as the table top, and how different table top materials affect the image quality and radiation exposure has been investigated in this report. Radiation exposure measurements at different sites have been recorded with both a polymethyl methacrylate (PMMA) and an anthropomorphic phantom representing a patient. Image quality assessment was made in terms of contrast, signal-to-noise ratio and modular transfer function. The result showed that a higher table top attenuation or a higher tube voltage may lead to a lower dose but also a reduction in image quality. The preferred tube voltage and current, and resulting image quality is task dependent, i.e. dependent on the type of clinical procedure, which makes it hard to generalize in the selection of a certain table. Two methods have been proposed when selecting a table, one which can be considered as quantification of the trade-off between image quality and radiation dose, the other one is a fast method for comparing a large amount of tables. Student thesisinfo:eu-repo/semantics/bachelorThesistexthttp://urn.kb.se/resolve?urn=urn:nbn:se:kth:diva-172999TRITA-STH ; 2014:101application/pdfinfo:eu-repo/semantics/openAccess
collection NDLTD
language English
format Others
sources NDLTD
description Fluoroscopic procedures are commonly used in today’s healthcare and involve ionizing radiation exposure to personnel and patients. During these procedures the patient is placed on a surgical table. The board on which the patient is lying on is referred to as the table top, and how different table top materials affect the image quality and radiation exposure has been investigated in this report. Radiation exposure measurements at different sites have been recorded with both a polymethyl methacrylate (PMMA) and an anthropomorphic phantom representing a patient. Image quality assessment was made in terms of contrast, signal-to-noise ratio and modular transfer function. The result showed that a higher table top attenuation or a higher tube voltage may lead to a lower dose but also a reduction in image quality. The preferred tube voltage and current, and resulting image quality is task dependent, i.e. dependent on the type of clinical procedure, which makes it hard to generalize in the selection of a certain table. Two methods have been proposed when selecting a table, one which can be considered as quantification of the trade-off between image quality and radiation dose, the other one is a fast method for comparing a large amount of tables.
author Vinasco Korsfeldt, Ludvig
Tais, Mazin
spellingShingle Vinasco Korsfeldt, Ludvig
Tais, Mazin
The Effects of a Carbon Fiber Table Top on Radiation Dose and Image Quality During Fluoroscopy
author_facet Vinasco Korsfeldt, Ludvig
Tais, Mazin
author_sort Vinasco Korsfeldt, Ludvig
title The Effects of a Carbon Fiber Table Top on Radiation Dose and Image Quality During Fluoroscopy
title_short The Effects of a Carbon Fiber Table Top on Radiation Dose and Image Quality During Fluoroscopy
title_full The Effects of a Carbon Fiber Table Top on Radiation Dose and Image Quality During Fluoroscopy
title_fullStr The Effects of a Carbon Fiber Table Top on Radiation Dose and Image Quality During Fluoroscopy
title_full_unstemmed The Effects of a Carbon Fiber Table Top on Radiation Dose and Image Quality During Fluoroscopy
title_sort effects of a carbon fiber table top on radiation dose and image quality during fluoroscopy
publisher KTH, Medicinsk teknik
publishDate 2014
url http://urn.kb.se/resolve?urn=urn:nbn:se:kth:diva-172999
work_keys_str_mv AT vinascokorsfeldtludvig theeffectsofacarbonfibertabletoponradiationdoseandimagequalityduringfluoroscopy
AT taismazin theeffectsofacarbonfibertabletoponradiationdoseandimagequalityduringfluoroscopy
AT vinascokorsfeldtludvig ettkolfiberbordseffekterpastraldosochbildkvalitetvidfluoroskopi
AT taismazin ettkolfiberbordseffekterpastraldosochbildkvalitetvidfluoroskopi
AT vinascokorsfeldtludvig effectsofacarbonfibertabletoponradiationdoseandimagequalityduringfluoroscopy
AT taismazin effectsofacarbonfibertabletoponradiationdoseandimagequalityduringfluoroscopy
_version_ 1716824732088664065