Performance of a Micro-CT System : Characterisation of Hamamatsu X-ray source L10951-04 and flat panel C7942CA-22
This master thesis evaluated the performance of a micro-CT system consisting of Hamamatsu microfocus X-ray source L10951-04 and CMOS flat panel C7942CA-22. The X-ray source and flat panel have been characterised in terms of dark current, image noise and beam profile. Additionally, the micro-CT syste...
Main Author: | |
---|---|
Format: | Others |
Language: | English |
Published: |
KTH, Skolan för teknik och hälsa (STH)
2014
|
Subjects: | |
Online Access: | http://urn.kb.se/resolve?urn=urn:nbn:se:kth:diva-155454 |
id |
ndltd-UPSALLA1-oai-DiVA.org-kth-155454 |
---|---|
record_format |
oai_dc |
spelling |
ndltd-UPSALLA1-oai-DiVA.org-kth-1554542014-11-06T04:50:44ZPerformance of a Micro-CT System : Characterisation of Hamamatsu X-ray source L10951-04 and flat panel C7942CA-22engPrestanda hos ett Micro-CT System : Karaktärisering av Hamamatsu röntgenkälla L10951-04 och plattpanel C7942CA-22Baumann, MichaelKTH, Skolan för teknik och hälsa (STH)2014micro-CTcomputed tomographyflat panelcharacterisationperformance evaluationHamamatsuimage correctionThis master thesis evaluated the performance of a micro-CT system consisting of Hamamatsu microfocus X-ray source L10951-04 and CMOS flat panel C7942CA-22. The X-ray source and flat panel have been characterised in terms of dark current, image noise and beam profile. Additionally, the micro-CT system’s spatial resolution, detector lag and detector X-ray response have been measured. Guidance for full image correction and methods for characterisation and performance test of the X-ray source and detector is presented. A spatial resolution of 7 lp/mm at 10 % MTF was measured. A detector lag of 0.3 % was observed after ten minutes of radiation exposure. The performance of the micro-CT system was found to be sufficient for high resolution X-ray imaging. However, the detector lag effect is strong enough to reduce image quality during subsequent image acquisition and must either be avoided or corrected for. Student thesisinfo:eu-repo/semantics/bachelorThesistexthttp://urn.kb.se/resolve?urn=urn:nbn:se:kth:diva-155454TRITA-STH ; 2014:87application/pdfinfo:eu-repo/semantics/openAccess |
collection |
NDLTD |
language |
English |
format |
Others
|
sources |
NDLTD |
topic |
micro-CT computed tomography flat panel characterisation performance evaluation Hamamatsu image correction |
spellingShingle |
micro-CT computed tomography flat panel characterisation performance evaluation Hamamatsu image correction Baumann, Michael Performance of a Micro-CT System : Characterisation of Hamamatsu X-ray source L10951-04 and flat panel C7942CA-22 |
description |
This master thesis evaluated the performance of a micro-CT system consisting of Hamamatsu microfocus X-ray source L10951-04 and CMOS flat panel C7942CA-22. The X-ray source and flat panel have been characterised in terms of dark current, image noise and beam profile. Additionally, the micro-CT system’s spatial resolution, detector lag and detector X-ray response have been measured. Guidance for full image correction and methods for characterisation and performance test of the X-ray source and detector is presented. A spatial resolution of 7 lp/mm at 10 % MTF was measured. A detector lag of 0.3 % was observed after ten minutes of radiation exposure. The performance of the micro-CT system was found to be sufficient for high resolution X-ray imaging. However, the detector lag effect is strong enough to reduce image quality during subsequent image acquisition and must either be avoided or corrected for. |
author |
Baumann, Michael |
author_facet |
Baumann, Michael |
author_sort |
Baumann, Michael |
title |
Performance of a Micro-CT System : Characterisation of Hamamatsu X-ray source L10951-04 and flat panel C7942CA-22 |
title_short |
Performance of a Micro-CT System : Characterisation of Hamamatsu X-ray source L10951-04 and flat panel C7942CA-22 |
title_full |
Performance of a Micro-CT System : Characterisation of Hamamatsu X-ray source L10951-04 and flat panel C7942CA-22 |
title_fullStr |
Performance of a Micro-CT System : Characterisation of Hamamatsu X-ray source L10951-04 and flat panel C7942CA-22 |
title_full_unstemmed |
Performance of a Micro-CT System : Characterisation of Hamamatsu X-ray source L10951-04 and flat panel C7942CA-22 |
title_sort |
performance of a micro-ct system : characterisation of hamamatsu x-ray source l10951-04 and flat panel c7942ca-22 |
publisher |
KTH, Skolan för teknik och hälsa (STH) |
publishDate |
2014 |
url |
http://urn.kb.se/resolve?urn=urn:nbn:se:kth:diva-155454 |
work_keys_str_mv |
AT baumannmichael performanceofamicroctsystemcharacterisationofhamamatsuxraysourcel1095104andflatpanelc7942ca22 AT baumannmichael prestandahosettmicroctsystemkaraktariseringavhamamatsurontgenkallal1095104ochplattpanelc7942ca22 |
_version_ |
1716719828923842560 |