Performance of a Micro-CT System : Characterisation of Hamamatsu X-ray source L10951-04 and flat panel C7942CA-22

This master thesis evaluated the performance of a micro-CT system consisting of Hamamatsu microfocus X-ray source L10951-04 and CMOS flat panel C7942CA-22. The X-ray source and flat panel have been characterised in terms of dark current, image noise and beam profile. Additionally, the micro-CT syste...

Full description

Bibliographic Details
Main Author: Baumann, Michael
Format: Others
Language:English
Published: KTH, Skolan för teknik och hälsa (STH) 2014
Subjects:
Online Access:http://urn.kb.se/resolve?urn=urn:nbn:se:kth:diva-155454
id ndltd-UPSALLA1-oai-DiVA.org-kth-155454
record_format oai_dc
spelling ndltd-UPSALLA1-oai-DiVA.org-kth-1554542014-11-06T04:50:44ZPerformance of a Micro-CT System : Characterisation of Hamamatsu X-ray source L10951-04 and flat panel C7942CA-22engPrestanda hos ett Micro-CT System : Karaktärisering av Hamamatsu röntgenkälla L10951-04 och plattpanel C7942CA-22Baumann, MichaelKTH, Skolan för teknik och hälsa (STH)2014micro-CTcomputed tomographyflat panelcharacterisationperformance evaluationHamamatsuimage correctionThis master thesis evaluated the performance of a micro-CT system consisting of Hamamatsu microfocus X-ray source L10951-04 and CMOS flat panel C7942CA-22. The X-ray source and flat panel have been characterised in terms of dark current, image noise and beam profile. Additionally, the micro-CT system’s spatial resolution, detector lag and detector X-ray response have been measured. Guidance for full image correction and methods for characterisation and performance test of the X-ray source and detector is presented. A spatial resolution of 7 lp/mm at 10 % MTF was measured. A detector lag of 0.3 % was observed after ten minutes of radiation exposure. The performance of the micro-CT system was found to be sufficient for high resolution X-ray imaging. However, the detector lag effect is strong enough to reduce image quality during subsequent image acquisition and must either be avoided or corrected for. Student thesisinfo:eu-repo/semantics/bachelorThesistexthttp://urn.kb.se/resolve?urn=urn:nbn:se:kth:diva-155454TRITA-STH ; 2014:87application/pdfinfo:eu-repo/semantics/openAccess
collection NDLTD
language English
format Others
sources NDLTD
topic micro-CT
computed tomography
flat panel
characterisation
performance evaluation
Hamamatsu
image correction
spellingShingle micro-CT
computed tomography
flat panel
characterisation
performance evaluation
Hamamatsu
image correction
Baumann, Michael
Performance of a Micro-CT System : Characterisation of Hamamatsu X-ray source L10951-04 and flat panel C7942CA-22
description This master thesis evaluated the performance of a micro-CT system consisting of Hamamatsu microfocus X-ray source L10951-04 and CMOS flat panel C7942CA-22. The X-ray source and flat panel have been characterised in terms of dark current, image noise and beam profile. Additionally, the micro-CT system’s spatial resolution, detector lag and detector X-ray response have been measured. Guidance for full image correction and methods for characterisation and performance test of the X-ray source and detector is presented. A spatial resolution of 7 lp/mm at 10 % MTF was measured. A detector lag of 0.3 % was observed after ten minutes of radiation exposure. The performance of the micro-CT system was found to be sufficient for high resolution X-ray imaging. However, the detector lag effect is strong enough to reduce image quality during subsequent image acquisition and must either be avoided or corrected for.
author Baumann, Michael
author_facet Baumann, Michael
author_sort Baumann, Michael
title Performance of a Micro-CT System : Characterisation of Hamamatsu X-ray source L10951-04 and flat panel C7942CA-22
title_short Performance of a Micro-CT System : Characterisation of Hamamatsu X-ray source L10951-04 and flat panel C7942CA-22
title_full Performance of a Micro-CT System : Characterisation of Hamamatsu X-ray source L10951-04 and flat panel C7942CA-22
title_fullStr Performance of a Micro-CT System : Characterisation of Hamamatsu X-ray source L10951-04 and flat panel C7942CA-22
title_full_unstemmed Performance of a Micro-CT System : Characterisation of Hamamatsu X-ray source L10951-04 and flat panel C7942CA-22
title_sort performance of a micro-ct system : characterisation of hamamatsu x-ray source l10951-04 and flat panel c7942ca-22
publisher KTH, Skolan för teknik och hälsa (STH)
publishDate 2014
url http://urn.kb.se/resolve?urn=urn:nbn:se:kth:diva-155454
work_keys_str_mv AT baumannmichael performanceofamicroctsystemcharacterisationofhamamatsuxraysourcel1095104andflatpanelc7942ca22
AT baumannmichael prestandahosettmicroctsystemkaraktariseringavhamamatsurontgenkallal1095104ochplattpanelc7942ca22
_version_ 1716719828923842560