Development of Optical Resolution Photoacoustic Microscope for Micro-vasculature Imaging
碩士 === 國立清華大學 === 電機工程學系 === 102 === Blood vessels play an important role in many significant disease researches such as cancer study. Photoacoustic imaging is a novel bio-imaging modality based on the photoacoustic effect. For micro-vasculature imaging, it owns the advantages of label free high opt...
Main Authors: | , |
---|---|
Other Authors: | |
Format: | Others |
Language: | zh-TW |
Published: |
2014
|
Online Access: | http://ndltd.ncl.edu.tw/handle/4fa93j |
Summary: | 碩士 === 國立清華大學 === 電機工程學系 === 102 === Blood vessels play an important role in many significant disease researches such as cancer study. Photoacoustic imaging is a novel bio-imaging modality based on the photoacoustic effect. For micro-vasculature imaging, it owns the advantages of label free high optical absorption contrast and can be performed non-invasively. It also
can provide blood-related functional imaging capability for the measure of total hemoglobin concentration and hemoglobin oxygen saturation. In this thesis, we developed an unfocused ultrasound transducer based laser
scanning optical resolution photoacoustic microscope (OR-PAM) for
extended large field of view (FOV) in vivo micro-vasculature imaging of
small animals. Conventional OR-PAM employs a focused ultrasound
transducer to improve the signal-to-noise ratio and performs mechanical
scanning for imaging. However, mechanical scanning is time-consuming.
Such a problem is solved by optical scanning in this study while the
optical objective lens and focused ultrasound transducer limit the FOV
instead. In our design, the FOV is improved by using an unfocused
ultrasound transducer plus laser scanning. The experimental results
showed that the developed OR-PAM is with axial resolution of 120m
and lateral resolution of 4m when using a 10-MHz unfocused
transducer. The achievable FOV is at least 2x2 mm2. The resolving
power of the system was also demonstrated by imaging the in vivo
micro-vasculature of a mouse ear. The estimated noise-equivalent
penetration depth is 0.8 mm in vivo. Future work will focus on the
improvement of the imaging frame rate and spatial resolution and the
development of multi-wavelength functional micro-vascular imaging.
|
---|