Improvement and Performance Testing of Bioaerosol Generation & Sampling

碩士 === 中山醫學大學 === 職業安全衛生學系碩士班 === 102 === Andersen N6 single-stage bio-aerosol sampler is an impact sampler. In the process of collecting the sample, many microorganisms may die because of the impact. If the density of the sample is too high, there may be the masking effect. Therefore, the purpose...

Full description

Bibliographic Details
Main Authors: Wei-Ting Lin, 林威廷
Other Authors: 賴全裕
Format: Others
Language:zh-TW
Published: 2014
Online Access:http://ndltd.ncl.edu.tw/handle/65048471098355593316
id ndltd-TW-102CSMU5590009
record_format oai_dc
spelling ndltd-TW-102CSMU55900092016-02-21T04:33:01Z http://ndltd.ncl.edu.tw/handle/65048471098355593316 Improvement and Performance Testing of Bioaerosol Generation & Sampling 生物氣膠產生及採樣改良評估 Wei-Ting Lin 林威廷 碩士 中山醫學大學 職業安全衛生學系碩士班 102 Andersen N6 single-stage bio-aerosol sampler is an impact sampler. In the process of collecting the sample, many microorganisms may die because of the impact. If the density of the sample is too high, there may be the masking effect. Therefore, the purpose of this study is to develop a bio-aerosol generator, to improve the efficiency of the Andersen single-stage sampler, and to develop the dilutor for the sampling and studies its efficiency. This study used ultrasonic orifice atomizer to generate its mist/fog for its bio-aerosol generator. It studied the outcomes of different power on four different sizes of orifice of atomizer plates. This study further modified the Andersen single-stage sampler by using a 15cm culture medium, and diluted the dilutor ten times. Results showed that atomizer plate with 7μm set at 10V, 0.22A could have a more stable and higher concentration of bio-aerosols than Collison three-jet nebulizer at 15psi. However, the most efficient nebulization on the four different atomizer plates was not with the maximum power but the second highest power. By using 15cm culture medium on the Andersen single-stage bio-aerosol sampler in the collecting of the sample, the efficiency could increase as much as 24.2% and in addition there were three different kinds of bacteria found,--Arcanobacteriumpyogenes, Corynebacteriumamycolatum, and Enterococcus hirae. In the sampling, Staphylococcus species could increase 42-69% in density but Micrococcus species decreased its density to about 1/4. On the sampling of fungi, the density of Penicilliumcitrinum increased 3 times and the density of Cladosporiumoxysporum increased 1.6 times. There was about 22% loss of bio-aerosols because of vessel wall. 賴全裕 2014 學位論文 ; thesis 96 zh-TW
collection NDLTD
language zh-TW
format Others
sources NDLTD
description 碩士 === 中山醫學大學 === 職業安全衛生學系碩士班 === 102 === Andersen N6 single-stage bio-aerosol sampler is an impact sampler. In the process of collecting the sample, many microorganisms may die because of the impact. If the density of the sample is too high, there may be the masking effect. Therefore, the purpose of this study is to develop a bio-aerosol generator, to improve the efficiency of the Andersen single-stage sampler, and to develop the dilutor for the sampling and studies its efficiency. This study used ultrasonic orifice atomizer to generate its mist/fog for its bio-aerosol generator. It studied the outcomes of different power on four different sizes of orifice of atomizer plates. This study further modified the Andersen single-stage sampler by using a 15cm culture medium, and diluted the dilutor ten times. Results showed that atomizer plate with 7μm set at 10V, 0.22A could have a more stable and higher concentration of bio-aerosols than Collison three-jet nebulizer at 15psi. However, the most efficient nebulization on the four different atomizer plates was not with the maximum power but the second highest power. By using 15cm culture medium on the Andersen single-stage bio-aerosol sampler in the collecting of the sample, the efficiency could increase as much as 24.2% and in addition there were three different kinds of bacteria found,--Arcanobacteriumpyogenes, Corynebacteriumamycolatum, and Enterococcus hirae. In the sampling, Staphylococcus species could increase 42-69% in density but Micrococcus species decreased its density to about 1/4. On the sampling of fungi, the density of Penicilliumcitrinum increased 3 times and the density of Cladosporiumoxysporum increased 1.6 times. There was about 22% loss of bio-aerosols because of vessel wall.
author2 賴全裕
author_facet 賴全裕
Wei-Ting Lin
林威廷
author Wei-Ting Lin
林威廷
spellingShingle Wei-Ting Lin
林威廷
Improvement and Performance Testing of Bioaerosol Generation & Sampling
author_sort Wei-Ting Lin
title Improvement and Performance Testing of Bioaerosol Generation & Sampling
title_short Improvement and Performance Testing of Bioaerosol Generation & Sampling
title_full Improvement and Performance Testing of Bioaerosol Generation & Sampling
title_fullStr Improvement and Performance Testing of Bioaerosol Generation & Sampling
title_full_unstemmed Improvement and Performance Testing of Bioaerosol Generation & Sampling
title_sort improvement and performance testing of bioaerosol generation & sampling
publishDate 2014
url http://ndltd.ncl.edu.tw/handle/65048471098355593316
work_keys_str_mv AT weitinglin improvementandperformancetestingofbioaerosolgenerationsampling
AT línwēitíng improvementandperformancetestingofbioaerosolgenerationsampling
AT weitinglin shēngwùqìjiāochǎnshēngjícǎiyànggǎiliángpínggū
AT línwēitíng shēngwùqìjiāochǎnshēngjícǎiyànggǎiliángpínggū
_version_ 1718193986479325184