Development of the Lead Zirconate Titanate (PZT) High Frequency Ultrasound Needle Transducer and Clinical Verification
碩士 === 國立陽明大學 === 醫學工程研究所 === 101 === Epidural anesthesia is a highly experience dependent procedure. It is a challenge to properly position the epidural needle into the epidural space without dural puncture which damages the dura maters. Avoiding dural puncture, we develop a ultrasound needle trans...
Main Authors: | , |
---|---|
Other Authors: | |
Format: | Others |
Language: | zh-TW |
Published: |
2013
|
Online Access: | http://ndltd.ncl.edu.tw/handle/17221996773668723346 |
id |
ndltd-TW-101YM005530012 |
---|---|
record_format |
oai_dc |
spelling |
ndltd-TW-101YM0055300122016-03-18T04:41:51Z http://ndltd.ncl.edu.tw/handle/17221996773668723346 Development of the Lead Zirconate Titanate (PZT) High Frequency Ultrasound Needle Transducer and Clinical Verification 鋯鈦酸鉛高頻超音波探針製作及臨床驗證 Chiao-Ling Hsieh 謝巧羚 碩士 國立陽明大學 醫學工程研究所 101 Epidural anesthesia is a highly experience dependent procedure. It is a challenge to properly position the epidural needle into the epidural space without dural puncture which damages the dura maters. Avoiding dural puncture, we develop a ultrasound needle transducer epidural positioning system. We use lead zirconate titanate (PZT-5H) as active piezoelectric crystal, which is the most mature material in fabricating ultrasound transducer in Taiwan. In this study, a high-frequency single-element needle transducer with an aperture size of 0.5×0.5 mm2, needle length is 14 cm, and the center frequency is 30 MHz. After fabrication the measured -6 dB center frequency and bandwidth were 27.4 MHz ~ 36 MHz and 14% ~ 59.1%. In clinical animal experiments acquired the A-mode signals by needle ultrasound transducer. Than use these signals to identify the needle’s position inside the body. At present, we’ve successfully measured the signals of Dura Mater. In this study we provide the design and fabrication’s procedure of lead zirconate titanate high frequency ultrasound transducer. We’ve successfully verification that the ultrasound transducer is able to identify the position of epidural space. In the future, we will try to use this system in other surgery’s position. Huihua Kenny Chiang 江惠華 2013 學位論文 ; thesis 69 zh-TW |
collection |
NDLTD |
language |
zh-TW |
format |
Others
|
sources |
NDLTD |
description |
碩士 === 國立陽明大學 === 醫學工程研究所 === 101 === Epidural anesthesia is a highly experience dependent procedure. It is a challenge to properly position the epidural needle into the epidural space without dural puncture which damages the dura maters. Avoiding dural puncture, we develop a ultrasound needle transducer epidural positioning system.
We use lead zirconate titanate (PZT-5H) as active piezoelectric crystal, which is the most mature material in fabricating ultrasound transducer in Taiwan. In this study, a high-frequency single-element needle transducer with an aperture size of 0.5×0.5 mm2, needle length is 14 cm, and the center frequency is 30 MHz. After fabrication the measured -6 dB center frequency and bandwidth were 27.4 MHz ~ 36 MHz and 14% ~ 59.1%.
In clinical animal experiments acquired the A-mode signals by needle ultrasound transducer. Than use these signals to identify the needle’s position inside the body. At present, we’ve successfully measured the signals of Dura Mater.
In this study we provide the design and fabrication’s procedure of lead zirconate titanate high frequency ultrasound transducer. We’ve successfully verification that the ultrasound transducer is able to identify the position of epidural space. In the future, we will try to use this system in other surgery’s position.
|
author2 |
Huihua Kenny Chiang |
author_facet |
Huihua Kenny Chiang Chiao-Ling Hsieh 謝巧羚 |
author |
Chiao-Ling Hsieh 謝巧羚 |
spellingShingle |
Chiao-Ling Hsieh 謝巧羚 Development of the Lead Zirconate Titanate (PZT) High Frequency Ultrasound Needle Transducer and Clinical Verification |
author_sort |
Chiao-Ling Hsieh |
title |
Development of the Lead Zirconate Titanate (PZT) High Frequency Ultrasound Needle Transducer and Clinical Verification |
title_short |
Development of the Lead Zirconate Titanate (PZT) High Frequency Ultrasound Needle Transducer and Clinical Verification |
title_full |
Development of the Lead Zirconate Titanate (PZT) High Frequency Ultrasound Needle Transducer and Clinical Verification |
title_fullStr |
Development of the Lead Zirconate Titanate (PZT) High Frequency Ultrasound Needle Transducer and Clinical Verification |
title_full_unstemmed |
Development of the Lead Zirconate Titanate (PZT) High Frequency Ultrasound Needle Transducer and Clinical Verification |
title_sort |
development of the lead zirconate titanate (pzt) high frequency ultrasound needle transducer and clinical verification |
publishDate |
2013 |
url |
http://ndltd.ncl.edu.tw/handle/17221996773668723346 |
work_keys_str_mv |
AT chiaolinghsieh developmentoftheleadzirconatetitanatepzthighfrequencyultrasoundneedletransducerandclinicalverification AT xièqiǎolíng developmentoftheleadzirconatetitanatepzthighfrequencyultrasoundneedletransducerandclinicalverification AT chiaolinghsieh gàotàisuānqiāngāopínchāoyīnbōtànzhēnzhìzuòjílínchuángyànzhèng AT xièqiǎolíng gàotàisuānqiāngāopínchāoyīnbōtànzhēnzhìzuòjílínchuángyànzhèng |
_version_ |
1718207787155062784 |