A study of portable HPGe detector on method of calibration for BWR recirculation pipe radiation build-up measurement
碩士 === 國立清華大學 === 工程與系統科學系 === 101 === Using experimental and simulation method we found a general-purpose simulation software MCNP can be used in our special-configuration calibration of efficiency with the same type fitting curve on the measurement of BWR recirculation pipe build-up with the porta...
Main Authors: | , |
---|---|
Other Authors: | |
Format: | Others |
Language: | zh-TW |
Published: |
2013
|
Online Access: | http://ndltd.ncl.edu.tw/handle/81517570909946854383 |
Summary: | 碩士 === 國立清華大學 === 工程與系統科學系 === 101 === Using experimental and simulation method we found a general-purpose simulation software MCNP can be used in our special-configuration calibration of efficiency with the same type fitting curve on the measurement of BWR recirculation pipe build-up with the portable HPGe detector during the outage period. The MCNP simulation deviation can be controled under 10% while ISOCS application here isn’t fitted and only the simple type of sources and shielding can reach 5% deviation yet other complicated ones could be as large as around 60% based on our experiment.
|
---|