A multi-function driving circuit design with feedback control for phototherapy and electrical stimulation system

碩士 === 國立暨南國際大學 === 生物醫學科技研究所 === 96 === This thesis presents a multi-function driving circuit design with feedback control for phototherapy and electrical stimulation system. It is an all-in-one therapeutic instrument. It includes LCD component, switch circuit and constant current driving circuit....

Full description

Bibliographic Details
Main Authors: Yuan Ho, 賀淵
Other Authors: Tai-Ping Sun
Format: Others
Language:zh-TW
Published: 2008
Online Access:http://ndltd.ncl.edu.tw/handle/40125179523995416982
id ndltd-TW-096NCNU0114003
record_format oai_dc
spelling ndltd-TW-096NCNU01140032015-10-13T13:47:50Z http://ndltd.ncl.edu.tw/handle/40125179523995416982 A multi-function driving circuit design with feedback control for phototherapy and electrical stimulation system 多功能具迴授控制驅動電路設計應用於光療與電療治療儀系統 Yuan Ho 賀淵 碩士 國立暨南國際大學 生物醫學科技研究所 96 This thesis presents a multi-function driving circuit design with feedback control for phototherapy and electrical stimulation system. It is an all-in-one therapeutic instrument. It includes LCD component, switch circuit and constant current driving circuit. It has been designed by PIC to make up a multi-functional therapy system with light and electrical stimulation for medical application. This text will recommend the treatment of the light and electrical stimulation( micro current electrical stimulation ) to develop at first and overall system structure and theory secondly. Then, we have explained the sub circuit in the system structure, for instance: the driving circuit design and how to reject the offset voltage of op Amp. The phototherapy which is red, blue and green light has different power (under the frequency= 100Hz, duty cycle= 6% and current= 100mA) 12.34mW/cm2, 11.92 mW/cm2, 7 mW/cm2. separately. In this condition(frequency= 100Hz, duty cycle= 6% and current= 100mA ), we can find the better efficiency. After calculation, it can reach the record in the other paper. Micro-current electrical stimulation works at different constant current under frequency= 100Hz and we can find the better linearity at various loading resistance. Tai-Ping Sun 孫台平 2008 學位論文 ; thesis 67 zh-TW
collection NDLTD
language zh-TW
format Others
sources NDLTD
description 碩士 === 國立暨南國際大學 === 生物醫學科技研究所 === 96 === This thesis presents a multi-function driving circuit design with feedback control for phototherapy and electrical stimulation system. It is an all-in-one therapeutic instrument. It includes LCD component, switch circuit and constant current driving circuit. It has been designed by PIC to make up a multi-functional therapy system with light and electrical stimulation for medical application. This text will recommend the treatment of the light and electrical stimulation( micro current electrical stimulation ) to develop at first and overall system structure and theory secondly. Then, we have explained the sub circuit in the system structure, for instance: the driving circuit design and how to reject the offset voltage of op Amp. The phototherapy which is red, blue and green light has different power (under the frequency= 100Hz, duty cycle= 6% and current= 100mA) 12.34mW/cm2, 11.92 mW/cm2, 7 mW/cm2. separately. In this condition(frequency= 100Hz, duty cycle= 6% and current= 100mA ), we can find the better efficiency. After calculation, it can reach the record in the other paper. Micro-current electrical stimulation works at different constant current under frequency= 100Hz and we can find the better linearity at various loading resistance.
author2 Tai-Ping Sun
author_facet Tai-Ping Sun
Yuan Ho
賀淵
author Yuan Ho
賀淵
spellingShingle Yuan Ho
賀淵
A multi-function driving circuit design with feedback control for phototherapy and electrical stimulation system
author_sort Yuan Ho
title A multi-function driving circuit design with feedback control for phototherapy and electrical stimulation system
title_short A multi-function driving circuit design with feedback control for phototherapy and electrical stimulation system
title_full A multi-function driving circuit design with feedback control for phototherapy and electrical stimulation system
title_fullStr A multi-function driving circuit design with feedback control for phototherapy and electrical stimulation system
title_full_unstemmed A multi-function driving circuit design with feedback control for phototherapy and electrical stimulation system
title_sort multi-function driving circuit design with feedback control for phototherapy and electrical stimulation system
publishDate 2008
url http://ndltd.ncl.edu.tw/handle/40125179523995416982
work_keys_str_mv AT yuanho amultifunctiondrivingcircuitdesignwithfeedbackcontrolforphototherapyandelectricalstimulationsystem
AT hèyuān amultifunctiondrivingcircuitdesignwithfeedbackcontrolforphototherapyandelectricalstimulationsystem
AT yuanho duōgōngnéngjùhuíshòukòngzhìqūdòngdiànlùshèjìyīngyòngyúguāngliáoyǔdiànliáozhìliáoyíxìtǒng
AT hèyuān duōgōngnéngjùhuíshòukòngzhìqūdòngdiànlùshèjìyīngyòngyúguāngliáoyǔdiànliáozhìliáoyíxìtǒng
AT yuanho multifunctiondrivingcircuitdesignwithfeedbackcontrolforphototherapyandelectricalstimulationsystem
AT hèyuān multifunctiondrivingcircuitdesignwithfeedbackcontrolforphototherapyandelectricalstimulationsystem
_version_ 1717742605789298688