A Study on the Determination of Arsenic(III) and Arsenic(V) in Drinking Water and Arsenic(III) in Air by Graphite-Furnace Atomic Absorption Spectrophotometry

碩士 === 東海大學 === 環境科學系 === 86 === Arsenic(III) (12.5-200 ng) in drinking water (25 mL) reacted with 2,3-dimercaptopropane-1-sulfonate (DMPS,0.6 mg) in ammonium acetatebuffer (2.0 mmol, pH=5.5) and formed a complex of As(III)-DMPS. Thecomplex was concentrated on two Sep-Pak C18 cartridges in series an...

Full description

Bibliographic Details
Main Authors: Hsieh,Chuan-Jier, 謝傳傑
Other Authors: 郭茂松
Format: Others
Language:zh-TW
Published: 1998
Online Access:http://ndltd.ncl.edu.tw/handle/65425009911038954192
id ndltd-TW-086THU00518007
record_format oai_dc
spelling ndltd-TW-086THU005180072015-10-13T17:34:44Z http://ndltd.ncl.edu.tw/handle/65425009911038954192 A Study on the Determination of Arsenic(III) and Arsenic(V) in Drinking Water and Arsenic(III) in Air by Graphite-Furnace Atomic Absorption Spectrophotometry 以石墨電熱式原子吸光法測定飲用水中As(III),As(V)和空氣中As(III)之研究 Hsieh,Chuan-Jier 謝傳傑 碩士 東海大學 環境科學系 86 Arsenic(III) (12.5-200 ng) in drinking water (25 mL) reacted with 2,3-dimercaptopropane-1-sulfonate (DMPS,0.6 mg) in ammonium acetatebuffer (2.0 mmol, pH=5.5) and formed a complex of As(III)-DMPS. Thecomplex was concentrated on two Sep-Pak C18 cartridges in series and each cartridge was eluted with methanol and adjusted to 2.00 mL. After additionof 2.0 mg of Ni(II), a portion(20 mL) of the mixture was introduced into a graphite cuvette and was atomized according to a temperature program.The method detection limit (3s) was 0.11*0.01(mg/L) and the calibration graph was linear up to 8.00(mg/L) . Accuracies of 97.6-98.9% wereobtained when testing a QC (Strem Chem. Inc.) sample. Arsenic(V) (up to 100 ng) in drinking water was reduced to As(III) with L-cysteine(0.5 mL, 1.0% in HNO3 (0.2%,v/v)) prior to the formation of As(III)-DMPS complex. Accuracies of 98.5-99.3% were obtained with RSD of 2.5% by testing a SRM 1643d and a QC sample(HPS#290608). The proposed method could be applied to the determination of Arsenic(III) vapor in airwith a detection limit of 2.8 ng.-1 -aA Study on the Determination of Arsenic(III) and Arsenic(V) in Drinking Water and Arsenic(III) in Air by Graphite-Furnace Atomic Absorption Spectrophotometry 郭茂松 鄭政峰 葉玉堂 蔡素珍 1998 學位論文 ; thesis 97 zh-TW
collection NDLTD
language zh-TW
format Others
sources NDLTD
description 碩士 === 東海大學 === 環境科學系 === 86 === Arsenic(III) (12.5-200 ng) in drinking water (25 mL) reacted with 2,3-dimercaptopropane-1-sulfonate (DMPS,0.6 mg) in ammonium acetatebuffer (2.0 mmol, pH=5.5) and formed a complex of As(III)-DMPS. Thecomplex was concentrated on two Sep-Pak C18 cartridges in series and each cartridge was eluted with methanol and adjusted to 2.00 mL. After additionof 2.0 mg of Ni(II), a portion(20 mL) of the mixture was introduced into a graphite cuvette and was atomized according to a temperature program.The method detection limit (3s) was 0.11*0.01(mg/L) and the calibration graph was linear up to 8.00(mg/L) . Accuracies of 97.6-98.9% wereobtained when testing a QC (Strem Chem. Inc.) sample. Arsenic(V) (up to 100 ng) in drinking water was reduced to As(III) with L-cysteine(0.5 mL, 1.0% in HNO3 (0.2%,v/v)) prior to the formation of As(III)-DMPS complex. Accuracies of 98.5-99.3% were obtained with RSD of 2.5% by testing a SRM 1643d and a QC sample(HPS#290608). The proposed method could be applied to the determination of Arsenic(III) vapor in airwith a detection limit of 2.8 ng.-1 -aA Study on the Determination of Arsenic(III) and Arsenic(V) in Drinking Water and Arsenic(III) in Air by Graphite-Furnace Atomic Absorption Spectrophotometry
author2 郭茂松
author_facet 郭茂松
Hsieh,Chuan-Jier
謝傳傑
author Hsieh,Chuan-Jier
謝傳傑
spellingShingle Hsieh,Chuan-Jier
謝傳傑
A Study on the Determination of Arsenic(III) and Arsenic(V) in Drinking Water and Arsenic(III) in Air by Graphite-Furnace Atomic Absorption Spectrophotometry
author_sort Hsieh,Chuan-Jier
title A Study on the Determination of Arsenic(III) and Arsenic(V) in Drinking Water and Arsenic(III) in Air by Graphite-Furnace Atomic Absorption Spectrophotometry
title_short A Study on the Determination of Arsenic(III) and Arsenic(V) in Drinking Water and Arsenic(III) in Air by Graphite-Furnace Atomic Absorption Spectrophotometry
title_full A Study on the Determination of Arsenic(III) and Arsenic(V) in Drinking Water and Arsenic(III) in Air by Graphite-Furnace Atomic Absorption Spectrophotometry
title_fullStr A Study on the Determination of Arsenic(III) and Arsenic(V) in Drinking Water and Arsenic(III) in Air by Graphite-Furnace Atomic Absorption Spectrophotometry
title_full_unstemmed A Study on the Determination of Arsenic(III) and Arsenic(V) in Drinking Water and Arsenic(III) in Air by Graphite-Furnace Atomic Absorption Spectrophotometry
title_sort study on the determination of arsenic(iii) and arsenic(v) in drinking water and arsenic(iii) in air by graphite-furnace atomic absorption spectrophotometry
publishDate 1998
url http://ndltd.ncl.edu.tw/handle/65425009911038954192
work_keys_str_mv AT hsiehchuanjier astudyonthedeterminationofarseniciiiandarsenicvindrinkingwaterandarseniciiiinairbygraphitefurnaceatomicabsorptionspectrophotometry
AT xièchuánjié astudyonthedeterminationofarseniciiiandarsenicvindrinkingwaterandarseniciiiinairbygraphitefurnaceatomicabsorptionspectrophotometry
AT hsiehchuanjier yǐshímòdiànrèshìyuánzixīguāngfǎcèdìngyǐnyòngshuǐzhōngasiiiasvhékōngqìzhōngasiiizhīyánjiū
AT xièchuánjié yǐshímòdiànrèshìyuánzixīguāngfǎcèdìngyǐnyòngshuǐzhōngasiiiasvhékōngqìzhōngasiiizhīyánjiū
AT hsiehchuanjier studyonthedeterminationofarseniciiiandarsenicvindrinkingwaterandarseniciiiinairbygraphitefurnaceatomicabsorptionspectrophotometry
AT xièchuánjié studyonthedeterminationofarseniciiiandarsenicvindrinkingwaterandarseniciiiinairbygraphitefurnaceatomicabsorptionspectrophotometry
_version_ 1717781003278221312