A Preliminary Design of Central Monitoring╱Controlling System for Hemodialysis
碩士 === 中原大學 === 醫學工程學系 === 82 === In this paper, a preliminary monitoring and controlling system has been developed consisting of a personal computer with a dia- lysis machine. The serial interfaces including a singal process- ing card , Multi-8255 and an...
Main Authors: | , |
---|---|
Other Authors: | |
Format: | Others |
Language: | zh-TW |
Published: |
1994
|
Online Access: | http://ndltd.ncl.edu.tw/handle/93193929634935527868 |
id |
ndltd-TW-082CYCU0530017 |
---|---|
record_format |
oai_dc |
spelling |
ndltd-TW-082CYCU05300172016-02-10T04:08:56Z http://ndltd.ncl.edu.tw/handle/93193929634935527868 A Preliminary Design of Central Monitoring╱Controlling System for Hemodialysis 血液透析中心中央監控站初形開發 Yu, Keh Chung 于克忠 碩士 中原大學 醫學工程學系 82 In this paper, a preliminary monitoring and controlling system has been developed consisting of a personal computer with a dia- lysis machine. The serial interfaces including a singal process- ing card , Multi-8255 and an AD/DA interface card, were used for communication. By using the developed software in C language , graphics are created for monitoring the dialysis machine so that patient and system conditions may be easily understood. The moni- toring conditions include dialysate temperature,venous pressure, dialysate pressure , conductivity of dialysate ,total amount of ultrafilteration ,target of ultrafilteration , ultrafilteration rate, blood flow rate,blood leak and air bubble.The control sys- tem include the comfortable dialysis programing control of blood flow rate,the alteration of setting the mode of dialysis machine and the threshold of monitoring items. In this investigation, we execute preliminary for the central monitoring /controlling sys- tem and try to expand connecting with two dialysis machines. The singal processing card has been accomplished and produced.In hos- pital,the system is reliable and safe during the dialysis The system is now available that allow for monitoring and cont- roling one dialysis machine. In the future , it can connect with multi-dialysis machines to construct a network system . It will reduce the burden of physician and increase the safety of patient during dialysis treatment.84 Chang, Yen Lin 張炎林 1994 學位論文 ; thesis 120 zh-TW |
collection |
NDLTD |
language |
zh-TW |
format |
Others
|
sources |
NDLTD |
description |
碩士 === 中原大學 === 醫學工程學系 === 82 === In this paper, a preliminary monitoring and controlling system
has been developed consisting of a personal computer with a
dia- lysis machine. The serial interfaces including a singal
process- ing card , Multi-8255 and an AD/DA interface card,
were used for communication. By using the developed software in
C language , graphics are created for monitoring the dialysis
machine so that patient and system conditions may be easily
understood. The moni- toring conditions include dialysate
temperature,venous pressure, dialysate pressure , conductivity
of dialysate ,total amount of ultrafilteration ,target of
ultrafilteration , ultrafilteration rate, blood flow rate,blood
leak and air bubble.The control sys- tem include the
comfortable dialysis programing control of blood flow rate,the
alteration of setting the mode of dialysis machine and the
threshold of monitoring items. In this investigation, we
execute preliminary for the central monitoring /controlling
sys- tem and try to expand connecting with two dialysis
machines. The singal processing card has been accomplished and
produced.In hos- pital,the system is reliable and safe during
the dialysis The system is now available that allow for
monitoring and cont- roling one dialysis machine. In the future
, it can connect with multi-dialysis machines to construct a
network system . It will reduce the burden of physician and
increase the safety of patient during dialysis treatment.84
|
author2 |
Chang, Yen Lin |
author_facet |
Chang, Yen Lin Yu, Keh Chung 于克忠 |
author |
Yu, Keh Chung 于克忠 |
spellingShingle |
Yu, Keh Chung 于克忠 A Preliminary Design of Central Monitoring╱Controlling System for Hemodialysis |
author_sort |
Yu, Keh Chung |
title |
A Preliminary Design of Central Monitoring╱Controlling System for Hemodialysis |
title_short |
A Preliminary Design of Central Monitoring╱Controlling System for Hemodialysis |
title_full |
A Preliminary Design of Central Monitoring╱Controlling System for Hemodialysis |
title_fullStr |
A Preliminary Design of Central Monitoring╱Controlling System for Hemodialysis |
title_full_unstemmed |
A Preliminary Design of Central Monitoring╱Controlling System for Hemodialysis |
title_sort |
preliminary design of central monitoring╱controlling system for hemodialysis |
publishDate |
1994 |
url |
http://ndltd.ncl.edu.tw/handle/93193929634935527868 |
work_keys_str_mv |
AT yukehchung apreliminarydesignofcentralmonitoringcontrollingsystemforhemodialysis AT yúkèzhōng apreliminarydesignofcentralmonitoringcontrollingsystemforhemodialysis AT yukehchung xuèyètòuxīzhōngxīnzhōngyāngjiānkòngzhànchūxíngkāifā AT yúkèzhōng xuèyètòuxīzhōngxīnzhōngyāngjiānkòngzhànchūxíngkāifā AT yukehchung preliminarydesignofcentralmonitoringcontrollingsystemforhemodialysis AT yúkèzhōng preliminarydesignofcentralmonitoringcontrollingsystemforhemodialysis |
_version_ |
1718185716787183616 |