A collisional damping interface for a time-of-flight mass spectrometer with both electrospray ionization and matrix assisted laser desorption ionization

A collisional damping interface for orthogonal injection of either electrospray ionization (ESI) or matrix-assisted laser desorption/ionization (MALDI) ions into a reflecting TOF mass spectrometer (MANITOBA TOF III) has been designed. Ions produced by either ESI or MALDI are injected into an RF-quad...

Full description

Bibliographic Details
Main Author: Krutchinsky, Andrew
Language:en_US
Published: 2007
Online Access:http://hdl.handle.net/1993/1352
id ndltd-MANITOBA-oai-mspace.lib.umanitoba.ca-1993-1352
record_format oai_dc
spelling ndltd-MANITOBA-oai-mspace.lib.umanitoba.ca-1993-13522014-01-31T03:30:38Z A collisional damping interface for a time-of-flight mass spectrometer with both electrospray ionization and matrix assisted laser desorption ionization Krutchinsky, Andrew A collisional damping interface for orthogonal injection of either electrospray ionization (ESI) or matrix-assisted laser desorption/ionization (MALDI) ions into a reflecting TOF mass spectrometer (MANITOBA TOF III) has been designed. Ions produced by either ESI or MALDI are injected into an RF-quadrupole ion guide operating at pressure $\sim$70mTorr. As they pass through the quadrupole their motion is constrained by the RF field. Meanwhile, they lose energy by collisions with the buffer gas molecules. Computer simulations and experiments performed in the instrument with the new interface indicate that collisional cooling of ions improves the quality of the primary ion beams orthogonally injected into the mass spectrometer. Important characteristics of the mass spectrometer, such as resolution and sensitivity were greatly improved. In addition, mass discrimination was reduced, providing an observable mass range greater than 1 MDa with an electrospray ion source and up to $\sim$12,300 Da with a MALDI ion source. These improvements appeared to be very useful for the study of large biological compounds, noncovalent complexes in particular. The results of investigation of several noncovalent complexes are presented to illustrate the capabilities of the instrument with the new interface. The system studied consist of Echerichia coli citrate synthase dimers and hexamers interacting with an allosteric inhibitor, the reduced form of nicotinamide adenine nucleotide. 2007-05-17T12:33:30Z 2007-05-17T12:33:30Z 1998-07-01T00:00:00Z http://hdl.handle.net/1993/1352 en_US
collection NDLTD
language en_US
sources NDLTD
description A collisional damping interface for orthogonal injection of either electrospray ionization (ESI) or matrix-assisted laser desorption/ionization (MALDI) ions into a reflecting TOF mass spectrometer (MANITOBA TOF III) has been designed. Ions produced by either ESI or MALDI are injected into an RF-quadrupole ion guide operating at pressure $\sim$70mTorr. As they pass through the quadrupole their motion is constrained by the RF field. Meanwhile, they lose energy by collisions with the buffer gas molecules. Computer simulations and experiments performed in the instrument with the new interface indicate that collisional cooling of ions improves the quality of the primary ion beams orthogonally injected into the mass spectrometer. Important characteristics of the mass spectrometer, such as resolution and sensitivity were greatly improved. In addition, mass discrimination was reduced, providing an observable mass range greater than 1 MDa with an electrospray ion source and up to $\sim$12,300 Da with a MALDI ion source. These improvements appeared to be very useful for the study of large biological compounds, noncovalent complexes in particular. The results of investigation of several noncovalent complexes are presented to illustrate the capabilities of the instrument with the new interface. The system studied consist of Echerichia coli citrate synthase dimers and hexamers interacting with an allosteric inhibitor, the reduced form of nicotinamide adenine nucleotide.
author Krutchinsky, Andrew
spellingShingle Krutchinsky, Andrew
A collisional damping interface for a time-of-flight mass spectrometer with both electrospray ionization and matrix assisted laser desorption ionization
author_facet Krutchinsky, Andrew
author_sort Krutchinsky, Andrew
title A collisional damping interface for a time-of-flight mass spectrometer with both electrospray ionization and matrix assisted laser desorption ionization
title_short A collisional damping interface for a time-of-flight mass spectrometer with both electrospray ionization and matrix assisted laser desorption ionization
title_full A collisional damping interface for a time-of-flight mass spectrometer with both electrospray ionization and matrix assisted laser desorption ionization
title_fullStr A collisional damping interface for a time-of-flight mass spectrometer with both electrospray ionization and matrix assisted laser desorption ionization
title_full_unstemmed A collisional damping interface for a time-of-flight mass spectrometer with both electrospray ionization and matrix assisted laser desorption ionization
title_sort collisional damping interface for a time-of-flight mass spectrometer with both electrospray ionization and matrix assisted laser desorption ionization
publishDate 2007
url http://hdl.handle.net/1993/1352
work_keys_str_mv AT krutchinskyandrew acollisionaldampinginterfaceforatimeofflightmassspectrometerwithbothelectrosprayionizationandmatrixassistedlaserdesorptionionization
AT krutchinskyandrew collisionaldampinginterfaceforatimeofflightmassspectrometerwithbothelectrosprayionizationandmatrixassistedlaserdesorptionionization
_version_ 1716627947220107264