Calculation of water and graphite perturbation correction factors for the NACP-02 plane-parallel ionization chamber in high-energy electron beams

For megavoltage electron beams, current dosimetry protocols assume well-guarded plane-parallel ionization chambers have electron perturbation correction factors of unity. Papers by other researchers have shown the contrary. The National Physical Laboratory (NPL), Teddington UK, has an electron cal...

Full description

Bibliographic Details
Main Author: Chin, Erika
Other Authors: Frans Verhaegen (Internal/Supervisor)
Format: Others
Language:en
Published: McGill University 2008
Subjects:
Online Access:http://digitool.Library.McGill.CA:80/R/?func=dbin-jump-full&object_id=18699
id ndltd-LACETR-oai-collectionscanada.gc.ca-QMM.18699
record_format oai_dc
spelling ndltd-LACETR-oai-collectionscanada.gc.ca-QMM.186992014-02-13T03:53:32ZCalculation of water and graphite perturbation correction factors for the NACP-02 plane-parallel ionization chamber in high-energy electron beamsChin, ErikaPhysics - RadiationFor megavoltage electron beams, current dosimetry protocols assume well-guarded plane-parallel ionization chambers have electron perturbation correction factors of unity. Papers by other researchers have shown the contrary. The National Physical Laboratory (NPL), Teddington UK, has an electron calibration service that requires knowledge of the ratio of perturbation factors in water to graphite for their NACP-02 chamber. Using the Monte Carlo code EGSnrcMP, perturbation correction factors at depth zref in graphite for the NPL's Radiation Dynamics linac 4 – 19 MeV were calculated. These results combined with those in water calculated by Zakikhani (2006) showed that only the ratio of perturbation factors in water to graphite for the 4 MeV and 12 MeV beams, at 1.0080 ± 0.27 % and 1.0071 ± 0.28 %, were significantly different from unity. Perturbation correction factors in water and graphite for the Varian clinical linacs 4 – 18 MeV were also investigated. Important issues that arose requiring further study included the discrepancy in results caused by 1 keV vs. 10 keV transport cutoffs and whether the depth of the effective point of measurement should account for the phantom equivalent thickness of the chamber window.Pour les faisceaux d'électrons de haute énergie (MV), les protocoles de dosimétrie actuels supposent que les chambres d'ionisation à plaques parallèles ont des facteurs de correction de perturbation électronique égaux à l'unité. Plusieurs articles d'autres auteurs ont démontré le contraire. Le National Physical Laboratory (NPL), Teddington Royaume-Uni, possède un service d'étalonnage qui requiert de savoir les facteurs de perturbation de l'eau par rapport au graphite pour leur chambre NACP-02. A l'aide du code Monte Carlo EGSnrcMP, les facteurs de correction de perturbation à profondeur zref dans le graphite ont été calculés pour le linac Radiation Dynamics du NPL à des énergies de 4-19 MeV. Ces résultats, combinés à ceux dans l'eau calculés par Zakikhani (2006), ont montré que seulement les facteurs de perturbation de l'eau par rapport au graphite des faisceaux de 4 MeV et 12 MeV, à 1,0080 ± 0,27 % et 1,0071 ± 0,28 % respectivement, étaient significativement différents de l'unité. Les facteurs de correction de perturbation dans l'eau et dans le graphite ont aussi été étudiés pour les linacs cliniques de Varian à 4-18 MeV. D'importants problèmes ont alors été soulevés, notamment des différences dans les résultats causés par des énergies de coupure de 1 keV par rapport à 10 keV ainsi que la question de savoir si la profondeur du point de mesure devrait prendre en compte l'épaisseur effective de la fenêtre d'entrée. Ces problèmes requièrent davantage d'études.McGill UniversityFrans Verhaegen (Internal/Supervisor)Jan Peter Frans Seuntjens (Internal/Cosupervisor2)2008Electronic Thesis or Dissertationapplication/pdfenElectronically-submitted theses.All items in eScholarship@McGill are protected by copyright with all rights reserved unless otherwise indicated.Master of Science (Medical Physics Unit) http://digitool.Library.McGill.CA:80/R/?func=dbin-jump-full&object_id=18699
collection NDLTD
language en
format Others
sources NDLTD
topic Physics - Radiation
spellingShingle Physics - Radiation
Chin, Erika
Calculation of water and graphite perturbation correction factors for the NACP-02 plane-parallel ionization chamber in high-energy electron beams
description For megavoltage electron beams, current dosimetry protocols assume well-guarded plane-parallel ionization chambers have electron perturbation correction factors of unity. Papers by other researchers have shown the contrary. The National Physical Laboratory (NPL), Teddington UK, has an electron calibration service that requires knowledge of the ratio of perturbation factors in water to graphite for their NACP-02 chamber. Using the Monte Carlo code EGSnrcMP, perturbation correction factors at depth zref in graphite for the NPL's Radiation Dynamics linac 4 – 19 MeV were calculated. These results combined with those in water calculated by Zakikhani (2006) showed that only the ratio of perturbation factors in water to graphite for the 4 MeV and 12 MeV beams, at 1.0080 ± 0.27 % and 1.0071 ± 0.28 %, were significantly different from unity. Perturbation correction factors in water and graphite for the Varian clinical linacs 4 – 18 MeV were also investigated. Important issues that arose requiring further study included the discrepancy in results caused by 1 keV vs. 10 keV transport cutoffs and whether the depth of the effective point of measurement should account for the phantom equivalent thickness of the chamber window. === Pour les faisceaux d'électrons de haute énergie (MV), les protocoles de dosimétrie actuels supposent que les chambres d'ionisation à plaques parallèles ont des facteurs de correction de perturbation électronique égaux à l'unité. Plusieurs articles d'autres auteurs ont démontré le contraire. Le National Physical Laboratory (NPL), Teddington Royaume-Uni, possède un service d'étalonnage qui requiert de savoir les facteurs de perturbation de l'eau par rapport au graphite pour leur chambre NACP-02. A l'aide du code Monte Carlo EGSnrcMP, les facteurs de correction de perturbation à profondeur zref dans le graphite ont été calculés pour le linac Radiation Dynamics du NPL à des énergies de 4-19 MeV. Ces résultats, combinés à ceux dans l'eau calculés par Zakikhani (2006), ont montré que seulement les facteurs de perturbation de l'eau par rapport au graphite des faisceaux de 4 MeV et 12 MeV, à 1,0080 ± 0,27 % et 1,0071 ± 0,28 % respectivement, étaient significativement différents de l'unité. Les facteurs de correction de perturbation dans l'eau et dans le graphite ont aussi été étudiés pour les linacs cliniques de Varian à 4-18 MeV. D'importants problèmes ont alors été soulevés, notamment des différences dans les résultats causés par des énergies de coupure de 1 keV par rapport à 10 keV ainsi que la question de savoir si la profondeur du point de mesure devrait prendre en compte l'épaisseur effective de la fenêtre d'entrée. Ces problèmes requièrent davantage d'études.
author2 Frans Verhaegen (Internal/Supervisor)
author_facet Frans Verhaegen (Internal/Supervisor)
Chin, Erika
author Chin, Erika
author_sort Chin, Erika
title Calculation of water and graphite perturbation correction factors for the NACP-02 plane-parallel ionization chamber in high-energy electron beams
title_short Calculation of water and graphite perturbation correction factors for the NACP-02 plane-parallel ionization chamber in high-energy electron beams
title_full Calculation of water and graphite perturbation correction factors for the NACP-02 plane-parallel ionization chamber in high-energy electron beams
title_fullStr Calculation of water and graphite perturbation correction factors for the NACP-02 plane-parallel ionization chamber in high-energy electron beams
title_full_unstemmed Calculation of water and graphite perturbation correction factors for the NACP-02 plane-parallel ionization chamber in high-energy electron beams
title_sort calculation of water and graphite perturbation correction factors for the nacp-02 plane-parallel ionization chamber in high-energy electron beams
publisher McGill University
publishDate 2008
url http://digitool.Library.McGill.CA:80/R/?func=dbin-jump-full&object_id=18699
work_keys_str_mv AT chinerika calculationofwaterandgraphiteperturbationcorrectionfactorsforthenacp02planeparallelionizationchamberinhighenergyelectronbeams
_version_ 1716640793811222528