DESIGN AND MANUFACTURE OF A HIGH-FREQUENCY ANNULAR ARRAY ULTRASOUND SYSTEM FOR MEDICAL IMAGING

This thesis presents the design of a high-frequency annular array ultrasound system suitable for medical imaging. To reduce the cost of the system, off-the-shelf parts were used whenever possible. The system consists of four main components; 1) a transmit beamformer, 2) a high voltage pulse generato...

Full description

Bibliographic Details
Main Author: Lay, Holly Susan
Other Authors: Queen's University (Kingston, Ont.). Theses (Queen's University (Kingston, Ont.))
Language:en
en
Published: 2011
Subjects:
Online Access:http://hdl.handle.net/1974/6506
id ndltd-LACETR-oai-collectionscanada.gc.ca-OKQ.1974-6506
record_format oai_dc
spelling ndltd-LACETR-oai-collectionscanada.gc.ca-OKQ.1974-65062013-12-20T03:40:01ZDESIGN AND MANUFACTURE OF A HIGH-FREQUENCY ANNULAR ARRAY ULTRASOUND SYSTEM FOR MEDICAL IMAGINGLay, Holly SusanEngineering PhysicsMedical UltrasoundThis thesis presents the design of a high-frequency annular array ultrasound system suitable for medical imaging. To reduce the cost of the system, off-the-shelf parts were used whenever possible. The system consists of four main components; 1) a transmit beamformer, 2) a high voltage pulse generator, 3) an annular array transducer and 4) a receive beamformer. The transmit beamformer and pulser were designed for an 8-channel array but could be easily expanded for larger arrays. The pulser produces monocycle electrical pulses with centre frequencies that could be adjusted from 10-50 MHz and with amplitudes up to 90 Vpp. The annular array transducer has 12 equal area elements and a total active aperture of 6 mm. The transducer array produced pulses with a centre frequency of 20 MHz and 50% bandwidth. The resulting images had a lateral resolution of 172.5 μm at 10 mm and an axial resolution of 180 μm. A new fabrication method was developed that makes it easier to build the array. The receive beamformer was based on a commercial 8-channel analog-to-digital converter. The digital signals were transferred to a laptop where the beamforming was performed in software. This avoided the need to develop custom hardware and allowed it to be reconfigured for different transducers by simply modifying the software. The beamformer used a new interpolation method that reduced the required sampling frequency while maintaining a satisfactory radiation pattern. The system produces images at 10 frames/sec.Thesis (Ph.D, Physics, Engineering Physics and Astronomy) -- Queen's University, 2011-05-06 13:15:34.495Queen's University (Kingston, Ont.). Theses (Queen's University (Kingston, Ont.))2011-05-06 13:15:34.4952011-05-06T18:27:43Z2011-05-06T18:27:43Z2011-05-06T18:27:43ZThesishttp://hdl.handle.net/1974/6506enenCanadian thesesThis publication is made available by the authority of the copyright owner solely for the purpose of private study and research and may not be copied or reproduced except as permitted by the copyright laws without written authority from the copyright owner.
collection NDLTD
language en
en
sources NDLTD
topic Engineering Physics
Medical Ultrasound
spellingShingle Engineering Physics
Medical Ultrasound
Lay, Holly Susan
DESIGN AND MANUFACTURE OF A HIGH-FREQUENCY ANNULAR ARRAY ULTRASOUND SYSTEM FOR MEDICAL IMAGING
description This thesis presents the design of a high-frequency annular array ultrasound system suitable for medical imaging. To reduce the cost of the system, off-the-shelf parts were used whenever possible. The system consists of four main components; 1) a transmit beamformer, 2) a high voltage pulse generator, 3) an annular array transducer and 4) a receive beamformer. The transmit beamformer and pulser were designed for an 8-channel array but could be easily expanded for larger arrays. The pulser produces monocycle electrical pulses with centre frequencies that could be adjusted from 10-50 MHz and with amplitudes up to 90 Vpp. The annular array transducer has 12 equal area elements and a total active aperture of 6 mm. The transducer array produced pulses with a centre frequency of 20 MHz and 50% bandwidth. The resulting images had a lateral resolution of 172.5 μm at 10 mm and an axial resolution of 180 μm. A new fabrication method was developed that makes it easier to build the array. The receive beamformer was based on a commercial 8-channel analog-to-digital converter. The digital signals were transferred to a laptop where the beamforming was performed in software. This avoided the need to develop custom hardware and allowed it to be reconfigured for different transducers by simply modifying the software. The beamformer used a new interpolation method that reduced the required sampling frequency while maintaining a satisfactory radiation pattern. The system produces images at 10 frames/sec. === Thesis (Ph.D, Physics, Engineering Physics and Astronomy) -- Queen's University, 2011-05-06 13:15:34.495
author2 Queen's University (Kingston, Ont.). Theses (Queen's University (Kingston, Ont.))
author_facet Queen's University (Kingston, Ont.). Theses (Queen's University (Kingston, Ont.))
Lay, Holly Susan
author Lay, Holly Susan
author_sort Lay, Holly Susan
title DESIGN AND MANUFACTURE OF A HIGH-FREQUENCY ANNULAR ARRAY ULTRASOUND SYSTEM FOR MEDICAL IMAGING
title_short DESIGN AND MANUFACTURE OF A HIGH-FREQUENCY ANNULAR ARRAY ULTRASOUND SYSTEM FOR MEDICAL IMAGING
title_full DESIGN AND MANUFACTURE OF A HIGH-FREQUENCY ANNULAR ARRAY ULTRASOUND SYSTEM FOR MEDICAL IMAGING
title_fullStr DESIGN AND MANUFACTURE OF A HIGH-FREQUENCY ANNULAR ARRAY ULTRASOUND SYSTEM FOR MEDICAL IMAGING
title_full_unstemmed DESIGN AND MANUFACTURE OF A HIGH-FREQUENCY ANNULAR ARRAY ULTRASOUND SYSTEM FOR MEDICAL IMAGING
title_sort design and manufacture of a high-frequency annular array ultrasound system for medical imaging
publishDate 2011
url http://hdl.handle.net/1974/6506
work_keys_str_mv AT layhollysusan designandmanufactureofahighfrequencyannulararrayultrasoundsystemformedicalimaging
_version_ 1716621270693445632