Validation of conformal radiotherapy treatments in 3D using polymer gel dosimeters and optical computed tomography

Polymer gel dosimeters are a three dimensional (3D) dosimetry system that may be conveniently applied for verifying highly conformal radiation therapies where standard dosimetry techniques are insufficient. Polymer gel dosimetry with optical computed tomography (OptCT) can be used to measure spatia...

Full description

Bibliographic Details
Main Author: Holmes, OLIVER
Other Authors: Queen's University (Kingston, Ont.). Theses (Queen's University (Kingston, Ont.))
Format: Others
Language:en
en
Published: 2008
Subjects:
Online Access:http://hdl.handle.net/1974/1638
id ndltd-LACETR-oai-collectionscanada.gc.ca-OKQ.1974-1638
record_format oai_dc
spelling ndltd-LACETR-oai-collectionscanada.gc.ca-OKQ.1974-16382013-12-20T03:39:00ZValidation of conformal radiotherapy treatments in 3D using polymer gel dosimeters and optical computed tomographyHolmes, OLIVERRadiotherapyGel DosimetryPolymer gel dosimeters are a three dimensional (3D) dosimetry system that may be conveniently applied for verifying highly conformal radiation therapies where standard dosimetry techniques are insufficient. Polymer gel dosimetry with optical computed tomography (OptCT) can be used to measure spatial dose distributions with high resolution. While long experience with MRI has yielded many studies reporting on experiments involving validation of clinical deliveries using polymer gel dosimeters, there are very few studies of this type where OptCT is used. OptCT is a relatively new technique and consequently has not yet been adopted into the clinical environment. As a result, methods and software tools for integrating OptCT measurements into clinical systems are not available. Previous studies from the Medical Physics research group at the Cancer Centre of Southeastern Ontario (CCSEO) and Queen’s University have therefore been limited to simple deliveries and two dimensional (2D) comparisons. In this thesis various software tools and calibration techniques have been developed to allow comparative analysis between OptCT measurements with dose distributions calculated by treatment planning software. Further, a modification of the γ-evaluation (Low et al. 1998) is presented whereby the vector components of γ are used to identify the sources of disagreement between compared dose distributions. Test simulations of the new γ-tool revealed that individual vector components of γ, as well as the resulting vector field can be used to identify certain types of disagreements between dose distributions: especially spatial misalignments caused by geometric misses. The polymer gel dosimetry tools and analysis software were applied to a clinical validation mimicking a prostate conformal treatment with patient setup correction using image guidance. In one experiment greater than 90 % agreement was found between dose distributions in 4%T 50%C NIPAM/Bis dosimeters (Senden et al. 2006) measured with the Vista OptCT unit and dose distributions calculated by Eclipse treatment planning software.Thesis (Master, Physics, Engineering Physics and Astronomy) -- Queen's University, 2008-12-18 15:46:33.78Queen's University (Kingston, Ont.). Theses (Queen's University (Kingston, Ont.))2008-12-18 15:46:33.782008-12-18T21:57:51Z2008-12-18T21:57:51Z2008-12-18T21:57:51ZThesis7484251 bytesapplication/pdfhttp://hdl.handle.net/1974/1638enenCanadian thesesThis publication is made available by the authority of the copyright owner solely for the purpose of private study and research and may not be copied or reproduced except as permitted by the copyright laws without written authority from the copyright owner.
collection NDLTD
language en
en
format Others
sources NDLTD
topic Radiotherapy
Gel Dosimetry
spellingShingle Radiotherapy
Gel Dosimetry
Holmes, OLIVER
Validation of conformal radiotherapy treatments in 3D using polymer gel dosimeters and optical computed tomography
description Polymer gel dosimeters are a three dimensional (3D) dosimetry system that may be conveniently applied for verifying highly conformal radiation therapies where standard dosimetry techniques are insufficient. Polymer gel dosimetry with optical computed tomography (OptCT) can be used to measure spatial dose distributions with high resolution. While long experience with MRI has yielded many studies reporting on experiments involving validation of clinical deliveries using polymer gel dosimeters, there are very few studies of this type where OptCT is used. OptCT is a relatively new technique and consequently has not yet been adopted into the clinical environment. As a result, methods and software tools for integrating OptCT measurements into clinical systems are not available. Previous studies from the Medical Physics research group at the Cancer Centre of Southeastern Ontario (CCSEO) and Queen’s University have therefore been limited to simple deliveries and two dimensional (2D) comparisons. In this thesis various software tools and calibration techniques have been developed to allow comparative analysis between OptCT measurements with dose distributions calculated by treatment planning software. Further, a modification of the γ-evaluation (Low et al. 1998) is presented whereby the vector components of γ are used to identify the sources of disagreement between compared dose distributions. Test simulations of the new γ-tool revealed that individual vector components of γ, as well as the resulting vector field can be used to identify certain types of disagreements between dose distributions: especially spatial misalignments caused by geometric misses. The polymer gel dosimetry tools and analysis software were applied to a clinical validation mimicking a prostate conformal treatment with patient setup correction using image guidance. In one experiment greater than 90 % agreement was found between dose distributions in 4%T 50%C NIPAM/Bis dosimeters (Senden et al. 2006) measured with the Vista OptCT unit and dose distributions calculated by Eclipse treatment planning software. === Thesis (Master, Physics, Engineering Physics and Astronomy) -- Queen's University, 2008-12-18 15:46:33.78
author2 Queen's University (Kingston, Ont.). Theses (Queen's University (Kingston, Ont.))
author_facet Queen's University (Kingston, Ont.). Theses (Queen's University (Kingston, Ont.))
Holmes, OLIVER
author Holmes, OLIVER
author_sort Holmes, OLIVER
title Validation of conformal radiotherapy treatments in 3D using polymer gel dosimeters and optical computed tomography
title_short Validation of conformal radiotherapy treatments in 3D using polymer gel dosimeters and optical computed tomography
title_full Validation of conformal radiotherapy treatments in 3D using polymer gel dosimeters and optical computed tomography
title_fullStr Validation of conformal radiotherapy treatments in 3D using polymer gel dosimeters and optical computed tomography
title_full_unstemmed Validation of conformal radiotherapy treatments in 3D using polymer gel dosimeters and optical computed tomography
title_sort validation of conformal radiotherapy treatments in 3d using polymer gel dosimeters and optical computed tomography
publishDate 2008
url http://hdl.handle.net/1974/1638
work_keys_str_mv AT holmesoliver validationofconformalradiotherapytreatmentsin3dusingpolymergeldosimetersandopticalcomputedtomography
_version_ 1716620952315363328