DESIGN OF EXPERIMENTS TO ANALYZE THE INFLUENCE OF WATER ON THE UNCERTAINTY OF OIL FLOW MEASUREMENT WITH ULTRASONIC METERS

PONTIFÍCIA UNIVERSIDADE CATÓLICA DO RIO DE JANEIRO === COORDENAÇÃO DE APERFEIÇOAMENTO DO PESSOAL DE ENSINO SUPERIOR === PROGRAMA DE SUPORTE À PÓS-GRADUAÇÃO DE INSTS. DE ENSINO === O objetivo desta dissertação é avaliar a influência do teor de água e do fator do medidor na incerteza de malhas fiscais...

Full description

Bibliographic Details
Main Author: FILIPE CARELI DE ALMEIDA
Other Authors: CARLOS ROBERTO HALL BARBOSA
Language:Portuguese
Published: PONTIFÍCIA UNIVERSIDADE CATÓLICA DO RIO DE JANEIRO 2017
Online Access:http://www.maxwell.vrac.puc-rio.br/Busca_etds.php?strSecao=resultado&nrSeq=33051@1
http://www.maxwell.vrac.puc-rio.br/Busca_etds.php?strSecao=resultado&nrSeq=33051@2
id ndltd-IBICT-oai-MAXWELL.puc-rio.br-33051
record_format oai_dc
collection NDLTD
language Portuguese
sources NDLTD
description PONTIFÍCIA UNIVERSIDADE CATÓLICA DO RIO DE JANEIRO === COORDENAÇÃO DE APERFEIÇOAMENTO DO PESSOAL DE ENSINO SUPERIOR === PROGRAMA DE SUPORTE À PÓS-GRADUAÇÃO DE INSTS. DE ENSINO === O objetivo desta dissertação é avaliar a influência do teor de água e do fator do medidor na incerteza de malhas fiscais de medição de óleo utilizando medidores ultrassônicos. A motivação deste trabalho advém do fato do Regulamento Técnico de Medição de Petróleo e Gás Natural (RTM) existente limitar quais metodologias podem ser utilizadas para realizar a análise de teor de água e qual a incerteza máxima permitida para a malha de medição na qual esta se insere, porém não define a incerteza desta, havendo o risco de se adotar um dos métodos de análise permitidos e, ainda assim, a incerteza de medição da malha ultrapassar a definida pela legislação. Além disso, o RTM adota um critério de repetibilidade para a validação das calibrações dos medidores ultrassônicos, portanto avalia-se também a influência deste na incerteza da malha. A metodologia utilizada foi o planejamento de experimentos, utilizando superfície de resposta. Os resultados demonstram que é possível simular a influência da incerteza da análise da água e do fator do medidor na incerteza da malha. Por fim, as conclusões demonstram que as médias atualmente encontradas para incerteza do teor de água são adequadas para manter a incerteza da malha dentro da permitida pelo RTM, porém com pouca variação pode-se exceder este limite; a influência da incerteza do fator do medidor foi inferior ou similar à do teor de água e, por fim, sugere-se a adoção de um critério limite de incerteza ao invés de um critério de repetibilidade para os medidores ultrassônicos. === The aim of this dissertation is to evaluate the influence of water content dissolved in crude oil and the meter factor on the uncertainty of the fiscal oil metering using ultrasonic flow meters. The development of this work is motivated by the fact that the Technical Regulation for the Measurement of Oil and Natural Gas (RTM) limits which methodologies can be used for water analysis and what is the maximum permissible uncertainty for the oil flow meter in which it is inserted. However, the RTM doesn t define the uncertainty of the water analysis. There is a risk of adopting one of the permitted methods of analysis and, even so, the measurement uncertainty of the flow meter exceeds the defined by the legislation. In addition, the RTM adopts a repeatability standard when validating the calibration of ultrasonic meters, thus the influence of this on oil uncertainty is also evaluated. The methodology used is the design of experiments and response surface. The results demonstrate that the influence of the water content uncertainty analysis and of the meter factor on the uncertainty of the oil flow meter can be simulated. Finally, the conclusions show that the average currently found for uncertainty of water is adequate to maintain the oil uncertainty within the allowed by the RTM, but small variations may cause this limit to be exceeded. The influence of the meter factor uncertainty was inferior or similar to the water content and, finally, it is suggested the adoption of a standard limit of uncertainty instead of a standard of repeatability for the ultrasonic meters.
author2 CARLOS ROBERTO HALL BARBOSA
author_facet CARLOS ROBERTO HALL BARBOSA
FILIPE CARELI DE ALMEIDA
author FILIPE CARELI DE ALMEIDA
spellingShingle FILIPE CARELI DE ALMEIDA
DESIGN OF EXPERIMENTS TO ANALYZE THE INFLUENCE OF WATER ON THE UNCERTAINTY OF OIL FLOW MEASUREMENT WITH ULTRASONIC METERS
author_sort FILIPE CARELI DE ALMEIDA
title DESIGN OF EXPERIMENTS TO ANALYZE THE INFLUENCE OF WATER ON THE UNCERTAINTY OF OIL FLOW MEASUREMENT WITH ULTRASONIC METERS
title_short DESIGN OF EXPERIMENTS TO ANALYZE THE INFLUENCE OF WATER ON THE UNCERTAINTY OF OIL FLOW MEASUREMENT WITH ULTRASONIC METERS
title_full DESIGN OF EXPERIMENTS TO ANALYZE THE INFLUENCE OF WATER ON THE UNCERTAINTY OF OIL FLOW MEASUREMENT WITH ULTRASONIC METERS
title_fullStr DESIGN OF EXPERIMENTS TO ANALYZE THE INFLUENCE OF WATER ON THE UNCERTAINTY OF OIL FLOW MEASUREMENT WITH ULTRASONIC METERS
title_full_unstemmed DESIGN OF EXPERIMENTS TO ANALYZE THE INFLUENCE OF WATER ON THE UNCERTAINTY OF OIL FLOW MEASUREMENT WITH ULTRASONIC METERS
title_sort design of experiments to analyze the influence of water on the uncertainty of oil flow measurement with ultrasonic meters
publisher PONTIFÍCIA UNIVERSIDADE CATÓLICA DO RIO DE JANEIRO
publishDate 2017
url http://www.maxwell.vrac.puc-rio.br/Busca_etds.php?strSecao=resultado&nrSeq=33051@1
http://www.maxwell.vrac.puc-rio.br/Busca_etds.php?strSecao=resultado&nrSeq=33051@2
work_keys_str_mv AT filipecarelidealmeida designofexperimentstoanalyzetheinfluenceofwaterontheuncertaintyofoilflowmeasurementwithultrasonicmeters
AT filipecarelidealmeida planejamentodeexperimentosparaanalisedainfluenciadoteordeaguanaincertezademedicaodevazaodeoleocommedidoresultrassonicos
_version_ 1718989309622616064
spelling ndltd-IBICT-oai-MAXWELL.puc-rio.br-330512019-03-01T15:43:53Z DESIGN OF EXPERIMENTS TO ANALYZE THE INFLUENCE OF WATER ON THE UNCERTAINTY OF OIL FLOW MEASUREMENT WITH ULTRASONIC METERS PLANEJAMENTO DE EXPERIMENTOS PARA ANÁLISE DA INFLUÊNCIA DO TEOR DE ÁGUA NA INCERTEZA DE MEDIÇÃO DE VAZÃO DE ÓLEO COM MEDIDORES ULTRASSÔNICOS FILIPE CARELI DE ALMEIDA CARLOS ROBERTO HALL BARBOSA ELCIO CRUZ DE OLIVEIRA ELCIO CRUZ DE OLIVEIRA RICARDO QUEIROZ AUCELIO CARLOS ROBERTO HALL BARBOSA VERONICA MARIA DE ARAUJO CALADO ELCIO CRUZ DE OLIVEIRA ELCIO CRUZ DE OLIVEIRA PONTIFÍCIA UNIVERSIDADE CATÓLICA DO RIO DE JANEIRO COORDENAÇÃO DE APERFEIÇOAMENTO DO PESSOAL DE ENSINO SUPERIOR PROGRAMA DE SUPORTE À PÓS-GRADUAÇÃO DE INSTS. DE ENSINO O objetivo desta dissertação é avaliar a influência do teor de água e do fator do medidor na incerteza de malhas fiscais de medição de óleo utilizando medidores ultrassônicos. A motivação deste trabalho advém do fato do Regulamento Técnico de Medição de Petróleo e Gás Natural (RTM) existente limitar quais metodologias podem ser utilizadas para realizar a análise de teor de água e qual a incerteza máxima permitida para a malha de medição na qual esta se insere, porém não define a incerteza desta, havendo o risco de se adotar um dos métodos de análise permitidos e, ainda assim, a incerteza de medição da malha ultrapassar a definida pela legislação. Além disso, o RTM adota um critério de repetibilidade para a validação das calibrações dos medidores ultrassônicos, portanto avalia-se também a influência deste na incerteza da malha. A metodologia utilizada foi o planejamento de experimentos, utilizando superfície de resposta. Os resultados demonstram que é possível simular a influência da incerteza da análise da água e do fator do medidor na incerteza da malha. Por fim, as conclusões demonstram que as médias atualmente encontradas para incerteza do teor de água são adequadas para manter a incerteza da malha dentro da permitida pelo RTM, porém com pouca variação pode-se exceder este limite; a influência da incerteza do fator do medidor foi inferior ou similar à do teor de água e, por fim, sugere-se a adoção de um critério limite de incerteza ao invés de um critério de repetibilidade para os medidores ultrassônicos. The aim of this dissertation is to evaluate the influence of water content dissolved in crude oil and the meter factor on the uncertainty of the fiscal oil metering using ultrasonic flow meters. The development of this work is motivated by the fact that the Technical Regulation for the Measurement of Oil and Natural Gas (RTM) limits which methodologies can be used for water analysis and what is the maximum permissible uncertainty for the oil flow meter in which it is inserted. However, the RTM doesn t define the uncertainty of the water analysis. There is a risk of adopting one of the permitted methods of analysis and, even so, the measurement uncertainty of the flow meter exceeds the defined by the legislation. In addition, the RTM adopts a repeatability standard when validating the calibration of ultrasonic meters, thus the influence of this on oil uncertainty is also evaluated. The methodology used is the design of experiments and response surface. The results demonstrate that the influence of the water content uncertainty analysis and of the meter factor on the uncertainty of the oil flow meter can be simulated. Finally, the conclusions show that the average currently found for uncertainty of water is adequate to maintain the oil uncertainty within the allowed by the RTM, but small variations may cause this limit to be exceeded. The influence of the meter factor uncertainty was inferior or similar to the water content and, finally, it is suggested the adoption of a standard limit of uncertainty instead of a standard of repeatability for the ultrasonic meters. 2017-04-07 info:eu-repo/semantics/publishedVersion info:eu-repo/semantics/masterThesis http://www.maxwell.vrac.puc-rio.br/Busca_etds.php?strSecao=resultado&nrSeq=33051@1 http://www.maxwell.vrac.puc-rio.br/Busca_etds.php?strSecao=resultado&nrSeq=33051@2 por info:eu-repo/semantics/openAccess PONTIFÍCIA UNIVERSIDADE CATÓLICA DO RIO DE JANEIRO PPG EM METROLOGIA PUC-Rio BR reponame:Repositório Institucional da PUC_RIO instname:Pontifícia Universidade Católica do Rio de Janeiro instacron:PUC_RIO