The Sysmex CS-5100 coagulation analyzer offers comparable analytical performance and excellent throughput capabilities
Objectives: This study compared the new high-volume blood coagulation analyzer Sysmex CS-5100 System™ (Siemens Healthcare Diagnostics, Erlangen, Germany) to the mid-volume blood coagulation analyzer Sysmex CS-2000i System™ (Siemens) for analytical performance. Additionally, the operational performan...
Main Authors: | , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Elsevier
2016-12-01
|
Series: | Practical Laboratory Medicine |
Online Access: | http://www.sciencedirect.com/science/article/pii/S2352551716300324 |
id |
doaj-fc9e21539f7c4e9191ab639d2946cd54 |
---|---|
record_format |
Article |
spelling |
doaj-fc9e21539f7c4e9191ab639d2946cd542020-11-24T20:44:36ZengElsevierPractical Laboratory Medicine2352-55172016-12-0163847The Sysmex CS-5100 coagulation analyzer offers comparable analytical performance and excellent throughput capabilitiesT. Flieder0T. Gripp1C. Knabbe2I. Birschmann3Institute for Laboratory and Transfusion Medicine, Heart and Diabetes Center, Ruhr University Bochum, Georgstrasse 11, 32545 Bad Oeynhausen, GermanyInstitute for Laboratory and Transfusion Medicine, Heart and Diabetes Center, Ruhr University Bochum, Georgstrasse 11, 32545 Bad Oeynhausen, GermanyInstitute for Laboratory and Transfusion Medicine, Heart and Diabetes Center, Ruhr University Bochum, Georgstrasse 11, 32545 Bad Oeynhausen, GermanyCorresponding author.; Institute for Laboratory and Transfusion Medicine, Heart and Diabetes Center, Ruhr University Bochum, Georgstrasse 11, 32545 Bad Oeynhausen, GermanyObjectives: This study compared the new high-volume blood coagulation analyzer Sysmex CS-5100 System™ (Siemens Healthcare Diagnostics, Erlangen, Germany) to the mid-volume blood coagulation analyzer Sysmex CS-2000i System™ (Siemens) for analytical performance. Additionally, the operational performance of the Sysmex CS-5100 System was compared with the blood coagulation analyzer ACL TOP 700 (Instrumentation Laboratory, Werfen Group, Kirchheim bei Munchen, Germany). Materials and methods: We compared the Sysmex CS-5100 to the Sysmex CS-2000i and the ACL TOP analyzer for routine coagulation, chromogenic and immunological assays. Imprecision studies were performed for the Sysmex CS-5100 and Sysmex CS-2000i systems. A throughput and STAT analysis comparison of the CS-5100 and the ACL TOP was performed. A stress test was performed to characterize the robustness and the error rate of the CS-5100. We also performed correlation analysis between the CS-5100 and the CS-2000i or the ACL TOP in the measurement of patients’ samples. Results: The inter-assay precision using the CS systems was impressive (inter-assay CV generally <3.5%) and the correlation between the two Sysmex analyzers was excellent. In the throughput study, the CS-5100 completed the measurement of 100 samples (210 results) in less than 49 min. Conclusions: Our results demonstrated that the CS-5100 is a robust high-throughput analyzer, well-suited for coagulation laboratories. Keywords: Coagulation analyzer, CS-5100, Evaluation, Automationhttp://www.sciencedirect.com/science/article/pii/S2352551716300324 |
collection |
DOAJ |
language |
English |
format |
Article |
sources |
DOAJ |
author |
T. Flieder T. Gripp C. Knabbe I. Birschmann |
spellingShingle |
T. Flieder T. Gripp C. Knabbe I. Birschmann The Sysmex CS-5100 coagulation analyzer offers comparable analytical performance and excellent throughput capabilities Practical Laboratory Medicine |
author_facet |
T. Flieder T. Gripp C. Knabbe I. Birschmann |
author_sort |
T. Flieder |
title |
The Sysmex CS-5100 coagulation analyzer offers comparable analytical performance and excellent throughput capabilities |
title_short |
The Sysmex CS-5100 coagulation analyzer offers comparable analytical performance and excellent throughput capabilities |
title_full |
The Sysmex CS-5100 coagulation analyzer offers comparable analytical performance and excellent throughput capabilities |
title_fullStr |
The Sysmex CS-5100 coagulation analyzer offers comparable analytical performance and excellent throughput capabilities |
title_full_unstemmed |
The Sysmex CS-5100 coagulation analyzer offers comparable analytical performance and excellent throughput capabilities |
title_sort |
sysmex cs-5100 coagulation analyzer offers comparable analytical performance and excellent throughput capabilities |
publisher |
Elsevier |
series |
Practical Laboratory Medicine |
issn |
2352-5517 |
publishDate |
2016-12-01 |
description |
Objectives: This study compared the new high-volume blood coagulation analyzer Sysmex CS-5100 System™ (Siemens Healthcare Diagnostics, Erlangen, Germany) to the mid-volume blood coagulation analyzer Sysmex CS-2000i System™ (Siemens) for analytical performance. Additionally, the operational performance of the Sysmex CS-5100 System was compared with the blood coagulation analyzer ACL TOP 700 (Instrumentation Laboratory, Werfen Group, Kirchheim bei Munchen, Germany). Materials and methods: We compared the Sysmex CS-5100 to the Sysmex CS-2000i and the ACL TOP analyzer for routine coagulation, chromogenic and immunological assays. Imprecision studies were performed for the Sysmex CS-5100 and Sysmex CS-2000i systems. A throughput and STAT analysis comparison of the CS-5100 and the ACL TOP was performed. A stress test was performed to characterize the robustness and the error rate of the CS-5100. We also performed correlation analysis between the CS-5100 and the CS-2000i or the ACL TOP in the measurement of patients’ samples. Results: The inter-assay precision using the CS systems was impressive (inter-assay CV generally <3.5%) and the correlation between the two Sysmex analyzers was excellent. In the throughput study, the CS-5100 completed the measurement of 100 samples (210 results) in less than 49 min. Conclusions: Our results demonstrated that the CS-5100 is a robust high-throughput analyzer, well-suited for coagulation laboratories. Keywords: Coagulation analyzer, CS-5100, Evaluation, Automation |
url |
http://www.sciencedirect.com/science/article/pii/S2352551716300324 |
work_keys_str_mv |
AT tflieder thesysmexcs5100coagulationanalyzerofferscomparableanalyticalperformanceandexcellentthroughputcapabilities AT tgripp thesysmexcs5100coagulationanalyzerofferscomparableanalyticalperformanceandexcellentthroughputcapabilities AT cknabbe thesysmexcs5100coagulationanalyzerofferscomparableanalyticalperformanceandexcellentthroughputcapabilities AT ibirschmann thesysmexcs5100coagulationanalyzerofferscomparableanalyticalperformanceandexcellentthroughputcapabilities AT tflieder sysmexcs5100coagulationanalyzerofferscomparableanalyticalperformanceandexcellentthroughputcapabilities AT tgripp sysmexcs5100coagulationanalyzerofferscomparableanalyticalperformanceandexcellentthroughputcapabilities AT cknabbe sysmexcs5100coagulationanalyzerofferscomparableanalyticalperformanceandexcellentthroughputcapabilities AT ibirschmann sysmexcs5100coagulationanalyzerofferscomparableanalyticalperformanceandexcellentthroughputcapabilities |