Polyethylene liner dissociation with the Pinnacle acetabular component: should we be concerned?

Between 2007 and 2018, 535 total hip arthroplasties using the uncemented Pinnacle acetabular component (DePuy Synthes, Warsaw, IN) and polyethylene liner were implanted in our unit. Of these, 6 patients presented acutely with liner polyethylene dissociation, giving a rate of liner dissociation of 1....

Full description

Bibliographic Details
Main Authors: Adeel R. Memon, MBBS, MCh, FRCS, David Gwynne-Jones, MA, FRCS, FRACS (Orth)
Format: Article
Language:English
Published: Elsevier 2020-03-01
Series:Arthroplasty Today
Online Access:http://www.sciencedirect.com/science/article/pii/S2352344119301621
id doaj-f776082c2307488ba3893e354d21db89
record_format Article
spelling doaj-f776082c2307488ba3893e354d21db892020-11-25T01:14:16ZengElsevierArthroplasty Today2352-34412020-03-016158Polyethylene liner dissociation with the Pinnacle acetabular component: should we be concerned?Adeel R. Memon, MBBS, MCh, FRCS0David Gwynne-Jones, MA, FRCS, FRACS (Orth)1Corresponding author. Dunedin Hospital, Great King street Dunedin, Otago, New Zealand. Tel.: +64 3 4740999.; Department of Trauma and Orthopaedics, Dunedin Hospital Dunedin, Dunedin, New ZealandDepartment of Trauma and Orthopaedics, Dunedin Hospital Dunedin, Dunedin, New ZealandBetween 2007 and 2018, 535 total hip arthroplasties using the uncemented Pinnacle acetabular component (DePuy Synthes, Warsaw, IN) and polyethylene liner were implanted in our unit. Of these, 6 patients presented acutely with liner polyethylene dissociation, giving a rate of liner dissociation of 1.11%. All dissociations were atraumatic. Failure occurred at mean 37 months (range 4.5 to 130 months). Radiologically, all acetabular components were within safe zone of abduction and mean anteversion was 10 degrees (range 2-20). In one case, there was posterior impingement against the femoral neck due to femoral malalignment. All patients underwent head and liner exchange with no repeat failures. Despite excellent long-term results, the frequency of dissociated polyethylene liners is a cause of concern with the Pinnacle acetabular component. Keywords: Pinnacle, Liner dissociation, Acetabular cuphttp://www.sciencedirect.com/science/article/pii/S2352344119301621
collection DOAJ
language English
format Article
sources DOAJ
author Adeel R. Memon, MBBS, MCh, FRCS
David Gwynne-Jones, MA, FRCS, FRACS (Orth)
spellingShingle Adeel R. Memon, MBBS, MCh, FRCS
David Gwynne-Jones, MA, FRCS, FRACS (Orth)
Polyethylene liner dissociation with the Pinnacle acetabular component: should we be concerned?
Arthroplasty Today
author_facet Adeel R. Memon, MBBS, MCh, FRCS
David Gwynne-Jones, MA, FRCS, FRACS (Orth)
author_sort Adeel R. Memon, MBBS, MCh, FRCS
title Polyethylene liner dissociation with the Pinnacle acetabular component: should we be concerned?
title_short Polyethylene liner dissociation with the Pinnacle acetabular component: should we be concerned?
title_full Polyethylene liner dissociation with the Pinnacle acetabular component: should we be concerned?
title_fullStr Polyethylene liner dissociation with the Pinnacle acetabular component: should we be concerned?
title_full_unstemmed Polyethylene liner dissociation with the Pinnacle acetabular component: should we be concerned?
title_sort polyethylene liner dissociation with the pinnacle acetabular component: should we be concerned?
publisher Elsevier
series Arthroplasty Today
issn 2352-3441
publishDate 2020-03-01
description Between 2007 and 2018, 535 total hip arthroplasties using the uncemented Pinnacle acetabular component (DePuy Synthes, Warsaw, IN) and polyethylene liner were implanted in our unit. Of these, 6 patients presented acutely with liner polyethylene dissociation, giving a rate of liner dissociation of 1.11%. All dissociations were atraumatic. Failure occurred at mean 37 months (range 4.5 to 130 months). Radiologically, all acetabular components were within safe zone of abduction and mean anteversion was 10 degrees (range 2-20). In one case, there was posterior impingement against the femoral neck due to femoral malalignment. All patients underwent head and liner exchange with no repeat failures. Despite excellent long-term results, the frequency of dissociated polyethylene liners is a cause of concern with the Pinnacle acetabular component. Keywords: Pinnacle, Liner dissociation, Acetabular cup
url http://www.sciencedirect.com/science/article/pii/S2352344119301621
work_keys_str_mv AT adeelrmemonmbbsmchfrcs polyethylenelinerdissociationwiththepinnacleacetabularcomponentshouldwebeconcerned
AT davidgwynnejonesmafrcsfracsorth polyethylenelinerdissociationwiththepinnacleacetabularcomponentshouldwebeconcerned
_version_ 1725157748444758016