Contrastive analysis of corneal horizontal diameter measurement by five instruments

AIM: To analyze the correlation and consistency of five corneal diameter measurements.<p>METHODS: Totally 25 cases(50 eyes)who underwent ICL implantation in West China Hospital. The preoperative horizontal corneal diameter was measured using measuring caliper, SIRIUS anterior eye assay system,...

Full description

Bibliographic Details
Main Authors: Xiao-Lan Zhang, Rui Gong, Shun-Qing Wang, Yong-Zhi Huang
Format: Article
Language:English
Published: Press of International Journal of Ophthalmology (IJO PRESS) 2019-04-01
Series:Guoji Yanke Zazhi
Subjects:
UBM
Online Access:http://ies.ijo.cn/cn_publish/2019/4/201904042.pdf
id doaj-f387515b63d6452fa6dd3bcaf77dc9e2
record_format Article
spelling doaj-f387515b63d6452fa6dd3bcaf77dc9e22020-11-24T21:54:24ZengPress of International Journal of Ophthalmology (IJO PRESS)Guoji Yanke Zazhi1672-51231672-51232019-04-0119470470810.3980/j.issn.1672-5123.2019.4.42Contrastive analysis of corneal horizontal diameter measurement by five instrumentsXiao-Lan Zhang0Rui Gong1Shun-Qing Wang2Yong-Zhi Huang3Department of Ophthalmology, West China Hospital, Sichuan University, Chengdu 610041, Sichuan Province, ChinaDepartment of Ophthalmology, West China Hospital, Sichuan University, Chengdu 610041, Sichuan Province, ChinaDepartment of Ophthalmology, West China Hospital, Sichuan University, Chengdu 610041, Sichuan Province, ChinaDepartment of Ophthalmology, West China Hospital, Sichuan University, Chengdu 610041, Sichuan Province, ChinaAIM: To analyze the correlation and consistency of five corneal diameter measurements.<p>METHODS: Totally 25 cases(50 eyes)who underwent ICL implantation in West China Hospital. The preoperative horizontal corneal diameter was measured using measuring caliper, SIRIUS anterior eye assay system, IOL Master500, anterior segment OCT and UBM. <p>RESULTS: The mean WTW distances were 11.54±0.30 mm as obtained with measuring caliper, 11.77±0.33mm with SIRIUS anterior eye assay system, 11.98±0.33mm with IOL Master500, 11.63±0.35mm with anterior segment OCT and 11.53±0.34mm with UBM. No statistical difference was found between measuring caliper and UBM, measuring caliper and anterior segment OCT, UBM and anterior segment OCT. The linear correlation analysis found significant correlation between the measurements of the five measurements. The Bland-Altman analysis for the measuring caliper and SIRIUS, measuring caliper and UBM, measuring caliper and anterior segment OCT found that the absolute values of 95% LOA upper and lower limits were less than 0.5mm.<p>CONCLUSION: The results of the four kinds of corneal horizontal diameter measurements can be interchanged including SIRIUS, UBM, anterior segment OCT and measuring caliper. IOL Master500 results are the largest, that cannot be used as a diagnostic basis for measuring the size of corneal horizontal diameter. The results of other measurement equipment should be combined with clinical practice.http://ies.ijo.cn/cn_publish/2019/4/201904042.pdfcorneal diametercaliperSIRIUS anterior eye assay systemIOL Master500anterior segment OCTUBMconsistency analysis
collection DOAJ
language English
format Article
sources DOAJ
author Xiao-Lan Zhang
Rui Gong
Shun-Qing Wang
Yong-Zhi Huang
spellingShingle Xiao-Lan Zhang
Rui Gong
Shun-Qing Wang
Yong-Zhi Huang
Contrastive analysis of corneal horizontal diameter measurement by five instruments
Guoji Yanke Zazhi
corneal diameter
caliper
SIRIUS anterior eye assay system
IOL Master500
anterior segment OCT
UBM
consistency analysis
author_facet Xiao-Lan Zhang
Rui Gong
Shun-Qing Wang
Yong-Zhi Huang
author_sort Xiao-Lan Zhang
title Contrastive analysis of corneal horizontal diameter measurement by five instruments
title_short Contrastive analysis of corneal horizontal diameter measurement by five instruments
title_full Contrastive analysis of corneal horizontal diameter measurement by five instruments
title_fullStr Contrastive analysis of corneal horizontal diameter measurement by five instruments
title_full_unstemmed Contrastive analysis of corneal horizontal diameter measurement by five instruments
title_sort contrastive analysis of corneal horizontal diameter measurement by five instruments
publisher Press of International Journal of Ophthalmology (IJO PRESS)
series Guoji Yanke Zazhi
issn 1672-5123
1672-5123
publishDate 2019-04-01
description AIM: To analyze the correlation and consistency of five corneal diameter measurements.<p>METHODS: Totally 25 cases(50 eyes)who underwent ICL implantation in West China Hospital. The preoperative horizontal corneal diameter was measured using measuring caliper, SIRIUS anterior eye assay system, IOL Master500, anterior segment OCT and UBM. <p>RESULTS: The mean WTW distances were 11.54±0.30 mm as obtained with measuring caliper, 11.77±0.33mm with SIRIUS anterior eye assay system, 11.98±0.33mm with IOL Master500, 11.63±0.35mm with anterior segment OCT and 11.53±0.34mm with UBM. No statistical difference was found between measuring caliper and UBM, measuring caliper and anterior segment OCT, UBM and anterior segment OCT. The linear correlation analysis found significant correlation between the measurements of the five measurements. The Bland-Altman analysis for the measuring caliper and SIRIUS, measuring caliper and UBM, measuring caliper and anterior segment OCT found that the absolute values of 95% LOA upper and lower limits were less than 0.5mm.<p>CONCLUSION: The results of the four kinds of corneal horizontal diameter measurements can be interchanged including SIRIUS, UBM, anterior segment OCT and measuring caliper. IOL Master500 results are the largest, that cannot be used as a diagnostic basis for measuring the size of corneal horizontal diameter. The results of other measurement equipment should be combined with clinical practice.
topic corneal diameter
caliper
SIRIUS anterior eye assay system
IOL Master500
anterior segment OCT
UBM
consistency analysis
url http://ies.ijo.cn/cn_publish/2019/4/201904042.pdf
work_keys_str_mv AT xiaolanzhang contrastiveanalysisofcornealhorizontaldiametermeasurementbyfiveinstruments
AT ruigong contrastiveanalysisofcornealhorizontaldiametermeasurementbyfiveinstruments
AT shunqingwang contrastiveanalysisofcornealhorizontaldiametermeasurementbyfiveinstruments
AT yongzhihuang contrastiveanalysisofcornealhorizontaldiametermeasurementbyfiveinstruments
_version_ 1725867228545417216