Contrastive analysis of corneal horizontal diameter measurement by five instruments
AIM: To analyze the correlation and consistency of five corneal diameter measurements.<p>METHODS: Totally 25 cases(50 eyes)who underwent ICL implantation in West China Hospital. The preoperative horizontal corneal diameter was measured using measuring caliper, SIRIUS anterior eye assay system,...
Main Authors: | , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Press of International Journal of Ophthalmology (IJO PRESS)
2019-04-01
|
Series: | Guoji Yanke Zazhi |
Subjects: | |
Online Access: | http://ies.ijo.cn/cn_publish/2019/4/201904042.pdf |
id |
doaj-f387515b63d6452fa6dd3bcaf77dc9e2 |
---|---|
record_format |
Article |
spelling |
doaj-f387515b63d6452fa6dd3bcaf77dc9e22020-11-24T21:54:24ZengPress of International Journal of Ophthalmology (IJO PRESS)Guoji Yanke Zazhi1672-51231672-51232019-04-0119470470810.3980/j.issn.1672-5123.2019.4.42Contrastive analysis of corneal horizontal diameter measurement by five instrumentsXiao-Lan Zhang0Rui Gong1Shun-Qing Wang2Yong-Zhi Huang3Department of Ophthalmology, West China Hospital, Sichuan University, Chengdu 610041, Sichuan Province, ChinaDepartment of Ophthalmology, West China Hospital, Sichuan University, Chengdu 610041, Sichuan Province, ChinaDepartment of Ophthalmology, West China Hospital, Sichuan University, Chengdu 610041, Sichuan Province, ChinaDepartment of Ophthalmology, West China Hospital, Sichuan University, Chengdu 610041, Sichuan Province, ChinaAIM: To analyze the correlation and consistency of five corneal diameter measurements.<p>METHODS: Totally 25 cases(50 eyes)who underwent ICL implantation in West China Hospital. The preoperative horizontal corneal diameter was measured using measuring caliper, SIRIUS anterior eye assay system, IOL Master500, anterior segment OCT and UBM. <p>RESULTS: The mean WTW distances were 11.54±0.30 mm as obtained with measuring caliper, 11.77±0.33mm with SIRIUS anterior eye assay system, 11.98±0.33mm with IOL Master500, 11.63±0.35mm with anterior segment OCT and 11.53±0.34mm with UBM. No statistical difference was found between measuring caliper and UBM, measuring caliper and anterior segment OCT, UBM and anterior segment OCT. The linear correlation analysis found significant correlation between the measurements of the five measurements. The Bland-Altman analysis for the measuring caliper and SIRIUS, measuring caliper and UBM, measuring caliper and anterior segment OCT found that the absolute values of 95% LOA upper and lower limits were less than 0.5mm.<p>CONCLUSION: The results of the four kinds of corneal horizontal diameter measurements can be interchanged including SIRIUS, UBM, anterior segment OCT and measuring caliper. IOL Master500 results are the largest, that cannot be used as a diagnostic basis for measuring the size of corneal horizontal diameter. The results of other measurement equipment should be combined with clinical practice.http://ies.ijo.cn/cn_publish/2019/4/201904042.pdfcorneal diametercaliperSIRIUS anterior eye assay systemIOL Master500anterior segment OCTUBMconsistency analysis |
collection |
DOAJ |
language |
English |
format |
Article |
sources |
DOAJ |
author |
Xiao-Lan Zhang Rui Gong Shun-Qing Wang Yong-Zhi Huang |
spellingShingle |
Xiao-Lan Zhang Rui Gong Shun-Qing Wang Yong-Zhi Huang Contrastive analysis of corneal horizontal diameter measurement by five instruments Guoji Yanke Zazhi corneal diameter caliper SIRIUS anterior eye assay system IOL Master500 anterior segment OCT UBM consistency analysis |
author_facet |
Xiao-Lan Zhang Rui Gong Shun-Qing Wang Yong-Zhi Huang |
author_sort |
Xiao-Lan Zhang |
title |
Contrastive analysis of corneal horizontal diameter measurement by five instruments |
title_short |
Contrastive analysis of corneal horizontal diameter measurement by five instruments |
title_full |
Contrastive analysis of corneal horizontal diameter measurement by five instruments |
title_fullStr |
Contrastive analysis of corneal horizontal diameter measurement by five instruments |
title_full_unstemmed |
Contrastive analysis of corneal horizontal diameter measurement by five instruments |
title_sort |
contrastive analysis of corneal horizontal diameter measurement by five instruments |
publisher |
Press of International Journal of Ophthalmology (IJO PRESS) |
series |
Guoji Yanke Zazhi |
issn |
1672-5123 1672-5123 |
publishDate |
2019-04-01 |
description |
AIM: To analyze the correlation and consistency of five corneal diameter measurements.<p>METHODS: Totally 25 cases(50 eyes)who underwent ICL implantation in West China Hospital. The preoperative horizontal corneal diameter was measured using measuring caliper, SIRIUS anterior eye assay system, IOL Master500, anterior segment OCT and UBM. <p>RESULTS: The mean WTW distances were 11.54±0.30 mm as obtained with measuring caliper, 11.77±0.33mm with SIRIUS anterior eye assay system, 11.98±0.33mm with IOL Master500, 11.63±0.35mm with anterior segment OCT and 11.53±0.34mm with UBM. No statistical difference was found between measuring caliper and UBM, measuring caliper and anterior segment OCT, UBM and anterior segment OCT. The linear correlation analysis found significant correlation between the measurements of the five measurements. The Bland-Altman analysis for the measuring caliper and SIRIUS, measuring caliper and UBM, measuring caliper and anterior segment OCT found that the absolute values of 95% LOA upper and lower limits were less than 0.5mm.<p>CONCLUSION: The results of the four kinds of corneal horizontal diameter measurements can be interchanged including SIRIUS, UBM, anterior segment OCT and measuring caliper. IOL Master500 results are the largest, that cannot be used as a diagnostic basis for measuring the size of corneal horizontal diameter. The results of other measurement equipment should be combined with clinical practice. |
topic |
corneal diameter caliper SIRIUS anterior eye assay system IOL Master500 anterior segment OCT UBM consistency analysis |
url |
http://ies.ijo.cn/cn_publish/2019/4/201904042.pdf |
work_keys_str_mv |
AT xiaolanzhang contrastiveanalysisofcornealhorizontaldiametermeasurementbyfiveinstruments AT ruigong contrastiveanalysisofcornealhorizontaldiametermeasurementbyfiveinstruments AT shunqingwang contrastiveanalysisofcornealhorizontaldiametermeasurementbyfiveinstruments AT yongzhihuang contrastiveanalysisofcornealhorizontaldiametermeasurementbyfiveinstruments |
_version_ |
1725867228545417216 |