Evaluation of the Accuracy and Related Factors of the Mechanical Torque-Limiting Device for Dental Implants

Objective: Accurate delivery of torque to implant screws is critical to generate ideal preload in the screw joint and to offer protection against screw loosening. Mechanical torque-limiting devices (MTLDs) are available for this reason. In this study, the accuracy of one type of friction-style and t...

Full description

Bibliographic Details
Main Authors: Mahmood Kazemi, Ahmad Rohanian, Abbas Monzavi, Mohammad Sadegh Nazari
Format: Article
Language:English
Published: Tehran University of Medical Sciences 2013-01-01
Series:Journal of Dentistry of Tehran University of Medical Sciences
Subjects:
Online Access:http://jdt.tums.ac.ir/index.php/jdt/article/view/827/399
id doaj-f2ed36359f424318bcfc7d78406ff1a5
record_format Article
spelling doaj-f2ed36359f424318bcfc7d78406ff1a52020-11-25T03:44:35ZengTehran University of Medical SciencesJournal of Dentistry of Tehran University of Medical Sciences 2008-21852013-01-01102112118Evaluation of the Accuracy and Related Factors of the Mechanical Torque-Limiting Device for Dental ImplantsMahmood KazemiAhmad RohanianAbbas MonzaviMohammad Sadegh NazariObjective: Accurate delivery of torque to implant screws is critical to generate ideal preload in the screw joint and to offer protection against screw loosening. Mechanical torque-limiting devices (MTLDs) are available for this reason. In this study, the accuracy of one type of friction-style and two types of spring-style MTLDs at baseline, following fatigue conditions and sterilization processes were determined.Materials and Methods: Five unused MTLDs were selected from each of Straumann (ITI), Astra TECH and CWM systems. To measure the output of each MTLD, a digital torque gauge with a 3-jaw chuck was used to hold the driver. Force was applied to the MTLDs until either the friction styles released at a pre-calibrated torque value or the spring styles flexed to a pre-calibrated limit (target torque value). The peak torque value was recorded and the procedure was repeated 5 times for each MTLD. Then MTLDs were subjected to fatigue conditions at 500 and 1000 times and steam sterilization processes at 50 and 100 times and the peak torque value was recorded again at each stage.Results: Adjusted difference between measured torque values and target torque values differed significantly between stages for all 3 systems. Adjusted difference did not differ significantly between systems at all stages, but differed significantly between two different styles at baseline and 500 times fatigue stages.Conclusion: Straumann (ITI) devices differed minimally from target torque values at all stages. MTLDs with Spring-style were significantly more accurate than Friction-style device in achieving their target torque values at baseline and 500 times fatiguehttp://jdt.tums.ac.ir/index.php/jdt/article/view/827/399Dental ImplantationTorqueCalibration
collection DOAJ
language English
format Article
sources DOAJ
author Mahmood Kazemi
Ahmad Rohanian
Abbas Monzavi
Mohammad Sadegh Nazari
spellingShingle Mahmood Kazemi
Ahmad Rohanian
Abbas Monzavi
Mohammad Sadegh Nazari
Evaluation of the Accuracy and Related Factors of the Mechanical Torque-Limiting Device for Dental Implants
Journal of Dentistry of Tehran University of Medical Sciences
Dental Implantation
Torque
Calibration
author_facet Mahmood Kazemi
Ahmad Rohanian
Abbas Monzavi
Mohammad Sadegh Nazari
author_sort Mahmood Kazemi
title Evaluation of the Accuracy and Related Factors of the Mechanical Torque-Limiting Device for Dental Implants
title_short Evaluation of the Accuracy and Related Factors of the Mechanical Torque-Limiting Device for Dental Implants
title_full Evaluation of the Accuracy and Related Factors of the Mechanical Torque-Limiting Device for Dental Implants
title_fullStr Evaluation of the Accuracy and Related Factors of the Mechanical Torque-Limiting Device for Dental Implants
title_full_unstemmed Evaluation of the Accuracy and Related Factors of the Mechanical Torque-Limiting Device for Dental Implants
title_sort evaluation of the accuracy and related factors of the mechanical torque-limiting device for dental implants
publisher Tehran University of Medical Sciences
series Journal of Dentistry of Tehran University of Medical Sciences
issn 2008-2185
publishDate 2013-01-01
description Objective: Accurate delivery of torque to implant screws is critical to generate ideal preload in the screw joint and to offer protection against screw loosening. Mechanical torque-limiting devices (MTLDs) are available for this reason. In this study, the accuracy of one type of friction-style and two types of spring-style MTLDs at baseline, following fatigue conditions and sterilization processes were determined.Materials and Methods: Five unused MTLDs were selected from each of Straumann (ITI), Astra TECH and CWM systems. To measure the output of each MTLD, a digital torque gauge with a 3-jaw chuck was used to hold the driver. Force was applied to the MTLDs until either the friction styles released at a pre-calibrated torque value or the spring styles flexed to a pre-calibrated limit (target torque value). The peak torque value was recorded and the procedure was repeated 5 times for each MTLD. Then MTLDs were subjected to fatigue conditions at 500 and 1000 times and steam sterilization processes at 50 and 100 times and the peak torque value was recorded again at each stage.Results: Adjusted difference between measured torque values and target torque values differed significantly between stages for all 3 systems. Adjusted difference did not differ significantly between systems at all stages, but differed significantly between two different styles at baseline and 500 times fatigue stages.Conclusion: Straumann (ITI) devices differed minimally from target torque values at all stages. MTLDs with Spring-style were significantly more accurate than Friction-style device in achieving their target torque values at baseline and 500 times fatigue
topic Dental Implantation
Torque
Calibration
url http://jdt.tums.ac.ir/index.php/jdt/article/view/827/399
work_keys_str_mv AT mahmoodkazemi evaluationoftheaccuracyandrelatedfactorsofthemechanicaltorquelimitingdevicefordentalimplants
AT ahmadrohanian evaluationoftheaccuracyandrelatedfactorsofthemechanicaltorquelimitingdevicefordentalimplants
AT abbasmonzavi evaluationoftheaccuracyandrelatedfactorsofthemechanicaltorquelimitingdevicefordentalimplants
AT mohammadsadeghnazari evaluationoftheaccuracyandrelatedfactorsofthemechanicaltorquelimitingdevicefordentalimplants
_version_ 1724513941513568256