Evaluation of the Accuracy and Related Factors of the Mechanical Torque-Limiting Device for Dental Implants
Objective: Accurate delivery of torque to implant screws is critical to generate ideal preload in the screw joint and to offer protection against screw loosening. Mechanical torque-limiting devices (MTLDs) are available for this reason. In this study, the accuracy of one type of friction-style and t...
Main Authors: | , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Tehran University of Medical Sciences
2013-01-01
|
Series: | Journal of Dentistry of Tehran University of Medical Sciences |
Subjects: | |
Online Access: | http://jdt.tums.ac.ir/index.php/jdt/article/view/827/399 |
id |
doaj-f2ed36359f424318bcfc7d78406ff1a5 |
---|---|
record_format |
Article |
spelling |
doaj-f2ed36359f424318bcfc7d78406ff1a52020-11-25T03:44:35ZengTehran University of Medical SciencesJournal of Dentistry of Tehran University of Medical Sciences 2008-21852013-01-01102112118Evaluation of the Accuracy and Related Factors of the Mechanical Torque-Limiting Device for Dental ImplantsMahmood KazemiAhmad RohanianAbbas MonzaviMohammad Sadegh NazariObjective: Accurate delivery of torque to implant screws is critical to generate ideal preload in the screw joint and to offer protection against screw loosening. Mechanical torque-limiting devices (MTLDs) are available for this reason. In this study, the accuracy of one type of friction-style and two types of spring-style MTLDs at baseline, following fatigue conditions and sterilization processes were determined.Materials and Methods: Five unused MTLDs were selected from each of Straumann (ITI), Astra TECH and CWM systems. To measure the output of each MTLD, a digital torque gauge with a 3-jaw chuck was used to hold the driver. Force was applied to the MTLDs until either the friction styles released at a pre-calibrated torque value or the spring styles flexed to a pre-calibrated limit (target torque value). The peak torque value was recorded and the procedure was repeated 5 times for each MTLD. Then MTLDs were subjected to fatigue conditions at 500 and 1000 times and steam sterilization processes at 50 and 100 times and the peak torque value was recorded again at each stage.Results: Adjusted difference between measured torque values and target torque values differed significantly between stages for all 3 systems. Adjusted difference did not differ significantly between systems at all stages, but differed significantly between two different styles at baseline and 500 times fatigue stages.Conclusion: Straumann (ITI) devices differed minimally from target torque values at all stages. MTLDs with Spring-style were significantly more accurate than Friction-style device in achieving their target torque values at baseline and 500 times fatiguehttp://jdt.tums.ac.ir/index.php/jdt/article/view/827/399Dental ImplantationTorqueCalibration |
collection |
DOAJ |
language |
English |
format |
Article |
sources |
DOAJ |
author |
Mahmood Kazemi Ahmad Rohanian Abbas Monzavi Mohammad Sadegh Nazari |
spellingShingle |
Mahmood Kazemi Ahmad Rohanian Abbas Monzavi Mohammad Sadegh Nazari Evaluation of the Accuracy and Related Factors of the Mechanical Torque-Limiting Device for Dental Implants Journal of Dentistry of Tehran University of Medical Sciences Dental Implantation Torque Calibration |
author_facet |
Mahmood Kazemi Ahmad Rohanian Abbas Monzavi Mohammad Sadegh Nazari |
author_sort |
Mahmood Kazemi |
title |
Evaluation of the Accuracy and Related Factors of the Mechanical Torque-Limiting Device for Dental Implants |
title_short |
Evaluation of the Accuracy and Related Factors of the Mechanical Torque-Limiting Device for Dental Implants |
title_full |
Evaluation of the Accuracy and Related Factors of the Mechanical Torque-Limiting Device for Dental Implants |
title_fullStr |
Evaluation of the Accuracy and Related Factors of the Mechanical Torque-Limiting Device for Dental Implants |
title_full_unstemmed |
Evaluation of the Accuracy and Related Factors of the Mechanical Torque-Limiting Device for Dental Implants |
title_sort |
evaluation of the accuracy and related factors of the mechanical torque-limiting device for dental implants |
publisher |
Tehran University of Medical Sciences |
series |
Journal of Dentistry of Tehran University of Medical Sciences |
issn |
2008-2185 |
publishDate |
2013-01-01 |
description |
Objective: Accurate delivery of torque to implant screws is critical to generate ideal preload in the screw joint and to offer protection against screw loosening. Mechanical torque-limiting devices (MTLDs) are available for this reason. In this study, the accuracy of one type of friction-style and two types of spring-style MTLDs at baseline, following fatigue conditions and sterilization processes were determined.Materials and Methods: Five unused MTLDs were selected from each of Straumann (ITI), Astra TECH and CWM systems. To measure the output of each MTLD, a digital torque gauge with a 3-jaw chuck was used to hold the driver. Force was applied to the MTLDs until either the friction styles released at a pre-calibrated torque value or the spring styles flexed to a pre-calibrated limit (target torque value). The peak torque value was recorded and the procedure was repeated 5 times for each MTLD. Then MTLDs were subjected to fatigue conditions at 500 and 1000 times and steam sterilization processes at 50 and 100 times and the peak torque value was recorded again at each stage.Results: Adjusted difference between measured torque values and target torque values differed significantly between stages for all 3 systems. Adjusted difference did not differ significantly between systems at all stages, but differed significantly between two different styles at baseline and 500 times fatigue stages.Conclusion: Straumann (ITI) devices differed minimally from target torque values at all stages. MTLDs with Spring-style were significantly more accurate than Friction-style device in achieving their target torque values at baseline and 500 times fatigue |
topic |
Dental Implantation Torque Calibration |
url |
http://jdt.tums.ac.ir/index.php/jdt/article/view/827/399 |
work_keys_str_mv |
AT mahmoodkazemi evaluationoftheaccuracyandrelatedfactorsofthemechanicaltorquelimitingdevicefordentalimplants AT ahmadrohanian evaluationoftheaccuracyandrelatedfactorsofthemechanicaltorquelimitingdevicefordentalimplants AT abbasmonzavi evaluationoftheaccuracyandrelatedfactorsofthemechanicaltorquelimitingdevicefordentalimplants AT mohammadsadeghnazari evaluationoftheaccuracyandrelatedfactorsofthemechanicaltorquelimitingdevicefordentalimplants |
_version_ |
1724513941513568256 |