Comparison of flow pressures in different 3-way infusion devices: an in-vitro study

Abstract Background The use of multiple infusions through one cannula is an increasingly common practice in anaesthesia. High pressures in the line often lead to occlusion alarms and pump disconnection. In this study, we aim to determine the pressures generated in common 3-way infusion devices, usin...

Full description

Bibliographic Details
Main Authors: Jonathan Chua, Arun Ratnavadivel
Format: Article
Language:English
Published: BMC 2018-06-01
Series:Patient Safety in Surgery
Subjects:
Online Access:http://link.springer.com/article/10.1186/s13037-018-0165-1
id doaj-ef84d346e710484ab8afa2617e7b16e7
record_format Article
spelling doaj-ef84d346e710484ab8afa2617e7b16e72020-11-24T20:40:21ZengBMCPatient Safety in Surgery1754-94932018-06-011211610.1186/s13037-018-0165-1Comparison of flow pressures in different 3-way infusion devices: an in-vitro studyJonathan Chua0Arun Ratnavadivel1Department of Anaesthesia, Nepean HospitalDepartment of Anaesthesia, Nepean HospitalAbstract Background The use of multiple infusions through one cannula is an increasingly common practice in anaesthesia. High pressures in the line often lead to occlusion alarms and pump disconnection. In this study, we aim to determine the pressures generated in common 3-way infusion devices, using simple low-cost equipment available and currently in use in the operating theatre environment. Methods We compared three different common and commercially available 3-way infusion devices that allowed multiple infusions through one cannula in vitro. One with anti-siphon valves, one without valves, and serial extension sets with side ports connected in series. An invasive blood pressure transducer was used to monitor line pressure. Seven different infusion rates were used to simulate different infusions. Results 3-way infusion devices with anti-siphon valves have 5.5 times the infusion pressures compared to devices without valves (P < 0.001). The highest pressures obtained across all devices were at the highest flow rate studied (400 ml/hr); this was 243 mmHg in the 3-way device with anti-siphon valves, compared to only 44 mmHg in the 3-way device without valves and 36 mmHg in the serial extension sets. Serial extension sets have the lowest pressures across all flow rates when compared to 3-way devices without valves. (P = 0.0001). Conclusions The presence of anti-siphon valves generate very high pressures in infusion lines that can contribute to occlusion alarm disconnection of a pump. However, when measured alone and in-vitro, these pressures are not sufficient to trigger occlusion alarms. There are 3-way infusion devices without anti-siphon valves that have lower line pressures, but clinicians should be aware of negative pressure scenarios which can lead to siphoning when using them.http://link.springer.com/article/10.1186/s13037-018-0165-1TransducersPressureInfusion pumps
collection DOAJ
language English
format Article
sources DOAJ
author Jonathan Chua
Arun Ratnavadivel
spellingShingle Jonathan Chua
Arun Ratnavadivel
Comparison of flow pressures in different 3-way infusion devices: an in-vitro study
Patient Safety in Surgery
Transducers
Pressure
Infusion pumps
author_facet Jonathan Chua
Arun Ratnavadivel
author_sort Jonathan Chua
title Comparison of flow pressures in different 3-way infusion devices: an in-vitro study
title_short Comparison of flow pressures in different 3-way infusion devices: an in-vitro study
title_full Comparison of flow pressures in different 3-way infusion devices: an in-vitro study
title_fullStr Comparison of flow pressures in different 3-way infusion devices: an in-vitro study
title_full_unstemmed Comparison of flow pressures in different 3-way infusion devices: an in-vitro study
title_sort comparison of flow pressures in different 3-way infusion devices: an in-vitro study
publisher BMC
series Patient Safety in Surgery
issn 1754-9493
publishDate 2018-06-01
description Abstract Background The use of multiple infusions through one cannula is an increasingly common practice in anaesthesia. High pressures in the line often lead to occlusion alarms and pump disconnection. In this study, we aim to determine the pressures generated in common 3-way infusion devices, using simple low-cost equipment available and currently in use in the operating theatre environment. Methods We compared three different common and commercially available 3-way infusion devices that allowed multiple infusions through one cannula in vitro. One with anti-siphon valves, one without valves, and serial extension sets with side ports connected in series. An invasive blood pressure transducer was used to monitor line pressure. Seven different infusion rates were used to simulate different infusions. Results 3-way infusion devices with anti-siphon valves have 5.5 times the infusion pressures compared to devices without valves (P < 0.001). The highest pressures obtained across all devices were at the highest flow rate studied (400 ml/hr); this was 243 mmHg in the 3-way device with anti-siphon valves, compared to only 44 mmHg in the 3-way device without valves and 36 mmHg in the serial extension sets. Serial extension sets have the lowest pressures across all flow rates when compared to 3-way devices without valves. (P = 0.0001). Conclusions The presence of anti-siphon valves generate very high pressures in infusion lines that can contribute to occlusion alarm disconnection of a pump. However, when measured alone and in-vitro, these pressures are not sufficient to trigger occlusion alarms. There are 3-way infusion devices without anti-siphon valves that have lower line pressures, but clinicians should be aware of negative pressure scenarios which can lead to siphoning when using them.
topic Transducers
Pressure
Infusion pumps
url http://link.springer.com/article/10.1186/s13037-018-0165-1
work_keys_str_mv AT jonathanchua comparisonofflowpressuresindifferent3wayinfusiondevicesaninvitrostudy
AT arunratnavadivel comparisonofflowpressuresindifferent3wayinfusiondevicesaninvitrostudy
_version_ 1716827244074106880