Patient and surgery related factors associated with fatigue type polyethylene wear on 49 PCA and DURACON retrievals at autopsy and revision

<p>Abstract</p> <p>Background</p> <p>Polyethylene wear is an important factor for longevity of total knee arthroplasty. Proven and suspicious factors causing wear can be grouped as material, patient and surgery related. There are more studies correlating design and/or b...

Full description

Bibliographic Details
Main Authors: Ochsner Peter E, Lüem Martin, Rohrbach Markus
Format: Article
Language:English
Published: BMC 2008-02-01
Series:Journal of Orthopaedic Surgery and Research
Online Access:http://www.josr-online.com/content/3/1/8
id doaj-e7f8fc2b8a7c484dab06323d331e79fe
record_format Article
spelling doaj-e7f8fc2b8a7c484dab06323d331e79fe2020-11-24T20:41:27ZengBMCJournal of Orthopaedic Surgery and Research1749-799X2008-02-0131810.1186/1749-799X-3-8Patient and surgery related factors associated with fatigue type polyethylene wear on 49 PCA and DURACON retrievals at autopsy and revisionOchsner Peter ELüem MartinRohrbach Markus<p>Abstract</p> <p>Background</p> <p>Polyethylene wear is an important factor for longevity of total knee arthroplasty. Proven and suspicious factors causing wear can be grouped as material, patient and surgery related. There are more studies correlating design and/or biomaterial factors to in vivo wear than those to patient and surgery related factors. Many retrieval studies just include revision implants and therefore may not be representative. This study is aimed to correlate patient- and surgery- related factors to visual wear score by minimizing design influence and include both autopsy and revision implants. Comparison between the groups was expected to unmask patient and surgery-related factors responsible for wear.</p> <p>Methods</p> <p>The amount of joint side wear on polyethylene retrievals was measured using a modification of an established visual wear score. Fatigue type wear was defined as summation of the most severe wear modes of delamination, pitting and cracks. Analysis of patient and surgery related variables suspicious to cause wear included prospectively sampled patient activity which was measured by self reported walking capacity. Statistical analysis was done by univariate analysis of variance. Activity level and implantation time were merged to an index of use and correlated to the wear score.</p> <p>Results</p> <p>Wear score after comparable implantation time was significantly less in the autopsy group. Even so, fatigue type wear accounted for 84 and 93 % of total wear score on autopsy and revision implants respectively. A highly significant influence on wear score was found in time of implantation (p = 0.002), level of activity (p = 0.025) and inserts belonging to revision group (p = 0.006). No influence was found for the kind of patella replacement (p = 0.483). Body mass index and accuracy of component alignment had no significant influence on visual wear score. Fatigue-type wear in the medial compartment was closely correlated to the index of use in the autopsy (R<sup>2 </sup>= 0.383) and the revision group (R<sup>2 </sup>= 0.813).</p> <p>Conclusion</p> <p>The present study's finding of substantial fatigue type wear in both autopsy and revision retrievals supports the theory that polyethylene fatigue strength is generally exceeded in this type of prosthesis. Furthermore, this study correlated fatigue-type polyethylene wear to an index of use as calculated by activity over time. Future retrieval studies may use activity over time as an important patient related factor correlated to the visual wear score. When evaluating total knee arthroplasty routine follow up, the surgeon must think of substantial wear present even without major clinical signs.</p> http://www.josr-online.com/content/3/1/8
collection DOAJ
language English
format Article
sources DOAJ
author Ochsner Peter E
Lüem Martin
Rohrbach Markus
spellingShingle Ochsner Peter E
Lüem Martin
Rohrbach Markus
Patient and surgery related factors associated with fatigue type polyethylene wear on 49 PCA and DURACON retrievals at autopsy and revision
Journal of Orthopaedic Surgery and Research
author_facet Ochsner Peter E
Lüem Martin
Rohrbach Markus
author_sort Ochsner Peter E
title Patient and surgery related factors associated with fatigue type polyethylene wear on 49 PCA and DURACON retrievals at autopsy and revision
title_short Patient and surgery related factors associated with fatigue type polyethylene wear on 49 PCA and DURACON retrievals at autopsy and revision
title_full Patient and surgery related factors associated with fatigue type polyethylene wear on 49 PCA and DURACON retrievals at autopsy and revision
title_fullStr Patient and surgery related factors associated with fatigue type polyethylene wear on 49 PCA and DURACON retrievals at autopsy and revision
title_full_unstemmed Patient and surgery related factors associated with fatigue type polyethylene wear on 49 PCA and DURACON retrievals at autopsy and revision
title_sort patient and surgery related factors associated with fatigue type polyethylene wear on 49 pca and duracon retrievals at autopsy and revision
publisher BMC
series Journal of Orthopaedic Surgery and Research
issn 1749-799X
publishDate 2008-02-01
description <p>Abstract</p> <p>Background</p> <p>Polyethylene wear is an important factor for longevity of total knee arthroplasty. Proven and suspicious factors causing wear can be grouped as material, patient and surgery related. There are more studies correlating design and/or biomaterial factors to in vivo wear than those to patient and surgery related factors. Many retrieval studies just include revision implants and therefore may not be representative. This study is aimed to correlate patient- and surgery- related factors to visual wear score by minimizing design influence and include both autopsy and revision implants. Comparison between the groups was expected to unmask patient and surgery-related factors responsible for wear.</p> <p>Methods</p> <p>The amount of joint side wear on polyethylene retrievals was measured using a modification of an established visual wear score. Fatigue type wear was defined as summation of the most severe wear modes of delamination, pitting and cracks. Analysis of patient and surgery related variables suspicious to cause wear included prospectively sampled patient activity which was measured by self reported walking capacity. Statistical analysis was done by univariate analysis of variance. Activity level and implantation time were merged to an index of use and correlated to the wear score.</p> <p>Results</p> <p>Wear score after comparable implantation time was significantly less in the autopsy group. Even so, fatigue type wear accounted for 84 and 93 % of total wear score on autopsy and revision implants respectively. A highly significant influence on wear score was found in time of implantation (p = 0.002), level of activity (p = 0.025) and inserts belonging to revision group (p = 0.006). No influence was found for the kind of patella replacement (p = 0.483). Body mass index and accuracy of component alignment had no significant influence on visual wear score. Fatigue-type wear in the medial compartment was closely correlated to the index of use in the autopsy (R<sup>2 </sup>= 0.383) and the revision group (R<sup>2 </sup>= 0.813).</p> <p>Conclusion</p> <p>The present study's finding of substantial fatigue type wear in both autopsy and revision retrievals supports the theory that polyethylene fatigue strength is generally exceeded in this type of prosthesis. Furthermore, this study correlated fatigue-type polyethylene wear to an index of use as calculated by activity over time. Future retrieval studies may use activity over time as an important patient related factor correlated to the visual wear score. When evaluating total knee arthroplasty routine follow up, the surgeon must think of substantial wear present even without major clinical signs.</p>
url http://www.josr-online.com/content/3/1/8
work_keys_str_mv AT ochsnerpetere patientandsurgeryrelatedfactorsassociatedwithfatiguetypepolyethylenewearon49pcaandduraconretrievalsatautopsyandrevision
AT luemmartin patientandsurgeryrelatedfactorsassociatedwithfatiguetypepolyethylenewearon49pcaandduraconretrievalsatautopsyandrevision
AT rohrbachmarkus patientandsurgeryrelatedfactorsassociatedwithfatiguetypepolyethylenewearon49pcaandduraconretrievalsatautopsyandrevision
_version_ 1716825041976426496