Fingerprint analysis of placenta polypeptide injection by high performance liquid chromatography

Objective: To develop the representative fingerprint for the quality control of placenta polypeptide injection. Methods: The chromatographic separation was performed using a Phenomenex Gemini C18 column (250 mmÃ4.6 mm, 5 μm) maintained at 30 °C. 0.1% aqueous trifluoroacetic acid (Solvent A) an...

Full description

Bibliographic Details
Main Authors: Li Huang, Xiao-Man Wu, Yu Ji, Yu Wang
Format: Article
Language:English
Published: Elsevier 2012-02-01
Series:Journal of Pharmaceutical Analysis
Online Access:http://www.sciencedirect.com/science/article/pii/S209517791100044X
Description
Summary:Objective: To develop the representative fingerprint for the quality control of placenta polypeptide injection. Methods: The chromatographic separation was performed using a Phenomenex Gemini C18 column (250 mmÃ4.6 mm, 5 μm) maintained at 30 °C. 0.1% aqueous trifluoroacetic acid (Solvent A) and acetonitrile contained 0.1% TFA (Solvent B) were used as mobile phase with a gradient elution. Detection wavelength was 280 nm with the sample injection volume of 50 μL; the flow rate was 1.0 mL/min. The fingerprints of different samples were investigated by similarity analysis. Results: Nine peaks were identified as the characteristic common peaks. The similarities of the fingerprints of the 10 batches of samples were above 0.992. Conclusion: This method showed high precision and good repeatability, and provided the basis for the improvement of the quality control of placenta polypeptide injection. Keywords: Placenta polypeptide injection, Fingerprint, Similarity analysis, High performance liquid chromatography
ISSN:2095-1779