Riata lead failure presenting as a life-threatening electrical storm: A novel manifestation of electrical lead failure

The Riata and Riata ST family of implantable cardioverter-defibrillator leads are prone to a unique type of structural failure involving exteriorization of the conductor cables, which may present as electrical failure. We report a mode of lead failure that occurred in a patient with a Riata 1570 ser...

Full description

Bibliographic Details
Main Authors: Jayaprakash Shenthar, MD, DM, FACC, Somasekhar Ghanta, MD, DM, Maneesh K. Rai, MD, DM, Rohit Walia, MD, DM
Format: Article
Language:English
Published: Wiley 2014-12-01
Series:Journal of Arrhythmia
Subjects:
Online Access:http://www.sciencedirect.com/science/article/pii/S1880427614000374
id doaj-e20ee8e7ef43460bbe73fb5be9529aeb
record_format Article
spelling doaj-e20ee8e7ef43460bbe73fb5be9529aeb2020-11-24T20:44:34ZengWileyJournal of Arrhythmia1880-42762014-12-0130652552810.1016/j.joa.2014.03.003Riata lead failure presenting as a life-threatening electrical storm: A novel manifestation of electrical lead failureJayaprakash Shenthar, MD, DM, FACCSomasekhar Ghanta, MD, DMManeesh K. Rai, MD, DMRohit Walia, MD, DMThe Riata and Riata ST family of implantable cardioverter-defibrillator leads are prone to a unique type of structural failure involving exteriorization of the conductor cables, which may present as electrical failure. We report a mode of lead failure that occurred in a patient with a Riata 1570 series dual coil 8F lead. In this case, the first appropriate shock for ventricular fibrillation resulted in noise, that in turn led to recurrent inappropriate shocks and proarrhythmia, that clinically mimicked a life-threatening electrical storm.http://www.sciencedirect.com/science/article/pii/S1880427614000374Implantable cardioverter defibrillator lead malfunctionICD shock deliveryRiata lead failureVentricular fibrillationElectrical storm
collection DOAJ
language English
format Article
sources DOAJ
author Jayaprakash Shenthar, MD, DM, FACC
Somasekhar Ghanta, MD, DM
Maneesh K. Rai, MD, DM
Rohit Walia, MD, DM
spellingShingle Jayaprakash Shenthar, MD, DM, FACC
Somasekhar Ghanta, MD, DM
Maneesh K. Rai, MD, DM
Rohit Walia, MD, DM
Riata lead failure presenting as a life-threatening electrical storm: A novel manifestation of electrical lead failure
Journal of Arrhythmia
Implantable cardioverter defibrillator lead malfunction
ICD shock delivery
Riata lead failure
Ventricular fibrillation
Electrical storm
author_facet Jayaprakash Shenthar, MD, DM, FACC
Somasekhar Ghanta, MD, DM
Maneesh K. Rai, MD, DM
Rohit Walia, MD, DM
author_sort Jayaprakash Shenthar, MD, DM, FACC
title Riata lead failure presenting as a life-threatening electrical storm: A novel manifestation of electrical lead failure
title_short Riata lead failure presenting as a life-threatening electrical storm: A novel manifestation of electrical lead failure
title_full Riata lead failure presenting as a life-threatening electrical storm: A novel manifestation of electrical lead failure
title_fullStr Riata lead failure presenting as a life-threatening electrical storm: A novel manifestation of electrical lead failure
title_full_unstemmed Riata lead failure presenting as a life-threatening electrical storm: A novel manifestation of electrical lead failure
title_sort riata lead failure presenting as a life-threatening electrical storm: a novel manifestation of electrical lead failure
publisher Wiley
series Journal of Arrhythmia
issn 1880-4276
publishDate 2014-12-01
description The Riata and Riata ST family of implantable cardioverter-defibrillator leads are prone to a unique type of structural failure involving exteriorization of the conductor cables, which may present as electrical failure. We report a mode of lead failure that occurred in a patient with a Riata 1570 series dual coil 8F lead. In this case, the first appropriate shock for ventricular fibrillation resulted in noise, that in turn led to recurrent inappropriate shocks and proarrhythmia, that clinically mimicked a life-threatening electrical storm.
topic Implantable cardioverter defibrillator lead malfunction
ICD shock delivery
Riata lead failure
Ventricular fibrillation
Electrical storm
url http://www.sciencedirect.com/science/article/pii/S1880427614000374
work_keys_str_mv AT jayaprakashshentharmddmfacc riataleadfailurepresentingasalifethreateningelectricalstormanovelmanifestationofelectricalleadfailure
AT somasekharghantamddm riataleadfailurepresentingasalifethreateningelectricalstormanovelmanifestationofelectricalleadfailure
AT maneeshkraimddm riataleadfailurepresentingasalifethreateningelectricalstormanovelmanifestationofelectricalleadfailure
AT rohitwaliamddm riataleadfailurepresentingasalifethreateningelectricalstormanovelmanifestationofelectricalleadfailure
_version_ 1716817035794579456