Investigation of the BNP level changes in blood stream in different modes and lead locations after pacemaker implementation

As has been proven, increase of mechanical strain could result in an increase of brain natriuretic peptide (BNP) in the blood stream of implanted patient pacemakers. We measured the BNP concentration in blood due to different mode and lead implantation location of pacemaker in the time period of 3 m...

Full description

Bibliographic Details
Main Authors: Alireza Mehri Dehnavi, Afshin Fakhrpour, Mohamad Bagher Tavakoli, Mohamad Hossein Nikoo
Format: Article
Language:English
Published: Wolters Kluwer Medknow Publications 2012-01-01
Series:Journal of Medical Signals and Sensors
Subjects:
BNP
Online Access:http://www.jmss.mui.ac.ir/article.asp?issn=2228-7477;year=2012;volume=2;issue=1;spage=38;epage=41;aulast=Dehnavi
id doaj-e0cc6ea2de3549bea100369fc5e8ae9f
record_format Article
spelling doaj-e0cc6ea2de3549bea100369fc5e8ae9f2020-11-24T22:42:57ZengWolters Kluwer Medknow PublicationsJournal of Medical Signals and Sensors2228-74772012-01-01213841Investigation of the BNP level changes in blood stream in different modes and lead locations after pacemaker implementationAlireza Mehri DehnaviAfshin FakhrpourMohamad Bagher TavakoliMohamad Hossein NikooAs has been proven, increase of mechanical strain could result in an increase of brain natriuretic peptide (BNP) in the blood stream of implanted patient pacemakers. We measured the BNP concentration in blood due to different mode and lead implantation location of pacemaker in the time period of 3 months. The aim of this study was to investigate the changing trend of BNP level after pacemaker implantation. One hundred and three pacemaker implanted patients were monitored. Patients were in the age span of 54±12 years, including 48 men and 55 women. A group of 44 were programmed in Dual Chamber Rate Adaptive (DDDR) Pacemaker mode and a group of 59 were programmed in Ventricular Rate Modulated Pacing (VVIR) mode by the recommendation of the cardiologist. Between these two groups, the pacing levels of pacemakers was divided to under and above 50%. Some of these pacemaker leads were located at the apex of the right ventricle and the others were located in the septum wall in the right ventricle. To evaluate BNP changes during a period of 3 months, the BNPs were measured in pg/ml within 24 h of implantation (BNP1) and after 3 months (BNP2). For different classes of pacemaker implantations, the ratio of final measurement (BNP2) is divided to after implantation measurements (BNP1). Results showed that in VVIR mode, the ratio is 1.54±0.3 and in DDDR mode, the ratio is 0.38±0.17, with acceptable standard error means (<0.04). Also, comparisons are made for lead location at two modes of DDDR and VVIR separately. In the DDDR mode, the ratio for apex location is 0.49±0.12 and for septum location is 0.22±0.34, with acceptable standard error means (<0.02). In the VVIR mode, the ratio for apex location is 1.71±0.27 and for septum location is 1.28±0.09, with acceptable standard error means (<0.04). Therefore, BNP decrease in DDDR mode is more than in VVIR mode programming. In both cases of DDDR and VVIR modes, the septum location of the leads would result in a greater decrease of BNP.http://www.jmss.mui.ac.ir/article.asp?issn=2228-7477;year=2012;volume=2;issue=1;spage=38;epage=41;aulast=DehnaviPacemakerBNPmodeimplantation position
collection DOAJ
language English
format Article
sources DOAJ
author Alireza Mehri Dehnavi
Afshin Fakhrpour
Mohamad Bagher Tavakoli
Mohamad Hossein Nikoo
spellingShingle Alireza Mehri Dehnavi
Afshin Fakhrpour
Mohamad Bagher Tavakoli
Mohamad Hossein Nikoo
Investigation of the BNP level changes in blood stream in different modes and lead locations after pacemaker implementation
Journal of Medical Signals and Sensors
Pacemaker
BNP
mode
implantation position
author_facet Alireza Mehri Dehnavi
Afshin Fakhrpour
Mohamad Bagher Tavakoli
Mohamad Hossein Nikoo
author_sort Alireza Mehri Dehnavi
title Investigation of the BNP level changes in blood stream in different modes and lead locations after pacemaker implementation
title_short Investigation of the BNP level changes in blood stream in different modes and lead locations after pacemaker implementation
title_full Investigation of the BNP level changes in blood stream in different modes and lead locations after pacemaker implementation
title_fullStr Investigation of the BNP level changes in blood stream in different modes and lead locations after pacemaker implementation
title_full_unstemmed Investigation of the BNP level changes in blood stream in different modes and lead locations after pacemaker implementation
title_sort investigation of the bnp level changes in blood stream in different modes and lead locations after pacemaker implementation
publisher Wolters Kluwer Medknow Publications
series Journal of Medical Signals and Sensors
issn 2228-7477
publishDate 2012-01-01
description As has been proven, increase of mechanical strain could result in an increase of brain natriuretic peptide (BNP) in the blood stream of implanted patient pacemakers. We measured the BNP concentration in blood due to different mode and lead implantation location of pacemaker in the time period of 3 months. The aim of this study was to investigate the changing trend of BNP level after pacemaker implantation. One hundred and three pacemaker implanted patients were monitored. Patients were in the age span of 54±12 years, including 48 men and 55 women. A group of 44 were programmed in Dual Chamber Rate Adaptive (DDDR) Pacemaker mode and a group of 59 were programmed in Ventricular Rate Modulated Pacing (VVIR) mode by the recommendation of the cardiologist. Between these two groups, the pacing levels of pacemakers was divided to under and above 50%. Some of these pacemaker leads were located at the apex of the right ventricle and the others were located in the septum wall in the right ventricle. To evaluate BNP changes during a period of 3 months, the BNPs were measured in pg/ml within 24 h of implantation (BNP1) and after 3 months (BNP2). For different classes of pacemaker implantations, the ratio of final measurement (BNP2) is divided to after implantation measurements (BNP1). Results showed that in VVIR mode, the ratio is 1.54±0.3 and in DDDR mode, the ratio is 0.38±0.17, with acceptable standard error means (<0.04). Also, comparisons are made for lead location at two modes of DDDR and VVIR separately. In the DDDR mode, the ratio for apex location is 0.49±0.12 and for septum location is 0.22±0.34, with acceptable standard error means (<0.02). In the VVIR mode, the ratio for apex location is 1.71±0.27 and for septum location is 1.28±0.09, with acceptable standard error means (<0.04). Therefore, BNP decrease in DDDR mode is more than in VVIR mode programming. In both cases of DDDR and VVIR modes, the septum location of the leads would result in a greater decrease of BNP.
topic Pacemaker
BNP
mode
implantation position
url http://www.jmss.mui.ac.ir/article.asp?issn=2228-7477;year=2012;volume=2;issue=1;spage=38;epage=41;aulast=Dehnavi
work_keys_str_mv AT alirezamehridehnavi investigationofthebnplevelchangesinbloodstreamindifferentmodesandleadlocationsafterpacemakerimplementation
AT afshinfakhrpour investigationofthebnplevelchangesinbloodstreamindifferentmodesandleadlocationsafterpacemakerimplementation
AT mohamadbaghertavakoli investigationofthebnplevelchangesinbloodstreamindifferentmodesandleadlocationsafterpacemakerimplementation
AT mohamadhosseinnikoo investigationofthebnplevelchangesinbloodstreamindifferentmodesandleadlocationsafterpacemakerimplementation
_version_ 1725697652723548160