Quantitative Assessment of the Impact of Blood Pulsation on Intraocular Pressure Measurement Results in Healthy Subjects

Background. Blood pulsation affects the results obtained using various medical devices in many different ways. Method. The paper proves the effect of blood pulsation on intraocular pressure measurements. Six measurements for each of the 10 healthy subjects were performed in various phases of blood p...

Full description

Bibliographic Details
Main Authors: Robert Koprowski, Lei Tian
Format: Article
Language:English
Published: Hindawi Limited 2017-01-01
Series:Journal of Ophthalmology
Online Access:http://dx.doi.org/10.1155/2017/9678041
id doaj-ddf26652d1974bd48ac825b1b6c8100a
record_format Article
spelling doaj-ddf26652d1974bd48ac825b1b6c8100a2020-11-25T00:47:25ZengHindawi LimitedJournal of Ophthalmology2090-004X2090-00582017-01-01201710.1155/2017/96780419678041Quantitative Assessment of the Impact of Blood Pulsation on Intraocular Pressure Measurement Results in Healthy SubjectsRobert Koprowski0Lei Tian1Department of Biomedical Computer Systems, University of Silesia, Faculty of Computer Science and Materials Science, Institute of Computer Science, Ul. Będzińska 39, 41-200 Sosnowiec, PolandBeijing Institute of Ophthalmology, Beijing Tongren Eye Center, Beijing Tongren Hospital, Capital Medical University, Beijing Ophthalmology & Visual Sciences Key Laboratory, Beijing 100730, ChinaBackground. Blood pulsation affects the results obtained using various medical devices in many different ways. Method. The paper proves the effect of blood pulsation on intraocular pressure measurements. Six measurements for each of the 10 healthy subjects were performed in various phases of blood pulsation. A total of 8400 corneal deformation images were recorded. The results of intraocular pressure measurements were related to the results of heartbeat phases measured with a pulse oximeter placed on the index finger of the subject’s left hand. Results. The correlation between the heartbeat phase measured with a pulse oximeter and intraocular pressure is 0.69±0.26 (p<0.05). The phase shift calculated for the maximum correlation is equal to 60±40° (p<0.05). When the moment of measuring intraocular pressure with an air-puff tonometer is not synchronized, the changes in IOP for the analysed group of subjects can vary in the range of ±2.31 mmHg (p<0.3). Conclusions. Blood pulsation has a statistically significant effect on the results of intraocular pressure measurement. For this reason, in modern ophthalmic devices, the measurement should be synchronized with the heartbeat phases. The paper proposes an additional method for synchronizing the time of pressure measurement with the blood pulsation phase.http://dx.doi.org/10.1155/2017/9678041
collection DOAJ
language English
format Article
sources DOAJ
author Robert Koprowski
Lei Tian
spellingShingle Robert Koprowski
Lei Tian
Quantitative Assessment of the Impact of Blood Pulsation on Intraocular Pressure Measurement Results in Healthy Subjects
Journal of Ophthalmology
author_facet Robert Koprowski
Lei Tian
author_sort Robert Koprowski
title Quantitative Assessment of the Impact of Blood Pulsation on Intraocular Pressure Measurement Results in Healthy Subjects
title_short Quantitative Assessment of the Impact of Blood Pulsation on Intraocular Pressure Measurement Results in Healthy Subjects
title_full Quantitative Assessment of the Impact of Blood Pulsation on Intraocular Pressure Measurement Results in Healthy Subjects
title_fullStr Quantitative Assessment of the Impact of Blood Pulsation on Intraocular Pressure Measurement Results in Healthy Subjects
title_full_unstemmed Quantitative Assessment of the Impact of Blood Pulsation on Intraocular Pressure Measurement Results in Healthy Subjects
title_sort quantitative assessment of the impact of blood pulsation on intraocular pressure measurement results in healthy subjects
publisher Hindawi Limited
series Journal of Ophthalmology
issn 2090-004X
2090-0058
publishDate 2017-01-01
description Background. Blood pulsation affects the results obtained using various medical devices in many different ways. Method. The paper proves the effect of blood pulsation on intraocular pressure measurements. Six measurements for each of the 10 healthy subjects were performed in various phases of blood pulsation. A total of 8400 corneal deformation images were recorded. The results of intraocular pressure measurements were related to the results of heartbeat phases measured with a pulse oximeter placed on the index finger of the subject’s left hand. Results. The correlation between the heartbeat phase measured with a pulse oximeter and intraocular pressure is 0.69±0.26 (p<0.05). The phase shift calculated for the maximum correlation is equal to 60±40° (p<0.05). When the moment of measuring intraocular pressure with an air-puff tonometer is not synchronized, the changes in IOP for the analysed group of subjects can vary in the range of ±2.31 mmHg (p<0.3). Conclusions. Blood pulsation has a statistically significant effect on the results of intraocular pressure measurement. For this reason, in modern ophthalmic devices, the measurement should be synchronized with the heartbeat phases. The paper proposes an additional method for synchronizing the time of pressure measurement with the blood pulsation phase.
url http://dx.doi.org/10.1155/2017/9678041
work_keys_str_mv AT robertkoprowski quantitativeassessmentoftheimpactofbloodpulsationonintraocularpressuremeasurementresultsinhealthysubjects
AT leitian quantitativeassessmentoftheimpactofbloodpulsationonintraocularpressuremeasurementresultsinhealthysubjects
_version_ 1725260069464965120