Interference of monitored electrocardiogram is common during different surgical procedures using electrical equipment. The electrical devices used induce artifacts in the electrocardiographic tracing, which may resemble serious arrhythmia. We describe a case of electrocardiographic artifact resembli...

Full description

Bibliographic Details
Main Authors: Sofi Khalid, Khalid Ab Salam, Mushtaq Gilani Kurdi
Format: Article
Language:English
Published: Wolters Kluwer Medknow Publications 2010-01-01
Series:Saudi Journal of Anaesthesia
Subjects:
Online Access:http://www.saudija.org/article.asp?issn=1658-354X;year=2010;volume=4;issue=2;spage=111;epage=113;aulast=Sofi
Description
Summary:Interference of monitored electrocardiogram is common during different surgical procedures using electrical equipment. The electrical devices used induce artifacts in the electrocardiographic tracing, which may resemble serious arrhythmia. We describe a case of electrocardiographic artifact resembling ventricular tachycardia with the use of a Storz unidrive microdebrider during inferior turbinectomy under general anesthesia. This case report highlights the importance of knowledge of various equipment-related electrocardiographic artifacts in avoiding unnecessary and harmful therapeutic interventions.