Accuracy and Reliability of AG501 Articulograph for Mandibular Movement Analysis: A Quantitative Descriptive Study

Electromagnetic articulography (EMA) have been mostly employed to study articulatory movements of speech. This technology appears to be very promising for studying mandibular movements within the field of dentistry. However, there are no studies reporting the validity of EMA for such purpose. The ai...

Full description

Bibliographic Details
Main Authors: María Florencia Lezcano, Fernando Dias, Alain Arias, Ramón Fuentes
Format: Article
Language:English
Published: MDPI AG 2020-11-01
Series:Sensors
Subjects:
Online Access:https://www.mdpi.com/1424-8220/20/21/6324
id doaj-c9563f83e550434b9a5b58bd206c94fe
record_format Article
spelling doaj-c9563f83e550434b9a5b58bd206c94fe2020-11-25T03:57:29ZengMDPI AGSensors1424-82202020-11-01206324632410.3390/s20216324Accuracy and Reliability of AG501 Articulograph for Mandibular Movement Analysis: A Quantitative Descriptive StudyMaría Florencia Lezcano0Fernando Dias1Alain Arias2Ramón Fuentes3Research Centre for Dental Sciences CICO, Department of Integral Adults Dentistry, Dental School, Universidad de La Frontera, Temuco 4811230, ChileResearch Centre for Dental Sciences CICO, Department of Integral Adults Dentistry, Dental School, Universidad de La Frontera, Temuco 4811230, ChileResearch Centre for Dental Sciences CICO, Department of Integral Adults Dentistry, Dental School, Universidad de La Frontera, Temuco 4811230, ChileResearch Centre for Dental Sciences CICO, Department of Integral Adults Dentistry, Dental School, Universidad de La Frontera, Temuco 4811230, ChileElectromagnetic articulography (EMA) have been mostly employed to study articulatory movements of speech. This technology appears to be very promising for studying mandibular movements within the field of dentistry. However, there are no studies reporting the validity of EMA for such purpose. The aim of this study is to assess accuracy and reliability of Carstens three-dimensional EMA AG501 in order to validate its use for mandibular movement analysis in dentistry. A set of tests was conducted attaching 16 sensors to a rotating rigid structure placed inside the measurement area. Another set of tests were conducted using a mouth anatomical model with human-like articulatory behaviour. A function of the EMA system called “head correction” was applied to normalize the data of every recording. The system reliability was higher at the centre of the measurement area and decreased toward the edges. Dispersion was greater for raw data than for normalized data. Bland-Altman analysis of agreement between the AG501 and a millimetre ruler used to measure the distance between the sensors revealed limits of agreement between 0.5 mm and −0.9 mm. The results suggest that EMA AG501 is valid for three-dimensional analysis of mandibular biomechanics allowing natural movements.https://www.mdpi.com/1424-8220/20/21/6324dentistrybiomechanicselectromagnetic articulographymandibular movements
collection DOAJ
language English
format Article
sources DOAJ
author María Florencia Lezcano
Fernando Dias
Alain Arias
Ramón Fuentes
spellingShingle María Florencia Lezcano
Fernando Dias
Alain Arias
Ramón Fuentes
Accuracy and Reliability of AG501 Articulograph for Mandibular Movement Analysis: A Quantitative Descriptive Study
Sensors
dentistry
biomechanics
electromagnetic articulography
mandibular movements
author_facet María Florencia Lezcano
Fernando Dias
Alain Arias
Ramón Fuentes
author_sort María Florencia Lezcano
title Accuracy and Reliability of AG501 Articulograph for Mandibular Movement Analysis: A Quantitative Descriptive Study
title_short Accuracy and Reliability of AG501 Articulograph for Mandibular Movement Analysis: A Quantitative Descriptive Study
title_full Accuracy and Reliability of AG501 Articulograph for Mandibular Movement Analysis: A Quantitative Descriptive Study
title_fullStr Accuracy and Reliability of AG501 Articulograph for Mandibular Movement Analysis: A Quantitative Descriptive Study
title_full_unstemmed Accuracy and Reliability of AG501 Articulograph for Mandibular Movement Analysis: A Quantitative Descriptive Study
title_sort accuracy and reliability of ag501 articulograph for mandibular movement analysis: a quantitative descriptive study
publisher MDPI AG
series Sensors
issn 1424-8220
publishDate 2020-11-01
description Electromagnetic articulography (EMA) have been mostly employed to study articulatory movements of speech. This technology appears to be very promising for studying mandibular movements within the field of dentistry. However, there are no studies reporting the validity of EMA for such purpose. The aim of this study is to assess accuracy and reliability of Carstens three-dimensional EMA AG501 in order to validate its use for mandibular movement analysis in dentistry. A set of tests was conducted attaching 16 sensors to a rotating rigid structure placed inside the measurement area. Another set of tests were conducted using a mouth anatomical model with human-like articulatory behaviour. A function of the EMA system called “head correction” was applied to normalize the data of every recording. The system reliability was higher at the centre of the measurement area and decreased toward the edges. Dispersion was greater for raw data than for normalized data. Bland-Altman analysis of agreement between the AG501 and a millimetre ruler used to measure the distance between the sensors revealed limits of agreement between 0.5 mm and −0.9 mm. The results suggest that EMA AG501 is valid for three-dimensional analysis of mandibular biomechanics allowing natural movements.
topic dentistry
biomechanics
electromagnetic articulography
mandibular movements
url https://www.mdpi.com/1424-8220/20/21/6324
work_keys_str_mv AT mariaflorencialezcano accuracyandreliabilityofag501articulographformandibularmovementanalysisaquantitativedescriptivestudy
AT fernandodias accuracyandreliabilityofag501articulographformandibularmovementanalysisaquantitativedescriptivestudy
AT alainarias accuracyandreliabilityofag501articulographformandibularmovementanalysisaquantitativedescriptivestudy
AT ramonfuentes accuracyandreliabilityofag501articulographformandibularmovementanalysisaquantitativedescriptivestudy
_version_ 1724460453381275648