INDIRECT METHODS OF MEASURING FLОW RATE AND PRESSURE DROP OF ROTARY BLOOD PUMPS

The article describes complex software to calculate the main parameters of the axial pump (pressure drop and flow rate) on the basis of an analysis of its technical characteristics (current consumption and pump speed). It is shown that the calculated value of pressure drop and flow rate throughout t...

Full description

Bibliographic Details
Main Authors: G. P. Itkin, I. A. Filatov, K. N. Dozorov, A. V. Adaskin
Format: Article
Language:Russian
Published: Federal Research Center of Transplantology and Artificial Organs named after V.I.Shumakov 2015-04-01
Series:Vestnik Transplantologii i Iskusstvennyh Organov
Subjects:
Online Access:https://journal.transpl.ru/vtio/article/view/501
Description
Summary:The article describes complex software to calculate the main parameters of the axial pump (pressure drop and flow rate) on the basis of an analysis of its technical characteristics (current consumption and pump speed). It is shown that the calculated value of pressure drop and flow rate throughout the range of operating loads and pump speeds is different from the measured one by no more than 10%. It is shown that this software package may be included in the control system of axial pump.
ISSN:1995-1191