Consistency analysis of Kappa angle measurement by Pentacam and Keratron Scout

AIM: To investigate the repeatability and agreement of Pentacam Front Section Analyzer and Keratron Scout Corneal Topographer in measuring Kappa angle.<p>METHODS: Totally 69 patients were randomly selected from Ophthalmology Department of Leshan People's Hospital. Three readings were take...

Full description

Bibliographic Details
Main Authors: Jiao Zhao, Geng-Sheng Hao, Li Zeng, Zong-Shun Liu, Bu-Dan Hu, Ke Mao
Format: Article
Language:English
Published: Press of International Journal of Ophthalmology (IJO PRESS) 2019-07-01
Series:Guoji Yanke Zazhi
Subjects:
Online Access:http://ies.ijo.cn/cn_publish/2019/7/201907043.pdf
id doaj-c6dbc84a58de46ddbc5db759f17ce8de
record_format Article
spelling doaj-c6dbc84a58de46ddbc5db759f17ce8de2020-11-25T02:45:48ZengPress of International Journal of Ophthalmology (IJO PRESS)Guoji Yanke Zazhi1672-51231672-51232019-07-011971260126210.3980/j.issn.1672-5123.2019.7.43Consistency analysis of Kappa angle measurement by Pentacam and Keratron ScoutJiao Zhao0Geng-Sheng Hao1Li Zeng2Zong-Shun Liu3Bu-Dan Hu4Ke Mao5Department of Ophthalmology, People's Hospital of Leshan, Leshan 614000, Sichuan Province, ChinaDepartment of Ophthalmology, People's Hospital of Leshan, Leshan 614000, Sichuan Province, ChinaDepartment of Ophthalmology, People's Hospital of Leshan, Leshan 614000, Sichuan Province, ChinaDepartment of Ophthalmology, People's Hospital of Leshan, Leshan 614000, Sichuan Province, ChinaDepartment of Ophthalmology, People's Hospital of Leshan, Leshan 614000, Sichuan Province, ChinaDepartment of Ophthalmology, People's Hospital of Leshan, Leshan 614000, Sichuan Province, ChinaAIM: To investigate the repeatability and agreement of Pentacam Front Section Analyzer and Keratron Scout Corneal Topographer in measuring Kappa angle.<p>METHODS: Totally 69 patients were randomly selected from Ophthalmology Department of Leshan People's Hospital. Three readings were taken by the same examiner. Results were recorded in(<i>X</i>, <i>Y</i>)coordinates format and assessed by intraclass correlation coefficient(ICC), Cronbach's Alpha Coefficient, <i>t-</i>test, Pearson correlation analysis and bland-altman analysis.<p>RESULTS: Both instruments showed excellent repeatability in Kappa angle assessment. The results significantly correlated(<i>X</i> value: <i>r</i>=0.90,<i> P</i><0.01; <i>Y</i> value: <i>r</i>=0.91, <i>P</i><0.01), however, no significant difference were found between the two instruments. The widest 95% LoA was -0.11-0.14mm for <i>X</i> coordinate and -0.10-0.11mm for <i>Y</i>.<p>CONCLUSION: Pentacam and Keratron Scout showed excellent repeatability and agreement in Kappa angle assessment. The two instruments could aid and verify each other.http://ies.ijo.cn/cn_publish/2019/7/201907043.pdfPentacam Front Section AnalyzerKeratron Scout Corneal TopographerKappa angleagreementrepeatability
collection DOAJ
language English
format Article
sources DOAJ
author Jiao Zhao
Geng-Sheng Hao
Li Zeng
Zong-Shun Liu
Bu-Dan Hu
Ke Mao
spellingShingle Jiao Zhao
Geng-Sheng Hao
Li Zeng
Zong-Shun Liu
Bu-Dan Hu
Ke Mao
Consistency analysis of Kappa angle measurement by Pentacam and Keratron Scout
Guoji Yanke Zazhi
Pentacam Front Section Analyzer
Keratron Scout Corneal Topographer
Kappa angle
agreement
repeatability
author_facet Jiao Zhao
Geng-Sheng Hao
Li Zeng
Zong-Shun Liu
Bu-Dan Hu
Ke Mao
author_sort Jiao Zhao
title Consistency analysis of Kappa angle measurement by Pentacam and Keratron Scout
title_short Consistency analysis of Kappa angle measurement by Pentacam and Keratron Scout
title_full Consistency analysis of Kappa angle measurement by Pentacam and Keratron Scout
title_fullStr Consistency analysis of Kappa angle measurement by Pentacam and Keratron Scout
title_full_unstemmed Consistency analysis of Kappa angle measurement by Pentacam and Keratron Scout
title_sort consistency analysis of kappa angle measurement by pentacam and keratron scout
publisher Press of International Journal of Ophthalmology (IJO PRESS)
series Guoji Yanke Zazhi
issn 1672-5123
1672-5123
publishDate 2019-07-01
description AIM: To investigate the repeatability and agreement of Pentacam Front Section Analyzer and Keratron Scout Corneal Topographer in measuring Kappa angle.<p>METHODS: Totally 69 patients were randomly selected from Ophthalmology Department of Leshan People's Hospital. Three readings were taken by the same examiner. Results were recorded in(<i>X</i>, <i>Y</i>)coordinates format and assessed by intraclass correlation coefficient(ICC), Cronbach's Alpha Coefficient, <i>t-</i>test, Pearson correlation analysis and bland-altman analysis.<p>RESULTS: Both instruments showed excellent repeatability in Kappa angle assessment. The results significantly correlated(<i>X</i> value: <i>r</i>=0.90,<i> P</i><0.01; <i>Y</i> value: <i>r</i>=0.91, <i>P</i><0.01), however, no significant difference were found between the two instruments. The widest 95% LoA was -0.11-0.14mm for <i>X</i> coordinate and -0.10-0.11mm for <i>Y</i>.<p>CONCLUSION: Pentacam and Keratron Scout showed excellent repeatability and agreement in Kappa angle assessment. The two instruments could aid and verify each other.
topic Pentacam Front Section Analyzer
Keratron Scout Corneal Topographer
Kappa angle
agreement
repeatability
url http://ies.ijo.cn/cn_publish/2019/7/201907043.pdf
work_keys_str_mv AT jiaozhao consistencyanalysisofkappaanglemeasurementbypentacamandkeratronscout
AT gengshenghao consistencyanalysisofkappaanglemeasurementbypentacamandkeratronscout
AT lizeng consistencyanalysisofkappaanglemeasurementbypentacamandkeratronscout
AT zongshunliu consistencyanalysisofkappaanglemeasurementbypentacamandkeratronscout
AT budanhu consistencyanalysisofkappaanglemeasurementbypentacamandkeratronscout
AT kemao consistencyanalysisofkappaanglemeasurementbypentacamandkeratronscout
_version_ 1724760026188349440