Consistency analysis of Kappa angle measurement by Pentacam and Keratron Scout
AIM: To investigate the repeatability and agreement of Pentacam Front Section Analyzer and Keratron Scout Corneal Topographer in measuring Kappa angle.<p>METHODS: Totally 69 patients were randomly selected from Ophthalmology Department of Leshan People's Hospital. Three readings were take...
Main Authors: | , , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Press of International Journal of Ophthalmology (IJO PRESS)
2019-07-01
|
Series: | Guoji Yanke Zazhi |
Subjects: | |
Online Access: | http://ies.ijo.cn/cn_publish/2019/7/201907043.pdf |
id |
doaj-c6dbc84a58de46ddbc5db759f17ce8de |
---|---|
record_format |
Article |
spelling |
doaj-c6dbc84a58de46ddbc5db759f17ce8de2020-11-25T02:45:48ZengPress of International Journal of Ophthalmology (IJO PRESS)Guoji Yanke Zazhi1672-51231672-51232019-07-011971260126210.3980/j.issn.1672-5123.2019.7.43Consistency analysis of Kappa angle measurement by Pentacam and Keratron ScoutJiao Zhao0Geng-Sheng Hao1Li Zeng2Zong-Shun Liu3Bu-Dan Hu4Ke Mao5Department of Ophthalmology, People's Hospital of Leshan, Leshan 614000, Sichuan Province, ChinaDepartment of Ophthalmology, People's Hospital of Leshan, Leshan 614000, Sichuan Province, ChinaDepartment of Ophthalmology, People's Hospital of Leshan, Leshan 614000, Sichuan Province, ChinaDepartment of Ophthalmology, People's Hospital of Leshan, Leshan 614000, Sichuan Province, ChinaDepartment of Ophthalmology, People's Hospital of Leshan, Leshan 614000, Sichuan Province, ChinaDepartment of Ophthalmology, People's Hospital of Leshan, Leshan 614000, Sichuan Province, ChinaAIM: To investigate the repeatability and agreement of Pentacam Front Section Analyzer and Keratron Scout Corneal Topographer in measuring Kappa angle.<p>METHODS: Totally 69 patients were randomly selected from Ophthalmology Department of Leshan People's Hospital. Three readings were taken by the same examiner. Results were recorded in(<i>X</i>, <i>Y</i>)coordinates format and assessed by intraclass correlation coefficient(ICC), Cronbach's Alpha Coefficient, <i>t-</i>test, Pearson correlation analysis and bland-altman analysis.<p>RESULTS: Both instruments showed excellent repeatability in Kappa angle assessment. The results significantly correlated(<i>X</i> value: <i>r</i>=0.90,<i> P</i><0.01; <i>Y</i> value: <i>r</i>=0.91, <i>P</i><0.01), however, no significant difference were found between the two instruments. The widest 95% LoA was -0.11-0.14mm for <i>X</i> coordinate and -0.10-0.11mm for <i>Y</i>.<p>CONCLUSION: Pentacam and Keratron Scout showed excellent repeatability and agreement in Kappa angle assessment. The two instruments could aid and verify each other.http://ies.ijo.cn/cn_publish/2019/7/201907043.pdfPentacam Front Section AnalyzerKeratron Scout Corneal TopographerKappa angleagreementrepeatability |
collection |
DOAJ |
language |
English |
format |
Article |
sources |
DOAJ |
author |
Jiao Zhao Geng-Sheng Hao Li Zeng Zong-Shun Liu Bu-Dan Hu Ke Mao |
spellingShingle |
Jiao Zhao Geng-Sheng Hao Li Zeng Zong-Shun Liu Bu-Dan Hu Ke Mao Consistency analysis of Kappa angle measurement by Pentacam and Keratron Scout Guoji Yanke Zazhi Pentacam Front Section Analyzer Keratron Scout Corneal Topographer Kappa angle agreement repeatability |
author_facet |
Jiao Zhao Geng-Sheng Hao Li Zeng Zong-Shun Liu Bu-Dan Hu Ke Mao |
author_sort |
Jiao Zhao |
title |
Consistency analysis of Kappa angle measurement by Pentacam and Keratron Scout |
title_short |
Consistency analysis of Kappa angle measurement by Pentacam and Keratron Scout |
title_full |
Consistency analysis of Kappa angle measurement by Pentacam and Keratron Scout |
title_fullStr |
Consistency analysis of Kappa angle measurement by Pentacam and Keratron Scout |
title_full_unstemmed |
Consistency analysis of Kappa angle measurement by Pentacam and Keratron Scout |
title_sort |
consistency analysis of kappa angle measurement by pentacam and keratron scout |
publisher |
Press of International Journal of Ophthalmology (IJO PRESS) |
series |
Guoji Yanke Zazhi |
issn |
1672-5123 1672-5123 |
publishDate |
2019-07-01 |
description |
AIM: To investigate the repeatability and agreement of Pentacam Front Section Analyzer and Keratron Scout Corneal Topographer in measuring Kappa angle.<p>METHODS: Totally 69 patients were randomly selected from Ophthalmology Department of Leshan People's Hospital. Three readings were taken by the same examiner. Results were recorded in(<i>X</i>, <i>Y</i>)coordinates format and assessed by intraclass correlation coefficient(ICC), Cronbach's Alpha Coefficient, <i>t-</i>test, Pearson correlation analysis and bland-altman analysis.<p>RESULTS: Both instruments showed excellent repeatability in Kappa angle assessment. The results significantly correlated(<i>X</i> value: <i>r</i>=0.90,<i> P</i><0.01; <i>Y</i> value: <i>r</i>=0.91, <i>P</i><0.01), however, no significant difference were found between the two instruments. The widest 95% LoA was -0.11-0.14mm for <i>X</i> coordinate and -0.10-0.11mm for <i>Y</i>.<p>CONCLUSION: Pentacam and Keratron Scout showed excellent repeatability and agreement in Kappa angle assessment. The two instruments could aid and verify each other. |
topic |
Pentacam Front Section Analyzer Keratron Scout Corneal Topographer Kappa angle agreement repeatability |
url |
http://ies.ijo.cn/cn_publish/2019/7/201907043.pdf |
work_keys_str_mv |
AT jiaozhao consistencyanalysisofkappaanglemeasurementbypentacamandkeratronscout AT gengshenghao consistencyanalysisofkappaanglemeasurementbypentacamandkeratronscout AT lizeng consistencyanalysisofkappaanglemeasurementbypentacamandkeratronscout AT zongshunliu consistencyanalysisofkappaanglemeasurementbypentacamandkeratronscout AT budanhu consistencyanalysisofkappaanglemeasurementbypentacamandkeratronscout AT kemao consistencyanalysisofkappaanglemeasurementbypentacamandkeratronscout |
_version_ |
1724760026188349440 |